Literature DB >> 28639254

Use of a digital phantom developed by QIBA for harmonizing SUVs obtained from the state-of-the-art SPECT/CT systems: a multicenter study.

Tadaki Nakahara1, Hiromitsu Daisaki2, Yasushi Yamamoto3, Takashi Iimori4, Kazuyuki Miyagawa5, Tomoya Okamoto6, Yoshiki Owaki7,8, Nobuhiro Yada3, Koichi Sawada4, Ryotaro Tokorodani5, Masahiro Jinzaki7.   

Abstract

BACKGROUND: Although quantitative analysis using standardized uptake value (SUV) becomes realistic in clinical single-photon emission computed tomography/computed tomography (SPECT/CT) imaging, reconstruction parameter settings can deliver different quantitative results among different SPECT/CT systems. This study aims to propose a use of the digital reference object (DRO), which is a National Electrical Manufacturers Association (NEMA) phantom-like object developed by the Quantitative Imaging Biomarker Alliance (QIBA) fluorodeoxyglucose-positron emission tomography technical committee, for the purpose of harmonizing SUVs in Tc-99m SPECT/CT imaging.
METHODS: The NEMA body phantom with determined Tc-99m concentration was scanned with the four state-of-the-art SPECT/CT systems. SPECT data were reconstructed using different numbers of the product of subset and iteration numbers (SI) and the width of 3D Gaussian filter (3DGF). The mean (SUVmean), maximal (SUVmax), and peak (SUVpeak) SUVs for six hot spheres (10, 13, 17, 22, 28, and 37 mm) were measured after converting SPECT count into SUV using Becquerel calibration factor. DRO smoothed by 3DGF with a FWHM of 17 mm (DRO17 mm) was generated, and the corresponding SUVs were measured. The reconstruction condition to yield the lowest root mean square error (RMSE) of SUVmeans for all the spheres between DRO17 mm and actual phantom images was determined as the harmonized condition for each SPECT/CT scanner. Then, inter-scanner variability in all quantitative metrics was measured before (i.e., according to the manufacturers' recommendation or the policies of their own departments) and after harmonization.
RESULTS: RMSE was lowest in the following reconstruction conditions: SI of 100 and 3DGF of 13 mm for Brightview XCT, SI of 160 and 3DGF of 3 pixels for Discovery NM/CT, SI of 60 and 3DGF of 2 pixels for Infinia, and SI of 140 and 3DGF of 15 mm for Symbia. In pre-harmonized conditions, coefficient of variations (COVs) among the SPECT/CT systems were greater than 10% for all quantitative metrics in three of the spheres, SUVmax and SUVmean, in one of the spheres. In contrast, all metrics except SUVmax in the 17-mm sphere yielded less than 10% of COVs after harmonization.
CONCLUSIONS: Our proposed method clearly reduced inter-scanner variability in SUVs. A digital phantom developed by QIBA would be useful for harmonizing SUVs in multicenter trials using SPECT/CT.

Entities:  

Keywords:  Harmonization; Multicenter study; SPECT/CT; SUV

Year:  2017        PMID: 28639254      PMCID: PMC5479776          DOI: 10.1186/s13550-017-0300-5

Source DB:  PubMed          Journal:  EJNMMI Res            Impact factor:   3.138


