Literature DB >> 22415635

A multivendor phantom study comparing the image quality produced from three state-of-the-art SPECT-CT systems.

Tyler Hughes1, Anna Celler.   

Abstract

OBJECTIVE: Ongoing advancements in single photon emission computed tomography with on-board X-ray computed tomography (SPECT-CT) hardware and software raise important questions regarding the relative performances of various cameras and their respective image-processing software. This phantom-based study compares images produced from three state-of-the-art cameras using four image quality measurements.
METHODS: A thorax phantom modeling the spine, lungs, a healthy heart, and three tumors (cylindrical bottles) was scanned using the following SPECT-CT systems: Philips' Precedence (PP), GE's Infinia-Hawkeye (GH), and Siemens' Symbia-T6 (SS). For each scan, Tc-99m solutions were injected into the heart, three bottles, and thorax to yield activity concentration ratios of roughly 6:1 for both heart:thorax and tumor:thorax. The data were reconstructed using the most advanced software available on the cameras, namely, Evolution for Bone and Evolution for Cardiac (EVB and EVC, respectively), Astonish (AST), and Flash3D (FLA) for GH, PP, and SS, respectively. In addition, all sets of data were reconstructed using our in-house software. The mean values of activity error, uniformity, signal-to-noise ratio, and contrast error were used as figures of merit (FOM).
RESULTS: No significant differences were observed for all FOM between all in-house reconstructions using PP, GH, and SS acquisition data. The mean activity error for the AST reconstruction (-24.0±1.6%) was significantly closer to the truth relative to EVB (-38.0±1.6%), EVC (-34.5±2.3%), and FLA (-33.8±1.6%). No significant differences were found between EVC and FLA for all FOM.
CONCLUSION: In this phantom-based study, Philips' AST provided the most quantitatively accurate and highest contrast images, whereas Siemens' FLA and GE's EVC provided relatively higher signal-to-noise ratios and more uniform images.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22415635     DOI: 10.1097/MNM.0b013e328351d549

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  6 in total

1.  Heterogeneity of intratumoral (111)In-ibritumomab tiuxetan and (18)F-FDG distribution in association with therapeutic response in radioimmunotherapy for B-cell non-Hodgkin's lymphoma.

Authors:  Kohei Hanaoka; Makoto Hosono; Yoichi Tatsumi; Kazunari Ishii; Sung-Woon Im; Norio Tsuchiya; Kenta Sakaguchi; Itaru Matsumura
Journal:  EJNMMI Res       Date:  2015-03-14       Impact factor: 3.138

2.  Performance of Myocardial Perfusion Imaging Using Multi-focus Fan Beam Collimator with Resolution Recovery Reconstruction in a Comparison with Conventional SPECT.

Authors:  Norikazu Matsutomo; Akio Nagaki; Masayuki Sasaki
Journal:  Asia Ocean J Nucl Med Biol       Date:  2014

3.  Use of a digital phantom developed by QIBA for harmonizing SUVs obtained from the state-of-the-art SPECT/CT systems: a multicenter study.

Authors:  Tadaki Nakahara; Hiromitsu Daisaki; Yasushi Yamamoto; Takashi Iimori; Kazuyuki Miyagawa; Tomoya Okamoto; Yoshiki Owaki; Nobuhiro Yada; Koichi Sawada; Ryotaro Tokorodani; Masahiro Jinzaki
Journal:  EJNMMI Res       Date:  2017-06-20       Impact factor: 3.138

4.  Towards standardization of absolute SPECT/CT quantification: a multi-center and multi-vendor phantom study.

Authors:  Steffie M B Peters; Niels R van der Werf; Marcel Segbers; Floris H P van Velden; Roel Wierts; Koos J A K Blokland; Mark W Konijnenberg; Sergiy V Lazarenko; Eric P Visser; Martin Gotthardt
Journal:  EJNMMI Phys       Date:  2019-12-26

5.  Impact of Reduced Acquisition Time on Bone Single-photon Emission Computed Tomography Images in Oncology Patients.

Authors:  Nusret Salkica; Amela Begic; Sandra Zubovic; Sejla Ceric; Amila Basic; Adnan Sehic; Fuad Julardzija; Enis Tinjak
Journal:  Acta Inform Med       Date:  2022-03

6.  Quantitative capabilities of four state-of-the-art SPECT-CT cameras.

Authors:  Alain Seret; Daniel Nguyen; Claire Bernard
Journal:  EJNMMI Res       Date:  2012-08-27       Impact factor: 3.138

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.