Literature DB >> 26547138

Mindray BC-6800 body fluid mode, performance of nucleated cells, and differential count in ascitic and pleural fluids.

S Buoro1, T Mecca1, G Azzarà1, M Seghezzi1, E Candiago2, A Gianatti2, A Crippa1, A La Gioia1.   

Abstract

INTRODUCTION: An accurate and rapid analysis of cells in body fluids (BFs) is important for diagnosis and follow-up in many pathological conditions. We evaluated the analytical performance of the module BF Mindray BC-6800 (BC-6800-BF) for cytometric analysis of ascitic and pleural fluids.
METHODS: A total of 99 ascitic and 45 pleural samples were collected and assessed with BC-6800-BF and optical microscopy. This study also includes the evaluation of limit blank (LoB), limit detection (LoD), limit quantitation, (LoQ), carryover, linearity, and diagnostic concordance between the two methods.
RESULTS: For TC-BF, LoB was 1 × 10(6) cells/L, LoD was 3 × 10(6) cells/L, and LoQ was 4 × 10(6) cells/L. Linearity was excellent (r(2) = 0.99) and carryover was negligible. TC-BF performed with the two methods showed Pearson's correlation of 0.99 (P < 0.0001), Passing-Bablok regression y = 1.04x - 1.17, and bias 33.7 cells. In ascitic fluids, polymorphonuclear cells (PMN) showed an area under curve (AUC) of 0.98 (P < 0.0001). In pleural fluids, mononuclear cells (MN) and PMN % displayed an AUC of 0.79 (P < 0.0001) and 0.93 (P < 0.0001), respectively.
CONCLUSIONS: BC-6800-BF in ascitic and pleural fluids offers rapid and accurate cell and differential counts in clinically relevant concentration ranges. The use of BC-6800-BF may allow to replace routine optical counting, except for samples displaying abnormal cell counts or abnormal DIFF scattergram.
© 2015 John Wiley & Sons Ltd.

Entities:  

Keywords:  Ascitic fluid; Mindray BC-6800; automated cell counting; body fluid; pleural fluid

Mesh:

Substances:

Year:  2015        PMID: 26547138     DOI: 10.1111/ijlh.12446

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer.

Authors:  Oscar Fuster; Belinda Andino; Amparo Pardo; Begoña Laiz
Journal:  J Clin Lab Anal       Date:  2017-04-26       Impact factor: 2.352

2.  Screening for malignant tumor cells in serous effusions with an automatic hematology analyzer using a novel diagnostic algorithm.

Authors:  Dehua Sun; Xinru Mao; Taixue An; Xiaojing He; Kai Qiu; Yuhong Luo; Zheyuan Qin; Yongjian He; Tie Xiong; Houmei Feng; Jin Li; Lei Zheng
Journal:  Ann Transl Med       Date:  2022-03