Statistical characterization of noise for spatial standardization of CT scans: Enabling comparison with multiple kernels and doses.

Gonzalo Vegas-Sánchez-Ferrero1, Maria J Ledesma-Carbayo2, George R Washko3, Raúl San José Estépar3.   

Abstract

Computerized tomography (CT) is a widely adopted modality for analyzing directly or indirectly functional, biological and morphological processes by means of the image characteristics. However, the potential utilization of the information obtained from CT images is often limited when considering the analysis of quantitative information involving different devices, acquisition protocols or reconstruction algorithms. Although CT scanners are calibrated as a part of the imaging workflow, the calibration is circumscribed to global reference values and does not circumvent problems that are inherent to the imaging modality. One of them is the lack of noise stationarity, which makes quantitative biomarkers extracted from the images less robust and stable. Some methodologies have been proposed for the assessment of non-stationary noise in reconstructed CT scans. However, those methods focused on the non-stationarity only due to the reconstruction geometry and are mainly based on the propagation of the variance of noise throughout the whole reconstruction process. Additionally, the philosophy followed in the state-of-the-art methods is based on the reduction of noise, but not in the standardization of it. This means that, even if the noise is reduced, the statistics of the signal remain non-stationary, which is insufficient to enable comparisons between different acquisitions with different statistical characteristics. In this work, we propose a statistical characterization of noise in reconstructed CT scans that leads to a versatile statistical model that effectively characterizes different doses, reconstruction kernels, and devices. The statistical model is generalized to deal with the partial volume effect via a localized mixture model that also describes the non-stationarity of noise. Finally, we propose a stabilization scheme to achieve stationary variance. The validation of the proposed methodology was performed with a physical phantom and clinical CT scans acquired with different configurations (kernels, doses, algorithms including iterative reconstruction). The results confirmed its suitability to enable comparisons with different doses, and acquisition protocols.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Computerized tomography; Non-stationary noise; Statistical characterization

Mesh:

Year:  2017        PMID: 28622587      PMCID: PMC5742643          DOI: 10.1016/j.media.2017.06.001

Source DB:  PubMed          Journal:  Med Image Anal        ISSN: 1361-8415            Impact factor:   8.545


  21 in total

1.  The noise power spectrum in CT with direct fan beam reconstruction.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

2.  Analytic noise-propagation in indirect fan-beam FBP reconstruction.

Authors:  Anja Borsdorf; Steffen Kappler; Rainer Raupach; Joachim Hornegger
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2008

3.  The use of transformations.

Authors:  M S BARTLETT
Journal:  Biometrics       Date:  1947-03       Impact factor: 2.571

4.  Quantification of coronary artery calcium using ultrafast computed tomography.

Authors:  A S Agatston; W R Janowitz; F J Hildner; N R Zusmer; M Viamonte; R Detrano
Journal:  J Am Coll Cardiol       Date:  1990-03-15       Impact factor: 24.094

5.  Quantitative imaging test approval and biomarker qualification: interrelated but distinct activities.

Authors:  Andrew J Buckler; Linda Bresolin; N Reed Dunnick; Daniel C Sullivan; Hugo J W L Aerts; Bernard Bendriem; Claus Bendtsen; Ronald Boellaard; John M Boone; Patricia E Cole; James J Conklin; Gary S Dorfman; Pamela S Douglas; Willy Eidsaunet; Cathy Elsinger; Richard A Frank; Constantine Gatsonis; Maryellen L Giger; Sandeep N Gupta; David Gustafson; Otto S Hoekstra; Edward F Jackson; Lisa Karam; Gary J Kelloff; Paul E Kinahan; Geoffrey McLennan; Colin G Miller; P David Mozley; Keith E Muller; Rick Patt; David Raunig; Mark Rosen; Haren Rupani; Lawrence H Schwartz; Barry A Siegel; A Gregory Sorensen; Richard L Wahl; John C Waterton; Walter Wolf; Gudrun Zahlmann; Brian Zimmerman
Journal:  Radiology       Date:  2011-02-15       Impact factor: 11.105

6.  Role of the Quantitative Imaging Biomarker Alliance in optimizing CT for the evaluation of lung cancer screen-detected nodules.

Authors:  James L Mulshine; David S Gierada; Samuel G Armato; Rick S Avila; David F Yankelevitz; Ella A Kazerooni; Michael F McNitt-Gray; Andrew J Buckler; Daniel C Sullivan
Journal:  J Am Coll Radiol       Date:  2015-04       Impact factor: 5.532

7.  Denoising of polychromatic CT images based on their own noise properties.

Authors:  Ji Hye Kim; Yongjin Chang; Jong Beom Ra
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

Review 8.  Impact of obesity on medical imaging and image-guided intervention.

Authors:  Raul N Uppot; Dushyant V Sahani; Peter F Hahn; Debra Gervais; Peter R Mueller
Journal:  AJR Am J Roentgenol       Date:  2007-02       Impact factor: 3.959

9.  The line spread function and modulation transfer function of a computed tomographic scanner.

Authors:  P F Judy
Journal:  Med Phys       Date:  1976 Jul-Aug       Impact factor: 4.071

10.  "Density mask". An objective method to quantitate emphysema using computed tomography.

Authors:  N L Müller; C A Staples; R R Miller; R T Abboud
Journal:  Chest       Date:  1988-10       Impact factor: 9.410

View more
  7 in total

1.  Automatic estimation of aortic and mitral valve displacements in dynamic CTA with 4D graph-cuts.

Authors:  Juan E Ortuño; Gonzalo Vegas-Sánchez-Ferrero; Juan J Gómez-Valverde; Marcus Y Chen; Andrés Santos; Elliot R McVeigh; María J Ledesma-Carbayo
Journal:  Med Image Anal       Date:  2020-06-06       Impact factor: 8.545

2.  A CT Scan Harmonization Technique to Detect Emphysema and Small Airway Diseases.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Raúl San Estépar José
Journal:  Image Anal Mov Organ Breast Thorac Images (2018)       Date:  2018-09-12

3.  Harmonization of chest CT scans for different doses and reconstruction methods.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Maria Jesus Ledesma-Carbayo; George R Washko; Raúl San José Estépar
Journal:  Med Phys       Date:  2019-06-07       Impact factor: 4.071

4.  Harmonization of in-plane resolution in CT using multiple reconstructions from single acquisitions.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Gabriel Ramos-Llordén; Raúl San José Estépar
Journal:  Med Phys       Date:  2021-09-14       Impact factor: 4.071

5.  Generative-based airway and vessel morphology quantification on chest CT images.

Authors:  Pietro Nardelli; James C Ross; Raúl San José Estépar
Journal:  Med Image Anal       Date:  2020-03-28       Impact factor: 8.545

6.  Autocalibration method for non-stationary CT bias correction.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Maria J Ledesma-Carbayo; George R Washko; Raúl San José Estépar
Journal:  Med Image Anal       Date:  2017-12-08       Impact factor: 8.545

7.  Statistical characterization of the linear attenuation coefficient in polychromatic CT scans.

Authors:  Gonzalo Vegas; Raúl San José Estépar
Journal:  Med Phys       Date:  2020-10-06       Impact factor: 4.071

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.