Literature DB >> 35538647

Consistency Analysis of 3 Detection Systems for Measuring Serum C-Reactive Protein.

Xueyan Chen1, Chuanyihong Huang1, Weichuan Zhong1, Suwen Qi2, Ting Huang3, Zihua Yang4.   

Abstract

BACKGROUND C-reactive protein (CRP) is an important clinical indicator. There are many methods and instruments for CRP measurement, and therefore the consistency of CRP values measured between instruments needs to be evaluated. This study aimed to compare the consistency of 3 serum CRP detection systems using turbidimetry. MATERIAL AND METHODS The consistency of CRP measured by 3 instruments, the Mindray BC-5390, Mindray BC-6800, and Johnson Vitros5600, was evaluated, and the consistency of blood routine measurement between the BC-5390 and BC-6800 was also evaluated. Pearson correlation analysis was used to evaluate the correlation of different instrument's test results (R, correlation coefficient). The consistency of instruments was assessed by Passing-Bablok analysis and weighted Deming analysis. RESULTS CRP data and route blood test data from 847 patients were used for analysis. The results showed that there were differences in the CRP values measured by the Mindray BC5390, Mindray BC6800, and Johnson Vitros5600 (χ²=78.573, P<0.001). The CRP measurement results of the BC5390 analyzer were consistent with those of the BC6800 analyzer (R=0.994, P<0.001) and Vitros5600 analyzer (R=0.983, P<0.001). However, there was a constant deviation in the CRP values measured by the BC-6800 and Vitros5600 analyzer (R=0.994, P<0.001). In the measurement of routine blood laboratory tests, the BC5390 analyzer and BC6800 analyzer were found to be interchangeable. CONCLUSIONS This study analyzed the consistency of CRP detection by 3 instruments, the Mindray BC-5390, Mindray BC-6800, and Johnson Vitros5600, and may provide a reference for the selection of CRP detection instruments.

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 35538647      PMCID: PMC9109607          DOI: 10.12659/MSM.935171

Source DB:  PubMed          Journal:  Med Sci Monit        ISSN: 1234-1010


Background

C-reactive protein (CRP) is an acute-phase reactive protein produced by the liver and is present in blood plasma [1]. After tissue injury, the blood CRP rises rapidly, with a CRP level that is 1000 times higher than the baseline level within 24 h [2,3]. This phenomenon makes the blood CRP level an early indicator of infection or inflammatory diseases, as well as a universal biomarker for many diseases [4-8]. Therefore, the detection of CRP levels in the blood of patients is essential for the diagnosis, monitoring, and intervention of disease. The analytical methods for assessing CRP levels should have the characteristics of a lower sample volume, sensitivity, good selectivity, rapidity, and trustworthiness. A variety of techniques have been reported for the in vitro detection of CRP, such as turbidimetry based assays (immunity transmission turbidity and immune scatter turbidity), enzyme-linked immunosorbent assay (ELISA), chemiluminescent, fluorescent, and electrochemical assays [9-13]. The ability of these methods to test the concentration of CRP is different: the turbidimetry-based assays can detect CRP in μg mL−1, while the methods of ELISA, chemiluminescent, fluorescent, and electrochemical assays can detect CRP in fg mL−1 [9]. For the measurement of CRP, different detection systems, reagents, and method principles can cause differences in detection results. Testing institutions commonly have different types of CRP testing instruments, and the consistency of different test results needs to be evaluated. In addition, the instrument should be easy to implement and use widely in clinical practice. The purpose of this study was to compare the consistency of 3 serum CRP detection systems, the Mindray BC-5390, Mindray BC-6800, and Johnson Vitros5600, and select the CRP detection instrument most suitable for widespread use in clinical practice.

Material and Methods

Study Design and Population

This was a cross-section study, and all the patients were recruited from October 2020 to February 2021 in the People’s Hospital of Longhua Shenzhen Hospital. All included patients were tested for CRP and routine blood laboratory values using 3 different instruments. This study was approved by the Institutional Review Board of the People’s Hospital of Longhua Shenzhen Hospital and written informed consent was obtained from all patients.