Background

Although physical quality of single-photon emission computed tomography/computed tomography (SPECT/CT) images such as image resolution and noise is worse than that of PET/CT images, recent studies suggested the possibility for the clinical application of quantitative SPECT/CT [1-3]. In 2010, Zeintl et al. reported that the advanced SPECT/CT technology facilitated quantitative Tc-99m SPECT imaging with excellent accuracy in both the phantom (error < 3.6%) and patient studies (error < 1.1%) [3]. In 2012, Seret et al. investigated the performance of the four state-of-the-art SPECT/CT systems (Philips Brightview XCT, General Electric Discovery NM/CT 670 and Infinia Hawkeye 4, and Siemens Symbia T6) in quantitative assessment using three-dimensional iterative reconstruction (3D-OSEM) with attenuation and scatter corrections and resolution recovery [1]. Quantitative errors of the four SPECT/CT systems were less than 10% if the targets were several times larger than the spatial resolution of these SPECT devices. In the same year, Hughes et al. also conducted a phantom study in order to compare the images obtained with three different SPECT/CT systems [2]. Interestingly, their study showed no significant differences in image quality when using their own algorithm, whereas image quality was different between images reconstructed with the vendors’ reconstruction software. These results seem to raise a problem with regard to the standardization of SPECT/CT quantitation among different nuclear medicine institutions. At present, common parameters used for quantitation in clinical SPECT/CT are the maximal standardized uptake value (SUVmax) [4, 5] and peak SUV (SUVpeak) [6]. SUV is the ratio of the radioactivity concentration in a voxel of the target to the average radioactivity concentration in the body, and SUVmax is the highest SUV within a volume of interest (VOI). Although SUVmax is preferably used in clinical PET imaging because it is not affected by ROI settings, optimization of reconstruction parameter settings is important to harmonize quantitative metrics among different PET cameras [7]. Since SUV is susceptible to spatial resolution and image noise, reconstruction conditions should be properly adjusted for each camera to provide reliable and robust SUVs in terms of the harmonization of SPECT/CT quantitation. In other words, harmonization-specific imaging protocol is crucial for clinical multicenter trials using quantitative SPECT/CT. This trend has been preceded by fluorodeoxyglucose-positron emission tomography/computed tomography (FDG-PET/CT) for multicenter trials [7, 8]. Recently, the Quantitative Imaging Biomarker Alliance (QIBA) FDG-PET technical committee has developed an FDG-PET/CT digital reference object (DRO) that is a synthetic test object representing an FDG-PET image volume in the Digital Imaging and Communications in Medicine (DICOM) format [9]. The DRO images in both PET and CT are based on the body phantom of National Electrical Manufacturers Association (NEMA) and International Electrotechnical Commission (IEC) [10]. Since the DRO is created synthetically with no random image noise, the DRO can be used as a reference standard to test SUV calculations. Pierce et al. used the DRO, which was smoothed by partial voxel computation in view of finite spatial resolution, to ensure the standardization of SUV computation in PET between medical image viewing workstations [11]. According to the Japanese guideline for oncological FDG-PET/CT imaging in 2009, a DRO-like digital phantom smoothed by a 3D Gaussian filter (3DGF) with a FWHM of 10 mm was used as a reference in order to define prerequisite image quality for detection of a 10-mm hot sphere with SUV of 4 [12]. In the present study, we propose a use of the DRO smoothed by 3DGF with a FWHM of 17 mm (DRO17 mm) for the purpose of harmonizing SUVs in Tc-99m SPECT/CT imaging. Our phantom study using the aforementioned four state-of-the-art SPECT/CT systems to image NEMA phantom showed that a 10-mm hot sphere was undetected and a 13-mm hot sphere was barely discernible, whereas all the scanners clearly depicted a 17-mm sphere. Based on the detectable feature, we hypothesized that DRO17 mm could be used as a reference to determine the harmonization-specific imaging protocol as a digital phantom with a smooth of 10 mm which was used in the Japanese PET guideline [12]. The aim of this study was to demonstrate the feasibility of SUV harmonization among these SPECT/CT using DRO17 mm as a reference standard.

Methods

Determination of Tc-99m concentration in NEMA phantom to simulate clinical Tc-99m SPECT/CT

In order to determine Tc-99m concentration enclosed in the NEMA body phantom, the following procedure was performed; first, SPECT/CT scans using an integrated SPECT/CT system (Discovery NM/CT 670pro, GE Healthcare) equipped with a low-energy high-resolution collimator were performed in 28 cancer patients 3 h after intravenous injection of 740 MBq of Tc-99m hydroxymethylene diphosphonate (Tc-99m HMDP) at one of the institutions participating in the present study. The SPECT data obtained from routine clinical examinations were used in order to determine Tc-99m concentration in the NEMA phantom, which was approved by the Institutional Review Board (IRB) in the hospital. The IRB officially granted permission for this retrospective review of the imaging data and waived the need for obtaining informed consent from the patients. SPECT counts of the lower abdominal portion were measured in order to obtain the reference counting rates (11.2 ± 3.5 kilo counts per second (kcps)) for the phantom study. Second, the body phantom in which six hot spheres (10, 13, 17, 22, 28, and 37 mm) were embedded was filled with Tc-99m solution so that the spheres had a 4:1 radioactivity ratio compared with the background. At the beginning of the SPECT scan, the radioactivity concentration and the SPECT counting rate of the phantom were 36 kBq/cc and 22.8 kcps, respectively (Fig. 1). Then, 6-min SPECT/CT scans were performed repeatedly with an interval of 60 min for 12 h. Based on the results of the correlation between the radioactivity concentration and the SPECT counting rate, the optimal radioactivity concentration for further phantom studies were determined.
Fig. 1

Correlation of Tc-99m concentration of the NEMA phantom and counting rate in Discovery NM/CT 670 from the beginning (A) to the end (B) of the SPECT scan. There was a linear relationship between the radioactivity and counting rate. Tc-99m concentration for further phantom study was determined based on the counting rate of clinical bone SPECT/CT

