Statistical characterization of the linear attenuation coefficient in polychromatic CT scans.

Gonzalo Vegas1, Raúl San José Estépar1.   

Abstract

PURPOSE: To provide a unifying statistical model that characterizes the integrated x-ray intensity at the detector after logarithmic transformation and can be extended to the characterization of computed tomography (CT) numbers in the reconstructed image.
METHODS: We study the statistical characteristics of polyenergetic x-ray beams in the detector. Firstly, we consider the characterization of the integrated x-ray intensity at the detector through a probabilistic model (compound Poisson) that describes its statistics. We analyze its properties and derive the probabilistic distribution after the logarithmic transformation analytically. Finally, we propose a more tractable probabilistic distribution with the same features observed in the characterization, the noncentral Gamma (nc-Gamma). This distribution exhibits desirable properties for the statistical characterization across the reconstruction process. We assess the assumptions adopted in the derivation of the statistical models throughout Monte Carlo simulations and validate them with a water phantom and a lung phantom acquired in a Siemens clinical CT scan. We evaluate the statistical similarities between the theoretical distribution and the nc-Gamma using a power analysis with a Kolmogorov-Smirnov test for a 95% confidence level.
RESULTS: The Kolmogorov-Smirnov goodness-of-fit test obtained for the Monte Carlo simulation shows an extremely high agreement between the empirical distribution of the post-logarithmic-integrated x-ray intensity and the nc-Gamma. The experimental validation performed with both phantoms confirmed the excellent match between the theoretical distribution, the proposed nc-Gamma, and sample distributions in all situations.
CONCLUSION: We derive an analytical model describing the post-log distribution of the linear attenuation coefficient in the sensor for polychromatic CT scans. We also demonstrate that the nc-Gamma distribution approximates well the theoretical distribution. This distribution also approximates well the CT numbers after reconstruction since it naturally extends across linear operations involved in filtered back projection reconstructions. This probabilistic model may provide the analytical foundation to define new likelihood-based reconstruction methodologies for polychromatic scans.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  CT; Gumbel distribution; filtered backprojection; iterative reconstruction; noncentral Gamma

Mesh:

Year:  2020        PMID: 32654155      PMCID: PMC7796558          DOI: 10.1002/mp.14384

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Monotonic algorithms for transmission tomography.

Authors:  H Erdoğan; J A Fessler
Journal:  IEEE Trans Med Imaging       Date:  1999-09       Impact factor: 10.048

2.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

3.  Properties of preprocessed sinogram data in x-ray computed tomography.

Authors:  Bruce R Whiting; Parinaz Massoumzadeh; Orville A Earl; Joseph A O'Sullivan; Donald L Snyder; Jeffrey F Williamson
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

4.  Electronic noise modeling in statistical iterative reconstruction.

Authors:  Jingyan Xu; Benjamin M W Tsui
Journal:  IEEE Trans Image Process       Date:  2009-04-24       Impact factor: 10.856

5.  An experimental study on the noise properties of x-ray CT sinogram data in Radon space.

Authors:  Jing Wang; Hongbing Lu; Zhengrong Liang; Daria Eremina; Guangxiang Zhang; Su Wang; John Chen; James Manzione
Journal:  Phys Med Biol       Date:  2008-06-03       Impact factor: 3.609

6.  Quantifying the Importance of the Statistical Assumption in Statistical X-ray CT Image Reconstruction.

Authors:  Jingyan Xu; Benjamin M W Tsui
Journal:  IEEE Trans Med Imaging       Date:  2013-08-30       Impact factor: 10.048

7.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

8.  Statistical characterization of noise for spatial standardization of CT scans: Enabling comparison with multiple kernels and doses.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Maria J Ledesma-Carbayo; George R Washko; Raúl San José Estépar
Journal:  Med Image Anal       Date:  2017-06-07       Impact factor: 8.545

9.  Correction for beam hardening in computed tomography.

Authors:  G T Herman
Journal:  Phys Med Biol       Date:  1979-01       Impact factor: 3.609

10.  Harmonization of chest CT scans for different doses and reconstruction methods.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Maria Jesus Ledesma-Carbayo; George R Washko; Raúl San José Estépar
Journal:  Med Phys       Date:  2019-06-07       Impact factor: 4.071

View more
  1 in total

1.  Harmonization of in-plane resolution in CT using multiple reconstructions from single acquisitions.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Gabriel Ramos-Llordén; Raúl San José Estépar
Journal:  Med Phys       Date:  2021-09-14       Impact factor: 4.071

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.