Sample Handling

For CRP testing, the Mindray BC-5390 and BC-6800 analyzers use whole blood samples and the Johnson Vitros5600 analyzer use serum samples. Two blood samples were collected from all patients under fasting conditions. For testing with the Mindray analyzers, 2 mL of blood was collected in a K2-EDTA anticoagulant test tube; for the Johnson Vitros5600 analyzer, 3 to 5 mL of blood was collected in a test tube without anticoagulant. The blood sample without anticoagulant was allowed to stand for 30 min and then centrifuged at a speed of 3500 r/min for 10 min; the serum was then taken for testing. If samples could not be tested in time, the protocol was to store them in an environment at 2°C to 8°C and test as soon as possible; however, the blood samples from all patients were measured in time.

Instruments

Three instruments were used to analyze the blood samples: the Mindray BC-5390 and Mindray BC-6800 (Mindray, Shenzhen, China), and the Johnson Vitros5600 (Ortho-Clinical Diagnostics, Inc., Raritan, NJ, USA). All reagents used were those matched with the instruments. All operations were carried out in strict accordance with the instrument and kit instructions. In addition, samples for each patient were tested by the 3 instruments. The sample volumes used by the different instruments during the test were as follows: (1) Mindray BC-5390 (sample type, whole blood; single blood routine model, 20 μL; single CRP model, 20 μL; blood routine combined blood routine model, 35 μL); (2) Mindray BC-6800 (sample type, whole blood; routine blood manual model, 150 μL; routine blood automatic model, 200 μL; CRP model, 20 μL); (3) Johnson Vitros5600 (sample type, serum; CRP model, 41 μL).

Data Collection

Test data of all patients were collected, and the following variables were extracted: CRP levels, white blood cells, neutrophils (×109/L), lymphocytes (×109/L), monocytes (×109/L), eosinophils (×109/L), basophils (×109/L), neutrophils (%), lymphocytes (%), monocytes (%), eosinophils (%), basophils (%), red blood cells (RBC), hemoglobin, hematocrit, mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), mean corpuscular hemoglobin concentration (MCHC), RBC distribution width-coefficient of variation (RDW-CV), RBC distribution width-standard deviation (RDW-SD), platelets, plateletcrit, mean platelet volume (MPV), and platelet distribution width.

Statistical Analysis

Statistical comparisons between combinations of 2 instruments (Mindray BC-5390 analyzer, Mindray BC-6800 analyzer, and Johnson Vitros5600 analyzer) were performed using Pearson correlation analysis, Passing-Bablok analysis, and weighted Deming analysis. Pearson correlation analysis was used to evaluate the correlation of the different instruments’ test results (R, correlation coefficient). The consistency analysis of the results was evaluated by Passing-Bablok analysis and weighted Deming analysis, and the proportional and constant systematic errors were evaluated by the slope and intercept, with 95% confidence intervals (95% CI) [14]. The bias and 95% agreement limit of CRP value were evaluated by Bland-Altman analysis and calculated as the mean difference between the instruments. Effect analysis was used to test the consistency of the indicator direction. For the comparison of routine blood indicators, the Shapiro test was used to test the normality of measurement data. Continuous variables with a normal distribution were displayed as mean±SD, and the t test was used for comparisons between 2 groups; non-normally distributed measurement data were represented by a median and interquartile range, and the Wilcoxon rank test was used between 2 groups. All statistical analyses were 2-sided, and R software (version 4.0.3) was used to perform the analysis. P<0.05 was considered to indicate statistical significance.

Results

Comparison of CRP Values Determined by BC-5390, BC-6800, and Vitros5600

Test information was initially collected from 1058 patients. There were 201 patients with missing CRP values due to failure to automatically upload instrument test data to the hospital laboratory information system, and the BC5390 analyzer was updating and debugging the new system, including 182 cases using BC-5390, 17 cases using the BC-6800, and 3 cases using Vitros5600. In addition, the CRP values measured by the BC-5390 and BC-6800 were 0 in 4 cases and 6 cases, respectively. After excluding patients with missing CRP values and CRP values of 0, data from a total of 847 patients were included to compare the consistency of CRP values measured by the BC-5390, BC-6800, and Vitros5600. Results of comparing the CRP values tested by the 3 analyzers showed a statistical difference in values (χ2=78.573, P<0.001). The results of the Wilcoxon test showed that the differences between every combination of 2 analyzers were statistically significant (all P<0.001). The detailed results are shown in Figure 1.
Figure 1

Difference analysis of C-reactive protein levels tested by the Mindray BC-5390, Mindray BC-6800, and Johnson Vitros5600.