Correlation of Tc-99m concentration of the NEMA phantom and counting rate in Discovery NM/CT 670 from the beginning (A) to the end (B) of the SPECT scan. There was a linear relationship between the radioactivity and counting rate. Tc-99m concentration for further phantom study was determined based on the counting rate of clinical bone SPECT/CT

Calculation of calibration factor for SUV measurement using a cylindrical phantom

A cylindrical phantom with a diameter of 160 mm and a height of 150 mm (3016 mL) filled with Tc-99m solution of known activity concentration (approximately 25 MBq) was scanned for 6 min. Data were reconstructed with 3D-OSEM with scatter and CT-based attenuation correction and were processed with various parameter settings including the pre-harmonized conditions used in each of the four SPECT/CT cameras. Basic performances of the SPECT/CT cameras were describe elsewhere [1], and the detailed imaging conditions and collimator configurations are shown in Table 1. Parameter settings are comprised of the product of subset and iteration numbers (SI, range 40–160) and 3DGF (range 1.0–4.0 pixel (Infinia and Discovery; pixel size, 4.4 mm) or 5–17 mm (Brightview and Symbia)). The processing of 3DGF in Philips Brightview XCT was performed using a commercially available software GI-PET (AZE Co., Ltd., Tokyo, Japan) because this filter option was not installed in any imaging workstation belonging to the institution with Philips Brightview XCT. Resolution recovery (RR) by compensating the distance-dependent detector response was used.
Table 1

Imaging conditions and collimator configurations regarding the four state-of-the-art SPECT/CT systems

Brightview XCTDiscovery NM/CT 670Infinia Hawkeye 4Symbia T6
Imaging condition
 Step and shoot image acquisition
  No. of step30303030
  Rotation angle6666
  No. of projection60606060
  Scan orbitbody contourbody contourbody contourbody contour
 Size for image acquisition
  Matrix (x, y, z)132, 132, 132128, 128, 128128, 128, 128128, 128, 128
  Pixel size and slice thickness (mm)4.74.44.44.8
 Smoothing filter3D Gaussian3D Gaussian3D Gaussian3D Gaussian
 Reconstructed image for ROI analysis
  Matrix (x, y, z)256, 256, 203256, 256, 207256, 256, 130256, 256, 187
  Pixel size and slice thickness (mm)2.02.02.22.0
 Energy window
  Main140.5 keV ± 10%140.5 keV ± 10%140 keV ± 10%140 keV ± 10%
  SubN.A.120 keV ± 5%120 keV ± 5%120 keV ± 5%
 Attenuation correctionCT-basedCT-basedCT-basedCT-based
 Scatter correctionESSEDEWDEWDEW
Collimator
 TypeCHRLEHRLEHRLEHR
 No. of holes (thousand)40.286.386.3148
 Hole shapeHexagonHexagonHexagonHexagon
 Hole length (mm)48353524.1
 Septal thickness (mm)0.150.200.200.16
 Hole diameter across the flats (mm)2.031.501.501.11

ESSE effective source scatter estimation method. DEW dual-energy window method, CHR cardiac high-resolution collimator, LEHR low-energy high-resolution collimator, N.A. not applicable

Imaging conditions and collimator configurations regarding the four state-of-the-art SPECT/CT systems ESSE effective source scatter estimation method. DEW dual-energy window method, CHR cardiac high-resolution collimator, LEHR low-energy high-resolution collimator, N.A. not applicable SPECT/CT data in each reconstruction condition were analyzed using a commercially available software GI-BONE (AZE Co., Ltd., Tokyo, Japan). With the software, slice thickness was automatically converted to be about 2 mm to allow isotropic voxel evaluation (Table 1). A circular ROI was drawn on the center of the cylindrical phantom in the central slice as well as in slices ±1 and ±2 cm away, measuring SPECT count density (count/cc). The calibration factor was calculated as the ratio of actual radioactivity concentration (as measured by the dose calibrator) in the phantom at the time of scanning (ACC) to the measured SPECT count density per scan duration (MC), and we call this factor Becquerel calibration factor (BCF). Consequently, the BCF is calculated as: The BCF should be dependent on the performance of SPECT/CT system and imaging conditions. The MC also should be affected by a scaling factor (multiplying pixel count in reconstruction with RR) in GE resolution modeling.

