Harmonization of chest CT scans for different doses and reconstruction methods.

Gonzalo Vegas-Sánchez-Ferrero1, Maria Jesus Ledesma-Carbayo2, George R Washko3, Raúl San José Estépar1.   

Abstract

PURPOSE: To develop and validate a computed tomography (CT) harmonization technique by combining noise-stabilization and autocalibration methodologies to provide reliable densitometry measurements in heterogeneous acquisition protocols.
METHODS: We propose to reduce the effects of spatially variant noise such as nonuniform patterns of noise and biases. The method combines the statistical characterization of the signal-to-noise relationship in the CT image intensities, which allows us to estimate both the signal and spatially variant variance of noise, with an autocalibration technique that reduces the nonuniform biases caused by noise and reconstruction techniques. The method is firstly validated with anthropomorphic synthetic images that simulate CT acquisitions with variable scanning parameters: different dosage, nonhomogeneous variance of noise, and various reconstruction methods. We finally evaluate these effects and the ability of our method to provide consistent densitometric measurements in a cohort of clinical chest CT scans from two vendors (Siemens, n = 54 subjects; and GE, n = 50 subjects) acquired with several reconstruction algorithms (filtered back-projection and iterative reconstructions) with high-dose and low-dose protocols.
RESULTS: The harmonization reduces the effect of nonhomogeneous noise without compromising the resolution of the images (25% RMSE reduction in both clinical datasets). An analysis through hierarchical linear models showed that the average biases induced by differences in dosage and reconstruction methods are also reduced up to 74.20%, enabling comparable results between high-dose and low-dose reconstructions. We also assessed the statistical similarity between acquisitions obtaining increases of up to 30% points and showing that the low-dose vs high-dose comparisons of harmonized data obtain similar and even higher similarity than the observed for high-dose vs high-dose comparisons of nonharmonized data.
CONCLUSION: The proposed harmonization technique allows to compare measures of low-dose with high-dose acquisitions without using a specific reconstruction as a reference. Since the harmonization does not require a precalibration with a phantom, it can be applied to retrospective studies. This approach might be suitable for multicenter trials for which a reference reconstruction is not feasible or hard to define due to differences in vendors, models, and reconstruction techniques.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  COPD; CT scanner; Hounsfield Unit correction; calibration; lung density; quantitative imaging

Mesh:

Year:  2019        PMID: 31069809      PMCID: PMC7251983          DOI: 10.1002/mp.13578

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

1.  Description and implementation of a quality control program in an imaging-based clinical trial.

Authors:  Christopher H Cagnon; Dianna D Cody; Michael F McNitt-Gray; J Anthony Seibert; Philip F Judy; Denise R Aberle
Journal:  Acad Radiol       Date:  2006-11       Impact factor: 3.173

Review 2.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

3.  Improved correlation between CT emphysema quantification and pulmonary function test by density correction of volumetric CT data based on air and aortic density.

Authors:  Song Soo Kim; Joon Beom Seo; Namkug Kim; Eun Jin Chae; Young Kyung Lee; Yeon Mok Oh; Sang Do Lee
Journal:  Eur J Radiol       Date:  2012-05-20       Impact factor: 3.528

4.  Current Status and Future Opportunities in Lung Precision Medicine Research with a Focus on Biomarkers. An American Thoracic Society/National Heart, Lung, and Blood Institute Research Statement.

Authors:  Ann Chen Wu; James P Kiley; Patricia J Noel; Shashi Amur; Esteban G Burchard; John P Clancy; Joshua Galanter; Maki Inada; Tiffanie K Jones; Jonathan A Kropski; James E Loyd; Lawrence M Nogee; Benjamin A Raby; Angela J Rogers; David A Schwartz; Don D Sin; Avrum Spira; Scott T Weiss; Lisa R Young; Blanca E Himes
Journal:  Am J Respir Crit Care Med       Date:  2018-12-15       Impact factor: 21.405

5.  Sinogram Affirmed Iterative Reconstruction (SAFIRE) versus weighted filtered back projection (WFBP) effects on quantitative measure in the COPDGene 2 test object.

Authors:  J P Sieren; E A Hoffman; M K Fuld; K S Chan; J Guo; J D Newell
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

6.  Denoising of polychromatic CT images based on their own noise properties.

Authors:  Ji Hye Kim; Yongjin Chang; Jong Beom Ra
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

7.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

8.  Statistical characterization of noise for spatial standardization of CT scans: Enabling comparison with multiple kernels and doses.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Maria J Ledesma-Carbayo; George R Washko; Raúl San José Estépar
Journal:  Med Image Anal       Date:  2017-06-07       Impact factor: 8.545

Review 9.  Methods and challenges in quantitative imaging biomarker development.

Authors:  Richard G Abramson; Kirsteen R Burton; John-Paul J Yu; Ernest M Scalzetti; Thomas E Yankeelov; Andrew B Rosenkrantz; Mishal Mendiratta-Lala; Brian J Bartholmai; Dhakshinamoorthy Ganeshan; Leon Lenchik; Rathan M Subramaniam
Journal:  Acad Radiol       Date:  2015-01       Impact factor: 3.173

10.  Effects of CT section thickness and reconstruction kernel on emphysema quantification relationship to the magnitude of the CT emphysema index.

Authors:  David S Gierada; Andrew J Bierhals; Cliff K Choong; Seth T Bartel; Jon H Ritter; Nitin A Das; Cheng Hong; Thomas K Pilgram; Kyongtae T Bae; Bruce R Whiting; Jason C Woods; James C Hogg; Barbara A Lutey; Richard J Battafarano; Joel D Cooper; Bryan F Meyers; G Alexander Patterson
Journal:  Acad Radiol       Date:  2010-02       Impact factor: 3.173

View more
  2 in total

1.  Harmonization of in-plane resolution in CT using multiple reconstructions from single acquisitions.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Gabriel Ramos-Llordén; Raúl San José Estépar
Journal:  Med Phys       Date:  2021-09-14       Impact factor: 4.071

2.  Statistical characterization of the linear attenuation coefficient in polychromatic CT scans.

Authors:  Gonzalo Vegas; Raúl San José Estépar
Journal:  Med Phys       Date:  2020-10-06       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.