Comparison of Consistency of CRP Values Measured by BC-5390, BC-6800, and Vitros5600

The CRP values measured by the Mindray BC-5390 analyzer were compared with those measured by the Mindray BC-6800 analyzer (Figure 2). Pearson correlation analysis showed that the correlation coefficient of the CRP values measured by the 2 instruments was 0.994 (P<0.001). The Passing-Bablok analysis yielded the equation y=0.999(95% CI: 0.989–1.011)x+0.565(95% CI: 0.305–0.824), in which the slope range contained 1. The weighted Deming analysis yielded the equation y=1.011(95% CI: 1.005–1.029)x+0.406(95% CI: −0.0937–0.459), in which the slope and intercept were close to 1 and equal to 0, respectively. The Bland-Altman test showed that the mean bias between the 2 different instruments’ tests was −1.34 mg/L. These results indicated that the CRP values measured by the BC-5390 and BC-6800 had a certain deviation, but they were still relatively consistent.
Figure 2

Consistency analysis and Bland-Altman difference plot of C-reactive protein levels between the Mindray BC-5390 and Mindray BC-6800. (A) Pearson correlation analysis; (B) Passing-Bablok analysis; (C) weighted Deming analysis; and (D) Bland-Altman test.

The consistency analyses of CRP between the Mindray BC-6800 analyzer and Johnson Vitros5600 analyzer are shown in Figure 3. The correlation coefficient of the CRP value measured by the 2 instruments was 0.994 (P<0.001) in Pearson correlation analysis. The Passing-Bablok analysis yielded the equation y=1.081(95% CI: 1.063–1.098)x–0.826(95% CI: −1.811–[−0.165]), in which the slope was close to 1 and the intercept was −0.826. The weighted Deming analysis yielded the equation y=1.005(95% CI: 0.982–1.028)x+1.152(95% CI: 0.366–1.957), in which the slope range contained 1 and the intercept was 1.152. In the Bland-Altman analysis, the mean bias between the 2 instruments’ tests was −4.31 mg/L. These results showed that there was a constant deviation in the CRP values measured by Mindray BC-6800 and Johnson Vitros5600.
Figure 3

Consistency analysis and Bland-Altman difference plot of C-reactive protein levels between the Johnson Vitros5600 and Mindray BC-6800. (A) Pearson correlation analysis; (B) Passing-Bablok analysis; (C) weighted Deming analysis; (D) Bland-Altman test.

The comparison results of the Mindray BC-5390 analyzer and Johnson Vitros5600 analyzer are shown in Figure 4. Pearson correlation analysis found that the correlation coefficient of the CRP values measured by the 2 instruments was 0.983 (P<0.001). The Passing-Bablok analysis yielded the equation y=1.088(95% CI: 1.066–1.109)x–0.310(95% CI: −1.068–0.475), in which the slope and intercept were close to 1 and equal to 0. The weighted Deming analysis yielded the equation y=1.025(95% CI: 0.996–1.052)x+1.306(CI: 0.372–2.217), in which the slope range contained 1. The Bland-Altman test indicated that the mean bias between the 2 different instrument tests was −5.66 mg/L. These results showed that the CRP test results between the Mindray BC-5390 and Johnson Vitros5600 were more consistent than those between other instruments.
Figure 4

Consistency analysis and Bland-Altman difference plot of C-reactive protein levels between the Johnson Vitros5600 and Mindray BC-5390. (A) Pearson correlation analysis; (B) Passing-Bablok analysis; (C) weighted Deming analysis; (D) Bland-Altman test.

In addition, there were differences in test cost and test time between the Mindray analyzers and Johnson Vitros5600 analyzer. The use of the Johnson Vitros5600 analyzer inspection costs $6.60/time, while Mindray analyzers (BC-5390 and BC-6800) costs $5.60/time. In terms of length of test time, the Johnson Vitros5600 analyzer takes 9 min/time, while the Mindray analyzer takes 3 min/time. Furthermore, the Johnson Vitros5600 analyzer requires centrifugation for 10 min/time before testing. In summary, the test cost and time of the Johnson Vitros5600 analyzer were higher than that of the Mindray analyzers.

