Literature DB >> 22892743

Differences in cyclic fatigue resistance at apical and coronal levels of Reciproc and WaveOne new files.

Ana Arias1, Juan J Perez-Higueras, José C de la Macorra.   

Abstract

INTRODUCTION: Recent improvements in alloys, kinematics, and concepts have been combined to increase the cyclic fatigue resistance of nickel-titanium (NiTi) instruments. The aim of this study was to compare the cyclic fatigue resistance of new M-Wire reciprocating WaveOne (Dentsply Maillefer, Ballaigues, Switzerland) and Reciproc (VDW GmbH, Munich, Germany) files at 2 levels.
METHODS: Sixty Reciproc and 60 WaveOne new files were fixed to a specifically designed device and tested in tempered steel canals with a 3-mm radius and a 60° angle of curvature. The motor used was programmed as defined by each manufacturer, and the specific reciprocating motion was followed. Thirty files of each brand were tested at 5 mm, and 30 were tested at 13 mm from their tips. The time to failure was registered. Weibull analysis was used to calculate the mean life, beta, and eta parameters.
RESULTS: The probability of the mean life was higher for Reciproc than WaveOne files at both levels, with the probability of the Reciproc mean life being 62% higher than that of WaveOne at 5 mm from the tip and 100% higher at 13 mm (all statistically significant). The probability of the mean life was higher at 5 mm than at 13 mm in both systems. The results may have been partially conditioned by the different motions that manufacturers propose for each system.
CONCLUSIONS: Reciproc files were more resistant to cyclic fatigue than WaveOne files at both distances from the tip. Both systems had greater cyclic fatigue resistance at 5 mm than at 13 mm from the tip.
Copyright © 2012 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22892743     DOI: 10.1016/j.joen.2012.05.022

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  10 in total

1.  Endodontic retreatment: clinical comparison of reciprocating systems versus rotary system in disinfecting root canals.

Authors:  Frederico C Martinho; Lilian F Freitas; Gustavo G Nascimento; Aleteia M Fernandes; Fabio R M Leite; Ana P M Gomes; Izabel C G Camões
Journal:  Clin Oral Investig       Date:  2014-11-21       Impact factor: 3.573

2.  Analysis of the Importance of the Motion Used in the Resistance of Different Mechanical Instrumentation Systems in Endodontics: A Comparative Study.

Authors:  Jesús Mena-Álvarez; Manuel Almanzor-López; Norberto Quispe-López; Ana De Pedro-Muñoz; Cristina Rico-Romano
Journal:  Materials (Basel)       Date:  2022-06-24       Impact factor: 3.748

3.  Effect of canal curvature location on the cyclic fatigue resistance of reciprocating files.

Authors:  Tyler Sobotkiewicz; Xiangya Huang; Markus Haapasalo; Christophe Mobuchon; Ahmed Hieawy; Jinghao Hu; Huimin Zhou; Zhejun Wang; Ya Shen
Journal:  Clin Oral Investig       Date:  2020-06-02       Impact factor: 3.573

4.  Histological evaluation of the cleaning effectiveness of two reciprocating single-file systems in severely curved root canals: Reciproc versus WaveOne.

Authors:  Maira de Souza Carvalho; Emílio Carlos Sponchiado; Angela Delfina Bitencourt Garrido; Lucas da Fonseca Roberti Garcia; André Augusto Franco Marques
Journal:  Eur J Dent       Date:  2015 Jan-Mar

Review 5.  Single file reciprocating systems: A systematic review and meta-analysis of the literature: Comparison of reciproc and WaveOne.

Authors:  Rana Alsilani; Fatima Jadu; Dania F Bogari; Ahmed M Jan; Turki Y Alhazzazi
Journal:  J Int Soc Prev Community Dent       Date:  2016-10-24

6.  Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments.

Authors:  Koray Yılmaz; Taha Özyürek
Journal:  Restor Dent Endod       Date:  2017-04-11

7.  Wear of the Primary WaveOne single file when shaping vestibular root canals of first maxillary molar.

Authors:  Daniel Aracena; Eduardo Borie; Pablo Betancourt; Angella Aracena; Mario Guzmán
Journal:  J Clin Exp Dent       Date:  2017-03-01

Review 8.  Effectiveness of rotary and reciprocating systems on microbial reduction: A systematic review.

Authors:  Riluwan Siddique; Malli Sureshbabu Nivedhitha
Journal:  J Conserv Dent       Date:  2019 Mar-Apr

9.  Bending Resistance and Cyclic Fatigue Life of a new Single-File Reciprocating Instrument Waveone Gold.

Authors:  Emmanuel João Nogueira Leal Silva; Justine Monnerat Tinoco; Eduardo Muniz Barretto Tinoco; Victor Talarico Leal Vieira; Luciana Moura Sassone; Hélio Pereira Lopes
Journal:  Eur Endod J       Date:  2016-10-11

10.  In vitro comparison of cyclic fatigue resistance of TruNatomy in single and double curvature canals compared with different nickel-titanium rotary instruments.

Authors:  Amr M Elnaghy; Shaymaa E Elsaka; Ayman O Mandorah
Journal:  BMC Oral Health       Date:  2020-02-04       Impact factor: 2.757

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.