Validation of the influence of CT slice thickness on the quantitative accuracy and image quality of single photon emission computed tomography.

Tomohiro Sato1, Takashi Takagi1.   

Abstract

OBJECTIVES: Computed tomography (CT) images are used for precise anatomical location of lesions and for accurate attenuation correction in single-photon emission computed tomography (SPECT) image reconstruction in SPECT/CT examination. The aim of this study was to verify the effects of varying CT collimation width and slice thickness on CT images and on CT attenuation corrected SPECT images.
METHODS: We acquired SPECT/CT images of a micro-coin phantom and the National Electrical Manufacturers Association body phantom filled with 99mTc-pertechnetate while varying the abovementioned CT parameters. The full width at half maximum of the slice sensitivity profile, the standard deviation of CT image background noise, and the radiation dose from CT scans were evaluated. Subsequently, the percentage contrast, background variability, and absolute recovery coefficient of the SPECT image were measured. Furthermore, we retrospectively reviewed the clinical bone SPECT images of 23 patients, and statistical testing of differences was performed.
RESULTS: As the collimation width and reconstruction slice thickness of the CT image increased, z-axis resolution deteriorated, and background noise decreased. In addition, CT radiation dose decreased with increasing collimation width. Meanwhile, SPECT image quality and quantitative accuracy were unchanged with varying CT collimation width and slice thickness. There were no notable variations in clinical SPECT images and no statistically significant differences.
CONCLUSION: When high-resolution CT slices on the z-axis are not required for clinical diagnosis, increasing collimation width or slice thickness can reduce the radiation dose and image noise with no influence on the quality of SPECT images .
© 2021 mums.ac.ir All rights reserved.

Entities:  

Keywords:  CT slice thickness; Image quality; OSCGM; Radiation dose; SPECT/CT

Year:  2021        PMID: 34250143      PMCID: PMC8255530          DOI: 10.22038/AOJNMB.2021.54670.1376

Source DB:  PubMed          Journal:  Asia Ocean J Nucl Med Biol        ISSN: 2322-5718


  14 in total

Review 1.  The use of standardized uptake values for assessing FDG uptake with PET in oncology: a clinical perspective.

Authors:  G Lucignani; G Paganelli; E Bombardieri
Journal:  Nucl Med Commun       Date:  2004-07       Impact factor: 1.690

2.  Ramp test objects for slice sensitivity profile measurement in spiral CT.

Authors:  A Mackie; G C Hart; J F Williams-Butt
Journal:  Br J Radiol       Date:  1997-09       Impact factor: 3.039

Review 3.  The Clinical Dilemma of Incidental Findings on the Low-Resolution CT Images from SPECT/CT MPI Studies.

Authors:  Joanne Coward; Julie Nightingale; Peter Hogg
Journal:  J Nucl Med Technol       Date:  2016-04-21

4.  Influence of the Pixel Sizes of Reference Computed Tomography on Single-photon Emission Computed Tomography Image Reconstruction Using Conjugate-gradient Algorithm.

Authors:  Kyohei Okuda; Shota Sakimoto; Susumu Fujii; Tomonobu Ida; Shigeru Moriyama
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2017

5.  Single-photon emission tomographic quantification in spherical objects: effects of object size and background.

Authors:  F Zito; M C Gilardi; P Magnani; F Fazio
Journal:  Eur J Nucl Med       Date:  1996-03

6.  The effect of x-ray beam quality and geometry on radiation utilization efficiency in multidetector CT imaging.

Authors:  Kostas Perisinakis; Antonios E Papadakis; John Damilakis
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

7.  PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients.

Authors:  C Burger; G Goerres; S Schoenes; A Buck; A H R Lonn; G K Von Schulthess
Journal:  Eur J Nucl Med Mol Imaging       Date:  2002-04-19       Impact factor: 9.236

8.  SPECT/CT Fusion in the Diagnosis of Hyperparathyroidism.

Authors:  Yoshio Monzen; Akihisa Tamura; Hajime Okazaki; Taichi Kurose; Masayuki Kobayashi; Masatsugu Kuraoka
Journal:  Asia Ocean J Nucl Med Biol       Date:  2015

9.  Use of a digital phantom developed by QIBA for harmonizing SUVs obtained from the state-of-the-art SPECT/CT systems: a multicenter study.

Authors:  Tadaki Nakahara; Hiromitsu Daisaki; Yasushi Yamamoto; Takashi Iimori; Kazuyuki Miyagawa; Tomoya Okamoto; Yoshiki Owaki; Nobuhiro Yada; Koichi Sawada; Ryotaro Tokorodani; Masahiro Jinzaki
Journal:  EJNMMI Res       Date:  2017-06-20       Impact factor: 3.138

10.  Impact of Novel Incorporation of CT-based Segment Mapping into a Conjugated Gradient Algorithm on Bone SPECT Imaging: Fundamental Characteristics of a Context-specific Reconstruction Method.

Authors:  Kyohei Okuda; Susumu Fujii; Shota Sakimoto
Journal:  Asia Ocean J Nucl Med Biol       Date:  2019
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.