Analysis of the Consistency of BC-5390 and BC-6800 in Routine Blood Indicators

The blood laboratory results from 847 patients were used to compare the consistency of the Mindray BC-5390 and BC-6800 in routine blood indicators. The results of difference analysis showed that there were statistically significant differences in all indicators except RDW-SD (P<0.05). The summary results of the Pearson correlation analysis, Passing-Bablok analysis, and Bland-Altman test indicated that the BC-5390 and BC-6800 had deviations in the measurement of hemoglobin, MCV, MCHC, RDW-CV, RDW-SD, and MPV (Table 1).
Table 1

Consistency of the Mindray BC-5390 and BC-6800 analyzers in routine blood indicators.

VariablesBC-5390BC-6800 P RPassing-BablokBland-Altman
InterceptSlope
WBC, 109/L, M (Q1, Q3)11.02 (8.11, 14.32)10.84 (8.00,13.88)<0.0010.9930.066 (−0.026, 0.160)0.972 (0.963, 0.981)−0.217 (−1.478, 1.045)
Neutrophils, %, mean±SD77.38±10.5277.76±10.31<0.0010.991.782 (1.094, 2.449)0.980 (0.971, 0.988)0.334 (−2.497, 3.165)
Lymphocytes, %, M (Q1, Q3)12.90 (8.50, 18.40)12.40 (8.20, 18.00)<0.0010.989−0.034 (−0.167, 0.099)0.976 (0.966, 0.985)−0.527 (−3.006, 1.953)
Monocytes, %, M (Q1, Q3)6.30 (4.90, 7.80)6.70 (5.20, 8.20)<0.0010.9170.092 (−0.028, 0.300)1.030 (1.000, 1.049)0.277 (−1.713, 2.267)
Eosinophils, %, M (Q1, Q3)0.70 (0.20, 1.50)0.50 (0.10, 1.30)<0.0010.986−0.100 (−0.100, −0.100)1.000 (1.000, 1.000)−0.106 (−0.852, 0.640)
Basophils, %, M (Q1, Q3)0.20 (0.10, 0.30)0.20 (0.10, 0.30)<0.0010.420.000 (0.000, 0.000)1.000 (1.000, 1.000)0.022 (−0.379, 0.423)
Neutrophils, 109/L, M (Q1, Q3)8.65 (5.90, 11.84)8.49 (5.94, 11.42)<0.0010.9940.123 (0.069, 0.190)0.966 (0.958, 0.975)−0.155 (−1.263, 0.953)
Lymphocytes, 109/L, M (Q1, Q3)1.38 (0.95, 1.85)1.32 (0.88, 1.76)<0.0010.982−0.036 (−0.056, −0.017)0.979 (0.964, 0.996)−0.069 (−0.356, 0.218)
Monocytes, 109/L, M (Q1, Q3)0.68 (0.49, 0.92)0.70 (0.50, 0.96)<0.0010.9620.003 (−0.008, 0.020)1.025 (1.000, 1.043)0.018 (−0.172, 0.208)
Eosinophils, 109/L, M (Q1, Q3)0.07 (0.03, 0.15)0.05 (0.01, 0.12)<0.0010.986−0.010 (−0.010, −0.010)1.000 (1.000, 1.000)−0.012 (−0.082, 0.059)
Basophils, 109/L, M (Q1, Q3)0.02 (0.01, 0.03)0.02 (0.01, 0.03)0.0090.3180.000 (0.000, 0.000)1.000 (1.000, 1.000)0.001 (−0.037, 0.039)
RBC, 109/L, mean±SD4.40±0.934.20±0.88<0.0010.99−0.048 (−0.085, −0.005)0.960 (0.950, 0.969)−0.223 (−0.491, 0.046)
Hemoglobin, g/L, mean±SD129.32±27.29126.38±27.60<0.0010.996−4.918 (−5.774, −3.000)1.013 (1.000, 1.020)−3.254 (−8.109, 1.601)
Hematocrit, %, mean±SD37.99±7.6437.28±7.61<0.0010.986−1.100 (−1.502, −0.757)1.000 (0.990, 1.011)−0.928 (−3.445, 1.589)
MCV, fL, mean±SD86.89±7.6689.33±7.820.0000.9832.400 (0.945, 3.321)1.000 (0.990, 1.018)2.333 (−0.337, 5.004)
MCH, pg, mean±SD29.55±3.0530.22±3.10<0.0010.9660.800 (0.202, 0.947)1.000 (0.992, 1.020)0.741 (−0.783, 2.265)
MCHC, g/L, mean±SD339.90±15.15337.89±10.810.0440.733128.276 (113.000, 141.085)0.621 (0.583, 0.667)−0.701 (−20.550, 19.148)
RDW-CV, %, mean±SD13.30±1.8613.41±1.80<0.0010.9581.033 (0.622, 1.450)0.933 (0.900, 0.967)0.107 (−0.825, 1.039)
RDW-SD, fL, mean±SD42.48±6.1842.56±5.420.8640.9557.183 (6.359, 8.102)0.833 (0.811, 0.854)0.010 (−3.384, 3.405)
Platelets, 109/L, M (Q1, Q3)230.00 (180.00, 284.00)225.00 (178.00, 276.00)<0.0010.9792.426 (−0.693, 5.297)0.966 (0.952, 0.980)−4.684 (−42.211, 32.842)
Plateletcrit, %, M (Q1, Q3)0.22 (0.18, 0.27)0.21 (0.17, 0.25)<0.0010.9580.002 (−0.002, 0.006)0.948 (0.929, 0.970)−0.009 (−0.054, 0.036)
MPV, fL, mean±SD9.69±1.189.51±1.29<0.0010.918−1.244 (−1.580, −0.857)1.111 (1.071, 1.148)−0.187 (−1.054, 0.681)
PDW, %, mean±SD15.96±0.7915.98±0.810.0090.8150.000 (0.000, 0.000)1.000 (1.000, 1.000)0.022 (−0.465, 0.510)