SUV conversion of NEMA body phantom image using BCF

In order to simulate clinical Tc-99m SPECT/CT scans, the activity concentration levels in the background and spheres in the NEMA body phantom were set at 18 and 54 kBq/cc, respectively (Fig. 1). The phantom was scanned for 6 min with the four different SPECT/CT systems. The phantom images were reconstructed in the same parameter setting as BCF images. The phantom data in each reconstruction condition were analyzed using the same software as BCF data. Six different target ROIs, whose diameters were equal to the physical inner diameters of the hot spheres, were placed on the target slice. The SUV is calculated as: In this phantom study, the reciprocal of body weight per injected dose was 9000 Bq/g so that background SUV was 1. Regarding calculation of SUVs, 10-, 13-, 17-, 22-, 28-, and 37-mm circular ROIs were drawn exactly on the corresponding spheres in the central slice by following the CT boundaries of the fused SPECT/CT images. Then, SUVpeak, SUVmax, and the mean SUV (SUVmean) for the spheres were measured. Peak SUV represents the average SUV obtained within a 1-cc sphere of region of interest (ROI) centered on a highest voxel of the target area.

Harmonization of SUVs using DRO17 mm

Simulated images of original DRO and DRO17 mm are shown in Fig. 2. SUVmean, SUVpeak, and SUVmax of the six spheres in DRO17 mm are described in Table 2. As a measure of harmonization of reconstruction conditions, the root mean square error (RMSE) was measured; RMSE is the square root of the variance in SUVmean of the six spheres between DRO17 mm and actual phantom images obtained with the SPECT/CT cameras. Therefore, RMSE is measured as:
Fig. 2

a Phantom configuration of a digital reference object (DRO) and b DRO filtered by a 17-mm Gaussian filter (DRO17 mm). The square boxes in DRO are by design (not used in the present study)

Table 2

SUV values derived from a digital reference object smoothed by a 17-mm Gaussian filter (DRO17 mm)

Quantitative metrics
Quantitative metrics
SUVmean SUVpeak SUVmax
Sphere diameter (mm)
101.181.171.20
131.361.381.42
171.671.771.86
222.122.402.53
282.553.143.29
372.943.823.91
a Phantom configuration of a digital reference object (DRO) and b DRO filtered by a 17-mm Gaussian filter (DRO17 mm). The square boxes in DRO are by design (not used in the present study) SUV values derived from a digital reference object smoothed by a 17-mm Gaussian filter (DRO17 mm) Reconstruction conditions according to the manufacturers’ recommendation or the policies of their own departments are shown in Table 3. SUVmean of the hot spheres and RMSE in the pre-harmonized conditions are also shown in Table 3.
Table 3

Reconstruction conditions according to the manufacturers’ recommendation or the policies of their own departments

SPECT/CT scanner
BrightViewDiscoveryInfiniaSymbia
Reconstruction parameter
Subset8101010
Iteration10101010
FilterGaussianGaussianGaussianGaussian
Cutoff value15 mm2.5 pixel2.5 pixel9 mm
Resolution recoveryAstonishEvolutionEvolutionFLASH 3D
SUVmean of the spheres
10 mm1.081.250.981.06
13 mm1.141.481.281.37
17 mm1.381.731.731.99
22 mm1.812.191.922.52
28 mm2.412.962.423.08
37 mm2.863.212.713.50
RMSE0.200.200.130.40
Reconstruction conditions according to the manufacturers’ recommendation or the policies of their own departments RMSE was calculated in the following conditions: SI, range 40–140; 3DGF, range 1.0–4.0 pixel (Infinia and Discovery) or 5–17 mm (Brightview and Symbia). Then, for every scanner examined, settings were found that showed a clear optimum for harmonization. In both pre- and post-harmonized conditions, coefficient of variation (COV) of SUVmean, SUVmax, and SUVpeak between the four scanners were calculated.

Results

Tc-99m concentration for phantom study

As shown in Fig. 1, Tc-99m concentration in the NEMA body phantom had linear correlation with the acquisition counting rate. The counting rates in human bone SPECT/CT (11.2 ± 3.5 kcps) were equivalent to Tc-99m concentration of 12.8–22.9 kBq/ml. Therefore, the activity concentration levels in the background and spheres for further evaluation were set at 18 and 54 kBq/cc, resulting in mean activity concentration of the entire phantom of approximately 18.3 kBq/cc.

