Harmonization of in-plane resolution in CT using multiple reconstructions from single acquisitions.

Gonzalo Vegas-Sánchez-Ferrero1, Gabriel Ramos-Llordén2, Raúl San José Estépar1.   

Abstract

PURPOSE: To providea methodology that removes the spatial variability of in-plane resolution using different CT reconstructions. The methodology does not require any training, sinogram, or specific reconstruction method.
METHODS: The methodology is formulated as a reconstruction problem. The desired sharp image is modeled as an unobservable variable to be estimated from an arbitrary number of observations with spatially variant resolution. The methodology comprises three steps: (1) density harmonization, which removes the density variability across reconstructions; (2) point spread function (PSF) estimation, which estimates a spatially variant PSF with arbitrary shape; (3) deconvolution, which is formulated as a regularized least squares problem. The assessment was performed with CT scans of phantoms acquired with three different Siemens scanners (Definition AS, Definition AS+, Drive). Four low-dose acquisitions reconstructed with backprojection and iterative methods were used for the resolution harmonization. A sharp, high-dose (HD) reconstruction was used as a validation reference. The different factors affecting the in-plane resolution (radial, angular, and longitudinal) were studied with regression analysis of the edge decay (between 10% and 90% of the edge spread function (ESF) amplitude).
RESULTS: Results showed that the in-plane resolution improves remarkably and the spatial variability is substantially reduced without compromising the noise characteristics. The modulated transfer function (MTF) also confirmed a pronounced increase in resolution. The resolution improvement was also tested by measuring the wall thickness of tubes simulating airways. In all scanners, the resolution harmonization obtained better performance than the HD, sharp reconstruction used as a reference (up to 50 percentage points). The methodology was also evaluated in clinical scans achieving a noise reduction and a clear improvement in thin-layered structures. The estimated ESF and MTF confirmed the resolution improvement.
CONCLUSION: We propose a versatile methodology to reduce the spatial variability of in-plane resolution in CT scans by leveraging different reconstructions available in clinical studies. The methodology does not require any sinogram, training, or specific reconstruction, and it is not limited to a fixed number of input images. Therefore, it can be easily adopted in multicenter studies and clinical practice. The results obtained with our resolution harmonization methodology evidence its suitability to reduce the spatially variant in-plane resolution in clinical CT scans without compromising the reconstruction's noise characteristics. We believe that the resolution increase achieved by our methodology may contribute in more accurate and reliable measurements of small structures such as vasculature, airways, and wall thickness.
© 2021 American Association of Physicists in Medicine.

Entities:  

Keywords:  CT scans; deconvolution; in-plane resolution; reconstruction

Mesh:

Year:  2021        PMID: 34432901      PMCID: PMC8710308          DOI: 10.1002/mp.15186

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  50 in total

1.  Enhancement of in-plane spatial resolution in volumetric Computed Tomography with focal spot wobbling - overcoming the constraint on number of projection views per gantry rotation.

Authors:  Xiangyang Tang; Suresh Narayanan; Jiang Hsieh; Jed D Pack; Scott M Mcolash; Paavana Sainath; Roy A Nilsen; Basel Taha
Journal:  J Xray Sci Technol       Date:  2010       Impact factor: 1.535

2.  Monte-Carlo sure: a black-box optimization of regularization parameters for general denoising algorithms.

Authors:  Sathish Ramani; Thierry Blu; Michael Unser
Journal:  IEEE Trans Image Process       Date:  2008-09       Impact factor: 10.856

3.  Channelized Hotelling observers for the assessment of volumetric imaging data sets.

Authors:  Ljiljana Platiša; Bart Goossens; Ewout Vansteenkiste; Subok Park; Brandon D Gallas; Aldo Badano; Wilfried Philips
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2011-06-01       Impact factor: 2.129

4.  Accurate measurement of intrathoracic airways.

Authors:  J M Reinhardt; N D D'Souza; E A Hoffman
Journal:  IEEE Trans Med Imaging       Date:  1997-12       Impact factor: 10.048

5.  Comparison of low-contrast detectability between two CT reconstruction algorithms using voxel-based 3D printed textured phantoms.

Authors:  Justin Solomon; Alexandre Ba; François Bochud; Ehsan Samei
Journal:  Med Phys       Date:  2016-12       Impact factor: 4.071

6.  Impact of miscentering on patient dose and image noise in x-ray CT imaging: phantom and clinical studies.

Authors:  M A Habibzadeh; M R Ay; A R Kamali Asl; H Ghadiri; H Zaidi
Journal:  Phys Med       Date:  2011-07-08       Impact factor: 2.685

7.  The line spread function and modulation transfer function of a computed tomographic scanner.

Authors:  P F Judy
Journal:  Med Phys       Date:  1976 Jul-Aug       Impact factor: 4.071

8.  Autocalibration method for non-stationary CT bias correction.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Maria J Ledesma-Carbayo; George R Washko; Raúl San José Estépar
Journal:  Med Image Anal       Date:  2017-12-08       Impact factor: 8.545

9.  Airway wall thickening on CT: Relation to smoking status and severity of COPD.

Authors:  Jean-Paul Charbonnier; Esther Pompe; Camille Moore; Stephen Humphries; Bram van Ginneken; Barry Make; Elizabeth Regan; James D Crapo; Eva M van Rikxoort; David A Lynch
Journal:  Respir Med       Date:  2018-11-20       Impact factor: 3.415

Review 10.  Model observers in medical imaging research.

Authors:  Xin He; Subok Park
Journal:  Theranostics       Date:  2013-10-04       Impact factor: 11.556

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.