R – correlation coefficient; WBC – white blood cells; RBC – red blood cells; MCV – mean corpuscular volume; MCH – mean corpuscular hemoglobin; MCHC – mean corpuscular hemoglobin concentration; RDW-CV – RBC distribution width-coefficient of variation); RDW-SD – RBC distribution width-standard deviation; MPV – mean platelet volume; PDW – platelet distribution width.

The consistency of the indicator direction was evaluated by effect analysis, as shown in Figure 5. The results showed that the distribution of the BC6800 and BC5390 analyzers in the 3 directions of low, normal, and high measured values were in the same direction (on the same side of the dot). The directions of the BC6800 and BC5390 were relatively consistent, indicating that the 2 instruments are interchangeable.
Figure 5

Consistency analysis of indicator direction between the Mindray BC-5390 and Mindray BC-6800.

Discussion

CRP is an acute-phase inflammatory protein, which shows increased expression under inflammatory conditions. The routine determination of CRP in clinical situations is based on analyzers, ELISA, and lateral flow assays [9]. In clinical practice, CRP levels can be measured by different types of instruments, and it is necessary to assess the consistency of the different instruments. This study evaluated the consistency of 3 instruments to measure CRP. The results showed that the CRP values measured by the Mindray BC5390, Mindray BC6800, and Johnson Vitros5600 were different. The CRP measurement results of the Mindray BC5390 analyzer were consistent with those of the Mindray BC6800 and Johnson Vitros5600 analyzers. However, there was a constant deviation in the CRP values measured by the Mindray BC-6800 and Johnson Vitros5600 analyzer. Furthermore, the Mindray analyzers consumed less time and cost in testing than the Johnson Vitros5600 analyzer. For the measurement of routine blood tests, the BC5390 analyzer and BC6800 analyzer were found to be interchangeable. The available evidence indicates that CRP is highly expressed in rheumatoid arthritis, certain cardiovascular diseases and infections, and other inflammatory diseases [6,15,16]. Transcriptional induction of the CRP gene occurs primarily in liver cells in response to increased levels of inflammatory cytokines, such as interleukin-6 [17]. CRP is an acute-phase protein, and therefore, its plasma concentrations change significantly in the presence of inflammatory disease [18,19]. During bacterial infections, the maximum concentration of serum CRP levels can increase by up to 1000 times that of baseline levels [20]. However, when the stimulation ends, the CRP value decreases exponentially for 18 to 20 h [21]. Because of these characteristics, CRP has become an important biomarker in clinical practice [22-25]. There are more than 10 detection methods for CRP, of which turbidimetry-based assays and ELISA are the most widely used in clinical practice [9,26,27]. Clinical chemistry analyzers based on turbidimetric principles are used in large clinical settings, while commercially approved ELISA kits are preferred in dispersed and remote settings [9]. In the present study, the Mindray BC-5390, Mindray BC-6800, and Johnson Vitros5600 analyzers were based on the principle of the turbidimetric method to detect CRP value [10,28]. The difference was that the Mindray analyzers used immunity transmission turbidity, while the Johnson Vitros5600 analyzer used immune scatter turbidity. Different detection systems, reagents, and methodological principles have different detection results for the same item. Our results showed that the CRP measurement results of the Mindray BC-5390 analyzer were consistent with the results of the Mindray BC-6800 and Johnson Vitros5600 analyzers. However, the CRP values measured by the Mindray BC-6800 and Johnson Vitros5600 analyzer had a constant deviation. The difference between the 2 analyzers may have been related to a certain difference in the performance of the different instruments. Furthermore, there are also reports on the use of the Mindray BC-6800 analyzer outside of China, such as in Poland [29], Spain [30], Korea [31], and Italy [32]. In the large-scale clinical practice environment, many factors should be considered in the test instrument, such as the price of the instrument, accuracy, test time, test cost, and the difficulty of sample acquisition. The results of this study indicated that the Mindray analyzers required less time and cost to test CRP values than the Johnson Vitros5600 analyzer. In terms of sample detection, the Mindray analyzers used whole blood for testing, and almost no manual intervention was required, while the Johnson Vitros5600 analyzer used serum, which required blood samples to be processed before measurement. In addition, the Johnson Vitros5600 analyzer (41 μL serum) consumed more blood when measuring CRP, while the Mindray analyzers (20 μL whole blood) used less blood, and can test routine blood values and CRP at the same time. For patients with difficulty in taking a blood sample, such as infants, the Mindray analyzers may be more suitable for detection in these populations.