BCF measurement

Table 4 shows the distribution of BCF among SPECT/CT systems with different reconstruction conditions. The difference in BCF value was small between BrightView and Symbia and between Infinia and Discovery. A scaling factor seemed to affect the BCF. Reconstruction conditions did not significantly affect the BCF (approximately less than 3% of mean value).
Table 4

Distribution of BCF among SPECT/CT systems with different reconstruction conditions

SPECT/CT scanner
BrightViewDiscoverya Infiniaa Symbia
Mean5309161715384914
Standard deviation3554832
Relative standard deviation (%)0.70.33.10.7

aA scaling factor is involved with the values

Distribution of BCF among SPECT/CT systems with different reconstruction conditions aA scaling factor is involved with the values

Effects of reconstruction parameter settings on SUVs and RMSE

Figure 3 shows RMSE for the four SPECT/CT systems. RMSE was lowest in the following reconstruction conditions: SI of 100 and 3DGF of 13 mm for Brightview XCT (RMSE = 0.115); SI of 160 and 3DGF of 3 pixels for Discovery NM/CT (RMSE = 0.085); SI of 60 and 3DGF of 2 pixels for Infinia (RMSE = 0.102); and SI of 140 and 3DGF of 15 mm for Symbia (RMSE = 0.117). It should be noted that the minimum RMSE was below 0.12 for each harmonized setting.
Fig. 3

The root mean square error (RMSE) in SUVmean of the six spheres between DRO17 mm and actual phantom images obtained with the SPECT/CT cameras

The root mean square error (RMSE) in SUVmean of the six spheres between DRO17 mm and actual phantom images obtained with the SPECT/CT cameras

SUVmean, SUVmax, and SUVpeak in both pre- and post-harmonized conditions

Figure 4 shows SUVmean, SUVmax, and SUVpeak of the hot spheres in both pre- and post-harmonized conditions. Table 5 shows COVs of these metrics between the four SPECT/CT systems. In pre-harmonized conditions, COVs were greater than 10% for all metrics in the 17-, 22-, and 28-mm spheres, SUVmax in the 13-mm sphere and SUVmean in the 37-mm sphere. In contrast, all metrics except SUVmax in the 17-mm sphere yielded less than 10% of COVs after harmonization.
Fig. 4

Inter-scanner variability in standardized uptake values (SUVs) a–c before and d–f after harmonization

Table 5

Coefficient of variations (COVs) of SUVs between the four SPECT/CT systems

Pre-harmonizationPost-harmonization
SpheresSUVmean SUVpeak SUVmax SUVmean SUVpeak SUVmax
10 mm9.008.239.817.476.427.48
13 mm9.469.3811.575.515.177.54
17 mm12.7012.5214.907.057.6610.77
22 mm12.9915.2017.302.241.002.94
28 mm11.2411.6312.932.114.856.56
37 mm10.025.826.212.671.552.51
Inter-scanner variability in standardized uptake values (SUVs) a–c before and d–f after harmonization Coefficient of variations (COVs) of SUVs between the four SPECT/CT systems

