| Literature DB >> 28503479 |
Abstract
OBJECTIVES: To compare the fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments under static model via artificial canals with different angles of curvature.Entities:
Keywords: Cyclic Fatigue; Reciproc; Static Model; Tango-Endo; WaveOne GOLD
Year: 2017 PMID: 28503479 PMCID: PMC5426221 DOI: 10.5395/rde.2017.42.2.134
Source DB: PubMed Journal: Restor Dent Endod ISSN: 2234-7658
The mean and standard deviation values of the time (unit: seconds) until tested NiTi files fractured
| Group | Canal 1 | Canal 2 | Canal 3 | |
|---|---|---|---|---|
| Angle of curvature | 45° | 60° | 90° | |
| Tango-Endo | 2171.36 ± 143.36a | 968.16 ± 63.91a | 290.44 ± 19.19a | |
| WaveOne GOLD | 434.28 ± 128.68b | 276.60 ± 18.26b | 116.18 ± 7.66b | |
| Reciproc | 415.63 ± 34.69b | 264.73 ± 22.10b | 111.18 ± 9.27b |
Different superscripts indicate statistically significant difference in the column (p < 0.05).
Figure 1Scanning electron microscopic appearances of the Tango-Endo, WaveOne GOLD, and Reciproc files after cyclic fatigue testing. (a, c, and e) General view of Tango-Endo, WaveOne GOLD, and Reciproc, respectively, (white arrows, showing fracture zone); (b, d, and f) High-magnification view of Tango-Endo, WaveOne GOLD, and Reciproc instruments showing typical fatigue striations of cyclic fatigue (white arrows, showing fatigue striations), respectively.