Validation of Fourier analysis of videokeratographic data.

Haris Sideroudi1,2, Georgios Labiris3,4, Fienke Ditzel5, Efi Tsaragli3, Kimonas Georgatzoglou3, Haralampos Siganos6, Vassilios Kozobolis3,4.   

Abstract

PURPOSE: The aim was to assess the repeatability of Fourier transfom analysis of videokeratographic data using Pentacam in normal (CG), keratoconic (KC) and post-CXL (CXL) corneas.
METHODS: This was a prospective, clinic-based, observational study. One randomly selected eye from all study participants was included in the analysis: 62 normal eyes (CG group), 33 keratoconus eyes (KC group), while 34 eyes, which had already received CXL treatment, formed the CXL group. Fourier analysis of keratometric data were obtained using Pentacam, by two different operators within each of two sessions. Precision, repeatability and Intraclass Correlation Coefficient (ICC), were calculated for evaluating intrassesion and intersession repeatability for the following parameters: Spherical Component (SphRmin, SphEcc), Maximum Decentration (Max Dec), Regular Astigmatism, and Irregularitiy (Irr). Bland-Altman analysis was used for assessing interobserver repeatability.
RESULTS: All parameters were presented to be repeatable, reliable and reproductible in all groups. Best intrasession and intersession repeatability and reliability were detected for parameters SphRmin, SphEcc and Max Dec parameters for both operators using ICC (intrasession: ICC > 98%, intersession: ICC > 94.7%) and within subject standard deviation. Best precision and lowest range of agreement was found for the SphRmin parameter (CG: 0.05, KC: 0.16, and CXL: 0.2) in all groups, while the lowest repeatability, reliability and reproducibility was detected for the Irr parameter.
CONCLUSIONS: The Pentacam system provides accurate measurements of Fourier tranform keratometric data. A single Pentacam scan will be sufficient for most clinical applications.

Entities:  

Keywords:  Corneal crosslinking; Fourier; Irregular astigmatism; Keratoconus; Reliability; Repeatability; Reproducibility; Scheimpflug camera

Mesh:

Year:  2017        PMID: 28620708     DOI: 10.1007/s10792-017-0602-7

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  23 in total

1.  Escalation of keratospeak: regular and irregular astigmatism.

Authors:  P Wyzinski
Journal:  Refract Corneal Surg       Date:  1990 May-Jun

2.  Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements.

Authors:  Hema Shankar; Deepa Taranath; Chandramalar T Santhirathelagan; Konrad Pesudovs
Journal:  J Cataract Refract Surg       Date:  2008-01       Impact factor: 3.351

3.  Measurement error.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-09-21

4.  Measurement error and correlation coefficients.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-07-06

5.  Variability in Scheimpflug image-derived posterior elevation measurements in keratoconus and collagen-crosslinked corneas.

Authors:  Georgios Labiris; Athanassios Giarmoukakis; Haris Sideroudi; Panagiota Bougatsou; Ilias Lazaridis; Vassilios P Kozobolis
Journal:  J Cataract Refract Surg       Date:  2012-09       Impact factor: 3.351

6.  Repeatability, reliability and reproducibility of posterior curvature and wavefront aberrations in keratoconic and cross-linked corneas.

Authors:  Haris Sideroudi; Georgios Labiris; Athanassios Giarmoulakis; Najia Bougatsou; Dimitrios Mikropoulos; Vassilios Kozobolis
Journal:  Clin Exp Optom       Date:  2013-04-25       Impact factor: 2.742

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Keratoconus detected by videokeratography in candidates for photorefractive keratectomy.

Authors:  A B Nesburn; S Bahri; J Salz; Y S Rabinowitz; E Maguen; J Hofbauer; M Berlin; J I Macy
Journal:  J Refract Surg       Date:  1995 May-Jun       Impact factor: 3.573

9.  Longitudinal changes in corneal irregular astigmatism and visual acuity in eyes with keratoconus.

Authors:  Mariko Suzuki; Shiro Amano; Norihiko Honda; Tomohiko Usui; Satoru Yamagami; Tetsuro Oshika
Journal:  Jpn J Ophthalmol       Date:  2007-08-03       Impact factor: 2.447

10.  Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry.

Authors:  Julio C Hernández-Camarena; Patricia Chirinos-Saldaña; Alejandro Navas; Arturo Ramirez-Miranda; Angle de la Mota; Aida Jimenez-Corona; Enrique O Graue-Hernindez
Journal:  J Refract Surg       Date:  2014-09       Impact factor: 3.573

View more
  2 in total

1.  Fourier analysis of corneal Scheimpflug imaging: clinical use in keratoconus.

Authors:  Dorukcan Akincioglu; Gokhan Ozge; Onder Ayyildiz; Gokcen Gokce; Umut Karaca; Fatih Mehmet Mutlu
Journal:  Int J Ophthalmol       Date:  2021-12-18       Impact factor: 1.779

Review 2.  Fourier Analysis of Keratometric Data in Epithelium Removal versus Epithelial Disruption Corneal Cross-linking.

Authors:  Shahram Bamdad; Seyed Mohammad Salar Zaheryani; Sahar Mohaghegh; Mohammad Shirvani
Journal:  J Ophthalmic Vis Res       Date:  2020-02-02
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.