Discussion

Recent advances in SPECT/CT technologies allowed major manufacturers to mass-produce commercial SPECT/CT systems for clinical application of not only SPECT/CT fusion imaging but also fully quantitative SPECT imaging. Although PET/CT has become an important diagnostic tool quantifying tracer uptake, only a small number of PET tracers have yet been approved in clinical practice. In contrast, there have already been various kinds of available radiopharmaceuticals labeled with single-photon emitters; much focus is being placed on the value of quantitative SPECT/CT [4, 5, 13–16]. Especially, clinical application of quantitative SPECT/CT using bone-seeking radiotracers is highly expected as shown in a successful report on the use of F-18 fluorine PET for prognostic assessment [17] as well as the accumulated evidences of quantitative planar bone scintigraphy in prostate cancer [18]. The bone scan index (BSI) [19, 20], which quantifies the total bone metastatic burden relative to the total skeletal mass on two-dimensional images, is getting wider acceptance as a biomarker for predicting survival in patients with prostate cancer [21-26]. However, there are substantial false-positive and false-negative findings when evaluating bone metastasis without SPECT/CT [27-31]. In addition, quantifying tracer accumulation on a per-lesion basis is limited by the projection of several overlying structures in a planar image. For instance, uptake in the sternum may contain some amounts of uptake in the thoracic spine in an anterior view of planar image, and quantitative analysis would therefore be difficult especially when metastasis occurs in these bones. We envisaged that harmonizing SUVs using the DRO could be applied to multicenter clinical trials using Tc-99m SPECT/CT; in particular, harmonized SUVs in bone SPECT/CT may become an alternative choice to BSI. In addition, the harmonization method might be utilized to reduce inter-scanner variability in measurement of SPECT/CT-derived absorbed doses in a variety of “theranostics” situations such as Tc-99m MAA SPECT/CT in Y-90 microsphere therapy, I-131 SPECT/CT in thyroid cancer therapy, and Lu-177-PSMA in prostate cancer therapy [32-34]. We used DRO for harmonization instead of a two-step approach of assessing how close each scanner can get to true SUV (i.e., SUV = 4) and then harmonizing to the lower common denominator based on the following reasons: We found that both SUVmax and SUVpeak fluctuated when acquisition time or phantom radioactivity was changed, probably due to image noise. In contrast, SUVmean did not (data not shown). Therefore, we thought that SUVmax and SUVpeak are not suitable parameters for harmonization in terms of test-retest reproducibility. Although SUVmean may be used for harmonization because of being unsusceptible to image noise, it never reached the uptake value of 4 even for the largest sphere (37 mm) due to partial volume effect. Hence, it seems impractical to assess how close each scanner can reach SUVmean of 4. Instead, DRO was smoothed to match SUVs of the targets in each of SPECT/CT systems with the corresponding SUVs in DRO. Presetting DRO as a reference allows a variety of institutions to easily join the harmonization projects without any revisions of previously harmonized protocols in enrolled institutions, whereas the two-step approach seems complex when many scanners need to be harmonized. It is important to know how accurate the current SPECT/CT technologies can be in terms of quantitation. It goes without saying, however, that even the state-of-the-art SPECT/CT systems are less reliable than general PET/CT systems especially in quantifying small lesions due to limited detector sensitivity and intrinsic spatial resolution; not surprisingly, a 10-mm hot sphere was undetected with any of the four SPECT/CT systems under all reconstruction conditions, and a 13-mm hot sphere was barely discernible in most of the reconstructed SPECT images (data not shown). Based on the fact that point spread function or line spread function of the SPECT detectors, which represents image blurring due to finite spatial resolution, can be geometrically approximated by Gaussian function; DRO10 mm or DRO13 mm was considered unsuitable to serve as a reference image. In contrast, the actual hot spheres measuring at least 17 mm were clearly observed irrespective of reconstruction conditions and SPECT/CT systems. Therefore, DRO17 mm was chosen as a reference image in the present study. 3DGF was used throughout the harmonization instead of using another filter such as Butterworth and Hanning filters. This is not only because DRO17 mm was generated with Gaussian filter, but because measurement of BCF with Gaussian filter was more stable than that with Butterworth or Hanning filter when changing reconstruction parameters such as SI and cutoff value of these filters (data not shown). We found that the lowest RMSE value was obtained with 3DGF of 8.8 to 15 mm and SI of 60 or 160 (Fig. 3). The results were partially different from the European Association of Nuclear Medicine (EANM) practice guidelines and recommendations of the camera manufacturers [35], which indicates a need for harmonization-specific imaging protocol. As shown in Fig. 4 and Table 5, inter-scanner variability in SUVs among the state-of-the-art SPECT/CT systems was clearly decreased after the proposed harmonization procedure. In this context, we propose the use of a digital phantom developed by QIBA for harmonizing SUVs in multicenter trials. In PET/CT, the EANM guidelines do not positively recommend the use of resolution recovery for quantitative assessment in multicenter studies due to Gibbs artifact [36]. We observed small amount of uptake biased to a peripheral side of the sphere in the larger spheres (e.g., 37-mm sphere), probably due to Gibbs phenomenon. This effect possibly resulted in a slight elevation of SUVmax over true value (i.e., SUV = 4) as shown in Fig. 4. Although resolution recovery was responsible for the overshoot, the lack of resolution recovery significantly underestimated SUVs. For instance, SUVmax and SUVmean of a 37-mm sphere without resolution recovery were about 3.3 and 2.5, respectively. In light of the principle of photon detection with collimator-dependent SPECT systems, it is reasonable to compensate the distance-dependent detector response for lesion-based quantitative assessment. At present, we consider that resolution recovery should be used at the sacrifice of the small amount of the overshoot. In the present study, the radioactivity of the NEMA phantom was determined on the basis of the bone SPECT data under a 6-min acquisition protocol. According to the Japanese technological guidelines on nuclear imaging, bone SPECT data should be collected for 5–6 min/bed [37]. On the other hand, the EANM practice guidelines indicate the acquisition time of 10–30 min/bed [35]. The difference may be due to the fact that radioactive dose administered to patients undergoing bone scintigraphy is different between Japan and Europe (mean dose, 740 vs 500 MBq). In addition, considering the difference in body weight and height between Japanese (light and short) and European people (heavy and tall), 6 min SPECT acquisition for Japanese patients would be equivalent to 10 min or more acquisition for European patients in terms of SPECT counts per bed position. There are several limitations in this study. First, DRO17 mm has no absolute and universal significance as a reference image. It does not seem to be necessary to smooth to a level which matches the sphere size. In other words, DRO14 mm, DRO15 mm, or DRO16 mm might serve as better references. In our preliminary study, DRO17 mm was arbitrarily determined as a reference. However, it is worth noting that the minimum RMSE was below 0.12 for each harmonized setting (Fig. 3) and that the harmonized SUV curves as a function of sphere size are close to the curve of DRO17 mm as shown in Fig. 4. Hence, we considered that DRO17 mm could be a suboptimal reference for a multicenter study and that DRO14 mm, DRO15 mm, and DRO16 mm may also be references for another multicenter study. In any case, it seems important to specify which DRO is regarded as a reference together with RMSE for each scanner. Our results suggest that RMSE of 0.12 may serve as an index of appropriateness of harmonization. Second, background and cold regions were not focused on in this study. This study was intended for a variety of multicenter SPECT studies such as bone SPECT/CT in which quantitation of background or cold regions would be unnecessary. In other words, the results of our study should not be applied to myocardial or cerebral perfusion SPECT/CT in which decrease in tracer uptake has significant impact on treatment strategy. In this context, we believe that our study is the first step to expand the use of DRO by QIBA for the future of quantitation using SPECT/CT. Finally, the currently available DRO has a contrast of 4:1. Therefore, we collected SPECT data of the NEMA phantom with 4:1 concentration ratio. Whether the results would be applicable for other contrast remains unknown. Examining this issue is one of the top priorities for further research.

