Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer.

Oscar Fuster1, Belinda Andino1, Amparo Pardo1, Begoña Laiz1.   

Abstract

BACKGROUND: Accurate evaluation of hematology analyzers is recommended before these devices can be broadly introduced for the routine testing of continuous ambulatory peritoneal dialysis (CAPD), ascitic, and pleural fluids.
METHODS: We evaluated the performance of Mindray BC-6800 for white blood cell (WBC) and differential cell count in 50 CAPD, 60 ascitic and 40 pleural compared with manual microscopy. Within-run precision, limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), and carryover were assessed.
RESULTS: The Passing-Bablok regression in all fluids showed the following equations: yWBC =1.05x+3.31 (95%CI slope 0.95 to 1.12; intercept -0.25 to 5.52); yMN =0.85x+15.63 (95%CI slope 0.72 to 1.05; intercept -24.18 to 84.47); and yPMN =1.21x+13.37 (95%CI slope 1.03 to 1.35; intercept 4.00 to 32.47) with bias 78 cells/μL. The AUC for clinical PMN cut-off was 0.88 (95%CI: 0.77 to 0.98). In ascitic, pleural, and CAPD fluids the AUC for clinical PMN cut-off were 0.88 (95%CI: 0.63 to 1.00), 0.83 (95%CI: 0.68 to 0.99), and 1.00 (95%CI: 1.00 to 1.00) respectively. CV ranged from 3%-34%. LoB of 3 cell/μL was verified. LoD and LoQ reported the same result (8 cells/μL). Carry over never exceeded 0.05%.
CONCLUSION: The effectiveness of BC-6800 to categorize cells from different body fluids was not compromised by the slight positive bias observed. This conclusion is supported by the high AUC and agreement between the automated method and the reference method. The results show that BC-6800 offers rapid, accurate, and reproducible results for clinical management of CAPD, ascitic, and pleural fluids.
© 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  automated hematology analyzer; cell count; differential count; serous fluid

Mesh:

Year:  2017        PMID: 28444939      PMCID: PMC6816949          DOI: 10.1002/jcla.22240

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  22 in total

1.  Investigation of a unilateral pleural effusion in adults: British Thoracic Society Pleural Disease Guideline 2010.

Authors:  Clare Hooper; Y C Gary Lee; Nick Maskell
Journal:  Thorax       Date:  2010-08       Impact factor: 9.139

2.  Peritoneal dialysis-related infections recommendations: 2010 update.

Authors:  Philip Kam-Tao Li; Cheuk Chun Szeto; Beth Piraino; Judith Bernardini; Ana E Figueiredo; Amit Gupta; David W Johnson; Ed J Kuijper; Wai-Choong Lye; William Salzer; Franz Schaefer; Dirk G Struijk
Journal:  Perit Dial Int       Date:  2010 Jul-Aug       Impact factor: 1.756

3.  Guidelines on the management of ascites in cirrhosis.

Authors:  K P Moore; G P Aithal
Journal:  Gut       Date:  2006-10       Impact factor: 23.059

4.  Automated evaluation of synovial and ascitic fluids with the Advia 2120 hematology analyzer.

Authors:  Paul Froom; Alaa Diab; Mira Barak
Journal:  Am J Clin Pathol       Date:  2013-12       Impact factor: 2.493

Review 5.  Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease.

Authors:  Chérina Fleming; Henk Russcher; Jan Lindemans; Robert de Jonge
Journal:  Clin Chem Lab Med       Date:  2015-10       Impact factor: 3.694

Review 6.  Body fluid cell counts by automated methods.

Authors:  Linda M Sandhaus
Journal:  Clin Lab Med       Date:  2014-12-04       Impact factor: 1.935

7.  Evaluation of white blood cell count in peritoneal fluid with five different hemocytometers.

Authors:  Giuseppe Lippi; Clarissa Cattabiani; Anna Benegiamo; Daniela Gennari; Fernanda Pavesi; Alberta Caleffi; Silvia Pipitone
Journal:  Clin Biochem       Date:  2012-11-02       Impact factor: 3.281

8.  Variations in pleural fluid WBC count and differential counts with different sample containers and different methods.

Authors:  Barrett D Conner; Y C Gary Lee; Paul Branca; Jeffrey T Rogers; R Michael Rodriguez; Richard W Light
Journal:  Chest       Date:  2003-04       Impact factor: 9.410

9.  UF-1000i: validation of the body fluid mode for counting cells in body fluids.

Authors:  Chérina Fleming; Rob Brouwer; Adriaan van Alphen; Jan Lindemans; Robert de Jonge
Journal:  Clin Chem Lab Med       Date:  2014-12       Impact factor: 3.694

10.  Mindray BC-6800 body fluid mode, performance of nucleated cells, and differential count in ascitic and pleural fluids.

Authors:  S Buoro; T Mecca; G Azzarà; M Seghezzi; E Candiago; A Gianatti; A Crippa; A La Gioia
Journal:  Int J Lab Hematol       Date:  2015-11-07       Impact factor: 2.877

View more
  3 in total

1.  Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer.

Authors:  Oscar Fuster; Belinda Andino; Amparo Pardo; Begoña Laiz
Journal:  J Clin Lab Anal       Date:  2017-04-26       Impact factor: 2.352

2.  Ascitic fluid polymorphic nuclear cell count impacts on outcome of cirrhotic patients with ascites.

Authors:  Benedikt Simbrunner; Annika Röthenbacher; Helmuth Haslacher; David Bauer; David Chromy; Theresa Bucsics; Philipp Schwabl; Rafael Paternostro; Bernhard Scheiner; Michael Trauner; Mattias Mandorfer; Ilse Schwarzinger; Thomas Reiberger
Journal:  United European Gastroenterol J       Date:  2019-04-05       Impact factor: 4.623

3.  Consistency Analysis of 3 Detection Systems for Measuring Serum C-Reactive Protein.

Authors:  Xueyan Chen; Chuanyihong Huang; Weichuan Zhong; Suwen Qi; Ting Huang; Zihua Yang
Journal:  Med Sci Monit       Date:  2022-05-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.