Conclusions

The consistency of 3 serum CRP analyzers using turbidimetry was analyzed. The measurement of CRP by the Mindray BC5390 analyzer was consistent with that of the Mindray BC6800 and Johnson Vitros5600 analyzers. A constant deviation was observed in the CRP values measured by the Mindray BC-6800 and Johnson Vitros5600 analyzer. The Mindray analyzers required less time and cost in CRP measurement than the Johnson Vitros5600 analyzer. In addition, the BC5390 analyzer and BC6800 analyzer were found to be interchangeable for the measurement of routine blood values.
  5 in total

1.  Mindray BC-6800 haematological analyser: 3D-DIFF scattergram usefulness in infectious mononucleosis diagnosis.

Authors:  Antonio La Gioia; Maurizio Fumi; Fabiana Fiorini; Maria Bombara; Nicola La Gioia; Ylenia Pancione; Silvia Sale; Marcello Fiorini; Vincenzo Rocco
Journal:  Int J Lab Hematol       Date:  2021-04-23       Impact factor: 2.877

2.  Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer.

Authors:  Oscar Fuster; Belinda Andino; Amparo Pardo; Begoña Laiz
Journal:  J Clin Lab Anal       Date:  2017-04-26       Impact factor: 2.352

3.  Performance evaluation of Mindray CAL 8000(BC-6800 and SC-120) hematology analyzer and slidemaker/stainer.

Authors:  Hwan Tae Lee; Pil-Whan Park; Yiel-Hea Seo; Kyung-Hee Kim; Ja Young Seo; Ji-Hun Jeong; Moon Jin Kim; Jeong-Yeal Ahn
Journal:  J Clin Lab Anal       Date:  2016-09-23       Impact factor: 2.352

4.  A Comparison of Mindray BC-6800, Sysmex XN-2000, and Beckman Coulter LH750 Automated Hematology Analyzers: A Pediatric Study.

Authors:  Olga Ciepiela; Iwona Kotuła; Szymon Kierat; Sandra Sieczkowska; Anna Podsiadłowska; Anna Jenczelewska; Karolina Księżarczyk; Urszula Demkow
Journal:  J Clin Lab Anal       Date:  2016-05-17       Impact factor: 2.352

5.  Numerical Modelling of the Optical Properties of Plasmonic and Latex Nanoparticles to Improve the Detection Limit of Immuno-Turbidimetric Assays.

Authors:  Giuliano Coletta; Vincenzo Amendola
Journal:  Nanomaterials (Basel)       Date:  2021-04-28       Impact factor: 5.076

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.