Conclusions

In the present study, the DRO smoothed by 3DGF with a FWHM of 17 mm was used for the purpose of harmonizing SUVs in Tc-99m SPECT/CT imaging. SUVs generated according to the manufacturers’ recommendation or the policies of their own departments had substantial inter-scanner variability, indicating a need for harmonization-specific imaging protocols. Our harmonization clearly reduced inter-scanner variability in all metrics except SUVmax in the 17-mm sphere with less than 10% of COVs. A digital phantom developed by QIBA would be useful for harmonizing SUVs in multicenter trials.
  33 in total

1.  Repeatability of Radiotracer Uptake in Normal Abdominal Organs with ¹¹¹In-Pentetreotide Quantitative SPECT/CT.

Authors:  Steven P Rowe; Esther Vicente; Nadège Anizan; Hao Wang; Jeffrey P Leal; Martin A Lodge; Eric C Frey; Richard L Wahl
Journal:  J Nucl Med       Date:  2015-05-14       Impact factor: 10.057

Review 2.  Scoring systems of quantitative bone scanning in prostate cancer: historical overview, current status and future perspectives.

Authors:  Athanasios Zafeirakis
Journal:  Hell J Nucl Med       Date:  2014-07-05       Impact factor: 1.102

3.  A Digital Reference Object to Analyze Calculation Accuracy of PET Standardized Uptake Value.

Authors:  Larry A Pierce; Brian F Elston; David A Clunie; Dennis Nelson; Paul E Kinahan
Journal:  Radiology       Date:  2015-05-19       Impact factor: 11.105

4.  Assessment of the bone scan index in a randomized placebo-controlled trial of tasquinimod in men with metastatic castration-resistant prostate cancer (mCRPC).

Authors:  Andrew J Armstrong; Reza Kaboteh; Michael A Carducci; Jan-Erik Damber; Walter M Stadler; Mats Hansen; Lars Edenbrandt; Göran Forsberg; Örjan Nordle; Roberto Pili; Michael J Morris
Journal:  Urol Oncol       Date:  2014-09-16       Impact factor: 3.498

5.  Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patients.

Authors:  Charline Lasnon; Cédric Desmonts; Elske Quak; Radj Gervais; Pascal Do; Catherine Dubos-Arvis; Nicolas Aide
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-04-06       Impact factor: 9.236

6.  Whole-body SPECT/CT for bone scintigraphy: diagnostic value and effect on patient management in oncological patients.

Authors:  H Palmedo; C Marx; A Ebert; B Kreft; Y Ko; A Türler; R Vorreuther; U Göhring; H H Schild; T Gerhardt; U Pöge; S Ezziddin; H-J Biersack; H Ahmadzadehfar
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-08-24       Impact factor: 9.236

7.  Bone Scan Index as a prognostic imaging biomarker during androgen deprivation therapy.

Authors:  Mariana Reza; Anders Bjartell; Mattias Ohlsson; Reza Kaboteh; Per Wollmer; Lars Edenbrandt; Elin Trägårdh
Journal:  EJNMMI Res       Date:  2014-10-17       Impact factor: 3.138

8.  3D SPECT/CT fusion using image data projection of bone SPECT onto 3D volume-rendered CT images: feasibility and clinical impact in the diagnosis of bone metastasis.

