UF-1000i: validation of the body fluid mode for counting cells in body fluids.

Chérina Fleming, Rob Brouwer, Adriaan van Alphen, Jan Lindemans, Robert de Jonge.   

Abstract

BACKGROUND: We evaluated the new body fluid mode on the UF-1000i urinalysis analyzer for counting total white blood cells (WBC) and red blood cells (RBC) in continuous ambulatory peritoneal dialysis (CAPD), ascites and pleural fluids.
METHODS: We collected 154 body fluid samples, and compared the results of the UF-1000i BF mode with the Fuchs-Rosenthal counting chamber and the XN-1000 BF mode. Linearity, carry over and precision were also assessed.
RESULTS: Method comparison results showed acceptable WBC agreement between UF-1000i and chamber (y=1.27x+3.13, n=135, r=0.99) and between UF-1000i and XN (y=1.15x+0.31, n=135, r=1.00). Comparison between the UF-1000i and both comparison methods showed good agreement for RBC counts. Overall results were better when UF-1000i was compared with the XN-1000 than with the Fuchs-Rosenthal chamber. The lower limit of quantitation was defined at 9×106 WBC/L and at 25×106 RBC/L. Linearity for both WBC (r=1.00) and RBC (r=0.99) was good. Carry over was negligible, and it never exceeded 0.01%. In one sample, a high discrepancy was observed between WBC results for both automated analyzers and the counting chamber. This discrepancy was due to interfering factors, such as bacteria and yeast cells, and it led to a false increased WBC count on both automated systems.
CONCLUSIONS: The UF-1000i BF mode offers rapid and reliable total WBC and RBC counts for initial screening of CAPD, ascites and pleural fluid, and can improve the workflow in a routine laboratory; however, when using automated analyzers, the inspection of scattergrams is required to ensure the most accurate results are obtained.

Entities:  

Mesh:

Year:  2014        PMID: 24964259     DOI: 10.1515/cclm-2014-0512

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  7 in total

1.  Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF-1000i.

Authors:  Sabrina Buoro; Sara Apassiti Esposito; MariaGrazia Alessio; Alberto Crippa; Cosimo Ottomano; Giuseppe Lippi
Journal:  J Clin Lab Anal       Date:  2015-08-24       Impact factor: 2.352

2.  Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer.

Authors:  Oscar Fuster; Belinda Andino; Amparo Pardo; Begoña Laiz
Journal:  J Clin Lab Anal       Date:  2017-04-26       Impact factor: 2.352

3.  Automated analysis for differentiating leukocytes in body fluids using the software "biological liquid application" on ADVIA2120/2120i hematology analyzer.

Authors:  Alessandra Falda; Paolo Doretto
Journal:  J Clin Lab Anal       Date:  2018-05-25       Impact factor: 2.352

4.  Evaluation of Sysmex XN-1000 hematology analyzer for cell count and screening of malignant cells of serous cavity effusion.

Authors:  Weiyi Xu; Qian Yu; Lixia Xie; Baode Chen; Ling Zhang
Journal:  Medicine (Baltimore)       Date:  2017-07       Impact factor: 1.889

5.  Validation and Search of the Ideal Cut-Off of the Sysmex UF-1000i® Flow Cytometer for the Diagnosis of Urinary Tract Infection in a Tertiary Hospital in Spain.

Authors:  María I Millán-Lou; Juan M García-Lechuz; María A Ruiz-Andrés; Concepción López; María J Aldea; María J Revillo; Antonio Rezusta
Journal:  Front Med (Lausanne)       Date:  2018-04-09

6.  Guidelines on the management of ascites in cirrhosis.

Authors:  Guruprasad P Aithal; Naaventhan Palaniyappan; Louise China; Suvi Härmälä; Lucia Macken; Jennifer M Ryan; Emilie A Wilkes; Kevin Moore; Joanna A Leithead; Peter C Hayes; Alastair J O'Brien; Sumita Verma
Journal:  Gut       Date:  2020-10-16       Impact factor: 23.059

7.  Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers.

Authors:  Jooyoung Cho; Joowon Oh; Sang Guk Lee; You Hee Lee; Jaewoo Song; Jeong Ho Kim
Journal:  Ann Lab Med       Date:  2020-03       Impact factor: 3.464

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.