The purpose of this study was to analyze the number and types of defects observed in single-use, rotary nickel-titanium instruments. Every ProFile Series 29.04 taper nickel-titanium instrument used during a 4-week period in an endodontic specialty practice was collected. All instruments were new and were used by experienced clinicians during a single patient visit. The instruments were routinely used in a crown-down manner with RC Prep lubrication and copious irrigation. The instruments were used in a MicroMega 324 air motor in a 6:1 gear reduction contra-angle at 333 rpm. The instruments were collected, ultrasonically cleaned, sterilized, and inspected at x16 magnification. Torsional, flexural, and fracture defects were recorded and statistical analysis was performed using the Kruskal-Wallis one-way analysis of variance. A total of 786 ProFile Series 29 nickel-titanium rotary instruments were evaluated; 115 (14.63%) showed some type of defect after one clinical use. Size 3 instruments had the highest defect rate (22.66%) followed by size 5 (17.30%), size 2 (17.24%), and size 4 instruments (16.10%). However, there was no statistically significant difference. The size 6 and size 7 instruments showed minimal defects (2.38% and 4.76%, respectfully). Seven of 786 files had fractured (0.891%). There was no statistically significant difference in the type of failure seen within each file size. This study does show that defects can occur even with new files in the hands of experienced endodontists, and for absolute safety a single-use approach should be followed.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14606792     DOI: 10.1097/00004770-200310000-00013

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  18 in total

1.  Non-destructive natural frequency tests of cyclic fatigue-loaded nickel-titanium rotary instruments.

Authors:  Sung-Chih Hsieh; Sheng-Yang Lee; Chun-Yu Ciou; Haw-Ming Huang
Journal:  Med Biol Eng Comput       Date:  2010-06       Impact factor: 2.602

2.  Microstructural, microchemical, and mechanical changes associated with the clinical reuse of two nickel-titanium endodontic instruments.

Authors:  Felipe Augusto Restrepo-Restrepo; Viviana Andrea Holguín-Vásquez; Syldana Julieth Cañas-Jiménez; Paula Andrea Villa-Machado; Sara Ochoa-Soto; Claudia Patricia Ossa-Orozco; Sergio Iván Tobón-Arroyave
Journal:  Dent Res J (Isfahan)       Date:  2021-06-22

3.  Effect of repetitive pecking at working length for glide path preparation using G-file.

Authors:  Jung-Hong Ha; Hyo-Jin Jeon; Rashid El Abed; Seok-Woo Chang; Sung-Kyo Kim; Hyeon-Cheol Kim
Journal:  Restor Dent Endod       Date:  2015-01-07

4.  Self-adjusting file (SAF) separation in clinical use: A preliminary survey among experienced SAF users regarding prevalence and retrieval methods.

Authors:  Michael Solomonov; Joe Ben-Itzhak; Anda Kfir; Oscar von Stetten; Elena Lipatova; Eleftherios T Farmakis
Journal:  J Conserv Dent       Date:  2015 May-Jun

5.  An in vitro comparison of root canal system prepared with either hand or rotary instruments.

Authors:  Azza A Dafalla; Neamat Hassan Abubakr; Yahia E Ibrahim
Journal:  Iran Endod J       Date:  2010-11-15

6.  Influence of the glide path on various parameters of root canal prepared with WaveOne reciprocating file using cone beam computed tomography.

Authors:  Anil Dhingra; Nidhi Nagar; Vipul Sapra
Journal:  Dent Res J (Isfahan)       Date:  2015 Nov-Dec

7.  Effect of adaptive motion on cyclic fatigue resistance of a nickel titanium instrument designed for retreatment.

Authors:  Taha Özyürek; Koray Yılmaz; Gülşah Uslu
Journal:  Restor Dent Endod       Date:  2016-12-19

8.  Shot peening increases resistance to cyclic fatigue fracture of endodontic files.

Authors:  Javier Nino-Barrera; Jose Sanchez-Aleman; Manuel Acosta-Humanez; Luis Gamboa-Martinez; Carlos Cortes-Rodriguez
Journal:  Sci Rep       Date:  2021-06-21       Impact factor: 4.379

9.  A survey of experience-based preference of Nickel-Titanium rotary files and incidence of fracture among general dentists.

Authors:  Woocheol Lee; Minju Song; Euiseong Kim; Hyojin Lee; Hyeon-Cheol Kim
Journal:  Restor Dent Endod       Date:  2012-11-21

10.  Cyclic fatigue resistance of OneShape, Reciproc, and WaveOne: An in vitro comparative study.

Authors:  Alberto Dagna; Claudio Poggio; Riccardo Beltrami; Marco Colombo; Marco Chiesa; Stefano Bianchi
Journal:  J Conserv Dent       Date:  2014-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.