Authors:  Yuji Ogata; Tadaki Nakahara; Kenichi Ode; Yohji Matsusaka; Mari Katagiri; Yu Iwabuchi; Kazunari Itoh; Akira Ichimura; Masahiro Jinzaki
Journal:  Ann Nucl Med       Date:  2017-02-27       Impact factor: 2.668

9.  Quantitative capabilities of four state-of-the-art SPECT-CT cameras.

Authors:  Alain Seret; Daniel Nguyen; Claire Bernard
Journal:  EJNMMI Res       Date:  2012-08-27       Impact factor: 3.138

10.  The EANM practice guidelines for bone scintigraphy.

Authors:  T Van den Wyngaert; K Strobel; W U Kampen; T Kuwert; W van der Bruggen; H K Mohan; G Gnanasegaran; R Delgado-Bolton; W A Weber; M Beheshti; W Langsteger; F Giammarile; F M Mottaghy; F Paycha
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-06-04       Impact factor: 9.236

View more
  12 in total

1.  Development of a new quantification method using partial volume effect correction for individual energy peaks in 111In-pentetreotide SPECT/CT.

Authors:  Kosuke Yamashita; Noriaki Miyaji; Kazuki Motegi; Takashi Terauchi; Shigeki Ito
Journal:  Asia Ocean J Nucl Med Biol       Date:  2022

2.  Assessment of inflammatory jaw pathologies using bone SPECT/CT maximum standardized uptake value.

Authors:  Hiroo Toshima; Ichiro Ogura
Journal:  Dentomaxillofac Radiol       Date:  2020-06-15       Impact factor: 2.419

3.  Validation of the influence of CT slice thickness on the quantitative accuracy and image quality of single photon emission computed tomography.

Authors:  Tomohiro Sato; Takashi Takagi
Journal:  Asia Ocean J Nucl Med Biol       Date:  2021

4.  Analysis of medication-related osteonecrosis of the jaw with bone SPECT/CT: relationship between patient characteristics and maximum standardized uptake value.

Authors:  Ruri Ogawa; Ichiro Ogura
Journal:  Dentomaxillofac Radiol       Date:  2021-05-21       Impact factor: 2.419

5.  Letter to the Editor.

Authors:  Tadaki Nakahara
Journal:  Asia Ocean J Nucl Med Biol       Date:  2018

6.  Phantom and clinical evaluation of bone SPECT/CT image reconstruction with xSPECT algorithm.

Authors:  Noriaki Miyaji; Kenta Miwa; Ayaka Tokiwa; Hajime Ichikawa; Takashi Terauchi; Mitsuru Koizumi; Masahisa Onoguchi
Journal:  EJNMMI Res       Date:  2020-06-29       Impact factor: 3.138

7.  Multicentre analysis of PET SUV using vendor-neutral software: the Japanese Harmonization Technology (J-Hart) study.

Authors:  Yuji Tsutsui; Hiromitsu Daisaki; Go Akamatsu; Takuro Umeda; Matsuyoshi Ogawa; Hironori Kajiwara; Shigeto Kawase; Minoru Sakurai; Hiroyuki Nishida; Keiichi Magota; Kazuaki Mori; Masayuki Sasaki
Journal:  EJNMMI Res       Date:  2018-08-20       Impact factor: 3.138

8.  Towards standardization of absolute SPECT/CT quantification: a multi-center and multi-vendor phantom study.

Authors:  Steffie M B Peters; Niels R van der Werf; Marcel Segbers; Floris H P van Velden; Roel Wierts; Koos J A K Blokland; Mark W Konijnenberg; Sergiy V Lazarenko; Eric P Visser; Martin Gotthardt
Journal:  EJNMMI Phys       Date:  2019-12-26

9.  Quantitative bone single-photon emission computed tomography imaging for uninfected nonunion: comparison of hypertrophic nonunion and non-hypertrophic nonunion.

Authors:  Keisuke Oe; Feibi Zeng; Tomoaki Fukui; Munenobu Nogami; Takamichi Murakami; Tomoyuki Matsumoto; Ryosuke Kuroda; Takahiro Niikura
Journal:  J Orthop Surg Res       Date:  2021-02-10       Impact factor: 2.359

Review 10.  Does quantification have a role to play in the future of bone SPECT?

Authors:  James C Ross; Dijana Vilić; Tom Sanderson; Stefan Vöö; John Dickson
Journal:  Eur J Hybrid Imaging       Date:  2019-05-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.