IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry.

Yang Zhang1, Lin Zheng1, Ailing Bian1, Qi Zhou2.   

Abstract

PURPOSE: To compare the postoperative intraocular pressure (IOP) of eyes following pars plana vitrectomy (PPV) combined with intravitreal silicone oil (SO) tamponade by Corneal Visualization Scheimpflug Technology (CST), Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT).
METHODS: Thirty-eight participants who had undergone PPV combined with SO tamponade to treat vitreoretinal diseases were enrolled. Postoperative IOP measurements were obtained using CST, NCT and GAT. Inter-device agreement was assessed by Bland-Altman analysis. The correlation coefficient was used to describe the potential postoperative factors affecting the postoperative IOP differences between each device.
RESULTS: Bland-Altman analysis revealed the bias between CST and GAT, between CST and NCT, and between GAT and NCT to be -0.2, 2.1 and 2.4 mmHg, respectively. CST and GAT correlated well with each other. NCT values were lower than those of GAT and CST (all p < 0.05), whereas CST values did not differ from the GAT readings. Central corneal thickness, corneal biomechanical properties and age showed significant correlation with the differences of CST-NCT and GAT-NCT.
CONCLUSIONS: In SO tamponade eyes, NCT obtains lower IOP than other tonometry techniques, and CST is highly consistent with GAT. CST offers an optional non-contact method for measuring postoperative IOP in SO tamponade eyes.

Entities:  

Keywords:  Corneal biomechanical properties; Corvis ST tonometer; Goldmann applanation tonometry; Intraocular pressure; Silicone oil tamponade

Mesh:

Substances:

Year:  2017        PMID: 28444523     DOI: 10.1007/s10792-017-0519-1

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  18 in total

1.  In vivo laser scanning confocal microscopy of the cornea in patients with silicone oil tamponade after vitreoretinal surgery.

Authors:  Qihua Le; Xin Wang; Jiahua Lv; Xinghuai Sun; Jianjiang Xu
Journal:  Cornea       Date:  2012-08       Impact factor: 2.651

2.  Time course of corneal biomechanical parameters after phacoemulsification with intraocular lens implantation.

Authors:  Kazutaka Kamiya; Kimiya Shimizu; Fumiko Ohmoto; Rie Amano
Journal:  Cornea       Date:  2010-11       Impact factor: 2.651

3.  Evaluation of tonometric correction factors.

Authors:  Pinakin Gunvant; Daniel James O'Leary; Mani Baskaran; David Charles Broadway; Russell Julian Watkins; Lingam Vijaya
Journal:  J Glaucoma       Date:  2005-10       Impact factor: 2.503

4.  Repeatability of ocular biomechanical data measurements with a Scheimpflug-based noncontact device on normal corneas.

Authors:  Gabor Nemeth; Ziad Hassan; Adrienne Csutak; Eszter Szalai; Andras Berta; Laszlo Modis
Journal:  J Refract Surg       Date:  2013-08       Impact factor: 3.573

5.  Sutureless 23-gauge versus 20-gauge vitrectomy with silicone oil injection in rhegmatogenous retinal detachment.

Authors:  Raja Narayanan; Nandkishore Tibra; Annie Mathai; Jay Chhablani; Baruch D Kuppermann
Journal:  Retina       Date:  2012-05       Impact factor: 4.256

Review 6.  Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.

Authors:  M J Doughty; M L Zaman
Journal:  Surv Ophthalmol       Date:  2000 Mar-Apr       Impact factor: 6.048

7.  An ultra-high-speed Scheimpflug camera for evaluation of corneal deformation response and its impact on IOP measurement.

Authors:  Christopher Kai-Shun Leung; Cong Ye; Robert N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-04-23       Impact factor: 4.799

8.  Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma.

Authors:  Lukas Reznicek; Daniel Muth; Anselm Kampik; Aljoscha S Neubauer; Christoph Hirneiss
Journal:  Br J Ophthalmol       Date:  2013-08-22       Impact factor: 4.638

9.  [Measurement of viscoelastic corneal parameters (corneal hysteresis) in patients with primary open angle glaucoma].

Authors:  B Schroeder; A Hager; A Kutschan; W Wiegand
Journal:  Ophthalmologe       Date:  2008-10       Impact factor: 1.059

Review 10.  Silicone oil induced glaucoma: a review.

Authors:  Parul Ichhpujani; Anjana Jindal; L Jay Katz
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-14       Impact factor: 3.117

View more
  2 in total

1.  Intraocular Pressure Measurement Using Ocular Response Analyzer, Dynamic Contour Tonometer, and Scheimpflug Analyzer Corvis ST.

Authors:  Lisa Ramm; Robert Herber; Eberhard Spoerl; Frederik Raiskup; Lutz E Pillunat; Naim Terai
Journal:  J Ophthalmol       Date:  2019-10-16       Impact factor: 1.909

Review 2.  Corneal Vibrations during Intraocular Pressure Measurement with an Air-Puff Method.

Authors:  Robert Koprowski; Sławomir Wilczyński
Journal:  J Healthc Eng       Date:  2018-02-11       Impact factor: 2.682

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.