Variations in pleural fluid WBC count and differential counts with different sample containers and different methods.

Barrett D Conner1, Y C Gary Lee, Paul Branca, Jeffrey T Rogers, R Michael Rodriguez, Richard W Light.   

Abstract

OBJECTIVE: To compare the results of pleural fluid analysis for WBC counts and differential cell counts as follows: (1) counting performed manually vs that performed by an automated cell counter; (2) cells collected in different types of specimen containers; and (3) cell counts performed at 4 and 24 h postthoracentesis.
METHODS: Twenty-eight pleural fluid samples were each collected in five different containers (ie, ethylenediaminetetraacetic acid (EDTA)-treated glass, citrate-treated glass, heparinized glass, plain glass, and plain plastic tubes). The WBC counts and differential cell counts were obtained manually (on the EDTA tube) and with an automated counter on all tubes within 4 h of collection, and again after 24 h of refrigeration.
RESULTS: There was a close correlation between the WBC counts obtained manually and those obtained with the automated counter from the pleural fluid samples collected in the EDTA tubes (r = 0.92). With the automated counter, the pleural fluid WBC counts were similar among the three tubes containing anticoagulants, but the counts obtained from the tubes without anticoagulants were significantly lower. The differential cell counts obtained manually and those obtained with the automated cell counter differed substantially. Although the percentage of lymphocytes was similar, the automated counter was inaccurate in differentiating neutrophils from large monocytes or mesothelial cells. The WBC counts obtained within 4 h of collection and 24 h after collection were virtually identical.
CONCLUSIONS: The WBC counts obtained manually and with the automated counter from pleural fluid samples in EDTA tubes correlated very closely. The pleural fluid WBC count was lower if the pleural fluids had been collected in tubes without an anticoagulant. Automated WBC counts from pleural fluid specimens were inaccurate, possibly due to difficulty in separating neutrophils from monocyte/mesothelial cells. Refrigerated storage for up to 24 h had no significant effect on the total WBC count or on the WBC count differential regardless of the tube utilized.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12684309     DOI: 10.1378/chest.123.4.1181

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  4 in total

1.  Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer.

Authors:  Oscar Fuster; Belinda Andino; Amparo Pardo; Begoña Laiz
Journal:  J Clin Lab Anal       Date:  2017-04-26       Impact factor: 2.352

2.  Collection of cerebrospinal fluid into EDTA versus plain tubes does not affect the standard analysis in dogs.

Authors:  Bodil Cathrine Koch; Lea Ophelia Daniels; Line Tang Thomsen; Michelle Brønniche Møller Nielsen; Mette Berendt; Hanne Gredal
Journal:  Acta Vet Scand       Date:  2019-05-06       Impact factor: 1.695

3.  Evaluation of Use of RPMI Medium to Preserve Cell Morphology for Pleural/Peritoneal Fluid Cytology.

Authors:  Usha Rani; Meeta Singh; Armaan Saith; Shayama L Jain; Anurag Aggarwal; Sunita Aggarwal
Journal:  J Cytol       Date:  2022-02-17       Impact factor: 1.000

4.  Analytical validation of the Sysmex XT-2000iV for cell counts in canine and feline effusions and concordance with cytologic diagnosis.

Authors:  Nazaré Pinto da Cunha; Alessia Giordano; Mario Caniatti; Saverio Paltrinieri
Journal:  Vet Clin Pathol       Date:  2009-02-02       Impact factor: 1.180

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.