Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device.

K L Pasma1, M L Dirkx, M Kroonwijk, A G Visser, B J Heijmen.   

Abstract

Dose distributions can often be significantly improved by modulating the two-dimensional intensity profile of the individual x-ray beams. One technique for delivering intensity modulated beams is dynamic multileaf collimation (DMLC). However, DMLC is complex and requires extensive quality assurance. In this paper a new method is presented for a pretreatment dosimetric verification of these intensity modulated beams utilizing a charge-coupled device camera based fluoroscopic electronic portal imaging device (EPID). In the absence of the patient, EPID images are acquired for all beams produced with DMLC. These images are then converted into two-dimensional dose distributions and compared with the calculated dose distributions. The calculations are performed with a pencil beam algorithm as implemented in a commercially available treatment planning system using the same absolute beam fluence profiles as used for calculation of the patient dose distribution. The method allows an overall verification of (i) the leaf trajectory calculation (including the models to incorporate collimator scatter and leaf transmission), (ii) the correct transfer of the leaf sequencing file to the treatment machine, and (iii) the mechanical and dosimetrical performance of the treatment unit. The method was tested for intensity modulated 10 and 25 MV photon beams; both model cases and real clinical cases were studied. Dose profiles measured with the EPID were also compared with ionization chamber measurements. In all cases both predictions and EPID measurements and EPID and ionization chamber measurements agreed within 2% (1 sigma). The study has demonstrated that the proposed method allows fast and accurate pretreatment verification of DMLC.

Entities:  

Mesh:

Year:  1999        PMID: 10587219     DOI: 10.1118/1.598752

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

1.  Verification of MLC based real-time tumor tracking using an electronic portal imaging device.

Authors:  Sarah Han-Oh; Byong Yong Yi; Fritz Lerma; Barry L Berman; Minzhi Gui; Cedric Yu
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  A study on dosimetric properties of electronic portal imaging device and its use as a quality assurance tool in Volumetric Modulated Arc Therapy.

Authors:  Prabakar Sukumar; Sriram Padmanaban; Prakash Jeevanandam; S A Syam Kumar; Vivekanandan Nagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2011-08-26

3.  Overestimation of low-dose radiation in intensity-modulated radiotherapy with sliding-window technique.

Authors:  Hilke Vorwerk; Daniela Wagner; Björn Seitz; Hans Christiansen; Hendrik A Wolff; Clemens F Hess
Journal:  Strahlenther Onkol       Date:  2009-12       Impact factor: 3.621

4.  Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Authors:  Arpita Agarwal; Nikhil Rastogi; K J Maria Das; S A Yoganathan; D Udayakumar; Shaleen Kumar
Journal:  J Med Phys       Date:  2017 Apr-Jun

5.  Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software.

Authors:  Mareike Held; Joey Cheung; Angelica Perez Andujar; François Husson; Olivier Morin
Journal:  Cureus       Date:  2018-02-02

6.  A fully electronic intensity-modulated radiation therapy quality assurance (IMRT QA) process implemented in a network comprised of independent treatment planning, record and verify, and delivery systems.

Authors:  Daniel W Bailey; Lalith Kumaraswamy; Matthew B Podgorsak
Journal:  Radiol Oncol       Date:  2010-05-24       Impact factor: 2.991

7.  Implementation and validation of a commercial portal dosimetry software for intensity-modulated radiation therapy pre-treatment verification.

Authors:  C Varatharaj; Eugenia Moretti; M Ravikumar; Maria Rosa Malisan; Sanjay S Supe; Renato Padovani
Journal:  J Med Phys       Date:  2010-10

8.  Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification.

Authors:  Ali Jomehzadeh; Parvaneh Shokrani; Mohammad Mohammadi; Alireza Amouheidari
Journal:  Adv Biomed Res       Date:  2016-11-28

9.  A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction.

Authors:  Wendel Dean Renner; Kevin Norton; Timothy Holmes
Journal:  J Appl Clin Med Phys       Date:  2005-11-21       Impact factor: 2.102

10.  The use of extended dose range film for dosimetric calibration of a scanning liquid-filled ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak; Paul Reich
Journal:  J Appl Clin Med Phys       Date:  2006-05-15       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.