Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine : An in vitro study.

F Elkholy1, B Mikhaiel2, F Schmidt3, B G Lapatki3.   

Abstract

INTRODUCTION: The six force-moment (F/M) components exerted by aligners of different thickness during simulated mesiorotation and distorotation of a mandibular canine were studied.
MATERIALS AND METHODS: An acrylic mandibular model with a separated right canine mounted on a hexapod via a 3D F/M sensor was used. Duran+® aligners (Scheu Dental, Germany) of thickness 0.5, 0.625, and 0.75 mm were fabricated on plaster models with the measurement tooth in its neutral position. The F/M values were recorded during progressive mesiorotation or distorotation of tooth 43 in 1° steps up to ±15°, corresponding to 0.5 mm displacements of the tooth's interdental contacts. Each rotation step included renewed seating of the aligner on the acrylic model. Three aligners were tested three times each for each thickness and direction of rotation.
RESULTS: The median rotational moments for the 0.5 mm aligner and 15° distorotation of tooth 43 was 27.49 Nmm (interquartile range, IQR 1.45 Nmm). The corresponding values for the 0.625 and 0.75 mm aligners were 41.04 Nmm (IQR 5.62 Nmm) and 42.48 Nmm (IQR 2.17 Nmm), respectively. The average rotational moments for distorotation were 15% higher than for mesiorotation (p = 0.01). Relatively high collateral F/M components, specifically an intrusive force and labiolingual and mesiodistal tipping moments, were observed.
CONCLUSION: To avoid overloading of periodontal structures, derotation of lower canines should be limited to 10° per setup step, leading to rotational moments of about 15 Nmm. The mechanical behavior of the 0.625 and 0.75 mm aligners were similar; thus, it may be omitted from the aligner sequence. Further studies are required to investigate specific aligner modifications or attachments for minimizing collateral F/M components or unwanted movements, respectively, during canine derotation.

Entities:  

Keywords:  Aligner; Canine derotation; Force; Moment; Orthodontic appliances; Thermoplastic appliance

Mesh:

Year:  2017        PMID: 28357471     DOI: 10.1007/s00056-017-0090-4

Source DB:  PubMed          Journal:  J Orofac Orthop        ISSN: 1434-5293            Impact factor:   1.938


  11 in total

1.  Torquing an upper central incisor with aligners--acting forces and biomechanical principles.

Authors:  Wolfram Hahn; Antonia Zapf; Henning Dathe; Julia Fialka-Fricke; Susanne Fricke-Zech; Rudolf Gruber; Dietmar Kubein-Meesenburg; Reza Sadat-Khonsari
Journal:  Eur J Orthod       Date:  2010-05-12       Impact factor: 3.075

2.  Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation.

Authors:  Fayez Elkholy; Thanapon Panchaphongsaphak; Fatih Kilic; Falko Schmidt; Bernd G Lapatki
Journal:  J Orofac Orthop       Date:  2015-11       Impact factor: 1.938

Review 3.  Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review.

Authors:  Gabriele Rossini; Simone Parrini; Tommaso Castroflorio; Andrea Deregibus; Cesare L Debernardi
Journal:  Angle Orthod       Date:  2014-11-20       Impact factor: 2.079

4.  Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: An in-vitro study.

Authors:  Fayez Elkholy; Falko Schmidt; Rudolf Jäger; Bernd G Lapatki
Journal:  Am J Orthod Dentofacial Orthop       Date:  2017-02       Impact factor: 2.650

5.  Stress relaxation properties of four orthodontic aligner materials: A 24-hour in vitro study.

Authors:  Luca Lombardo; Elisa Martines; Valentina Mazzanti; Angela Arreghini; Francesco Mollica; Giuseppe Siciliani
Journal:  Angle Orthod       Date:  2016-06-17       Impact factor: 2.079

6.  Forces and moments delivered by novel, thinner PET-G aligners during labiopalatal bodily movement of a maxillary central incisor: An in vitro study.

Authors:  Fayez Elkholy; Falko Schmidt; Rudolf Jäger; Bernd G Lapatki
Journal:  Angle Orthod       Date:  2016-05-25       Impact factor: 2.079

7.  The influence of occlusal forces on force delivery properties of aligners during rotation of an upper central incisor.

Authors:  Wolfram Hahn; Benjamin Engelke; Klaus Jung; Henning Dathe; Franz-Joseph Kramer; Tina Rödig; Dietmar Kubein-Meesenburg; Rudolf M Gruber
Journal:  Angle Orthod       Date:  2011-05-25       Impact factor: 2.079

8.  Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization.

Authors:  Mareike Simon; Ludger Keilig; Jörg Schwarze; Britta A Jung; Christoph Bourauel
Journal:  Am J Orthod Dentofacial Orthop       Date:  2014-06       Impact factor: 2.650

9.  Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor.

Authors:  Wolfram Hahn; Benjamin Engelke; Klaus Jung; Henning Dathe; Julia Fialka-Fricke; Dietmar Kubein-Meesenburg; Reza Sadat-Khonsari
Journal:  Angle Orthod       Date:  2010-03       Impact factor: 2.079

10.  Changes in force associated with the amount of aligner activation and lingual bodily movement of the maxillary central incisor.

Authors:  Xiaowei Li; Chaochao Ren; Zheyao Wang; Pai Zhao; Hongmei Wang; Yuxing Bai
Journal:  Korean J Orthod       Date:  2016-03-18       Impact factor: 1.372

View more
  11 in total

1.  Micro computed tomography evaluation of Invisalign aligner thickness homogeneity.

Authors:  Edoardo Mantovani; Simone Parrini; Emanuele Coda; Giovanni Cugliari; Nicola Scotti; Damiano Pasqualini; Andrea Deregibus; Tommaso Castroflorio
Journal:  Angle Orthod       Date:  2021-05-01       Impact factor: 2.079

2.  Scanning electron microscopy evaluation of aligner fit on teeth.

Authors:  Edoardo Mantovani; Enrico Castroflorio; Gabriele Rossini; Francesco Garino; Giovanni Cugliari; Andrea Deregibus; Tommaso Castroflorio
Journal:  Angle Orthod       Date:  2018-06-18       Impact factor: 2.079

3.  Force changes associated with different intrusion strategies for deep-bite correction by clear aligners.

Authors:  Yang Liu; Wei Hu
Journal:  Angle Orthod       Date:  2018-07-23       Impact factor: 2.079

4.  Clear aligner orthodontic therapy of rotated mandibular round-shaped teeth: A finite element study.

Authors:  Andrea Cortona; Gabriele Rossini; Simone Parrini; Andrea Deregibus; Tommaso Castroflorio
Journal:  Angle Orthod       Date:  2019-08-30       Impact factor: 2.079

5.  Effect of different attachment geometries on the mechanical load exerted by PET‑G aligners during derotation of mandibular canines : An in vitro study.

Authors:  F Elkholy; B Mikhaiel; S Repky; F Schmidt; B G Lapatki
Journal:  J Orofac Orthop       Date:  2019-10-08       Impact factor: 1.938

6.  Force decay of polyethylene terephthalate glycol aligner materials during simulation of typical clinical loading/unloading scenarios.

Authors:  Fayez Elkholy; Silva Schmidt; Falko Schmidt; Masoud Amirkhani; Bernd G Lapatki
Journal:  J Orofac Orthop       Date:  2021-12-09       Impact factor: 1.938

7.  Mechanical Characterization of Thermoplastic Aligner Materials: Recommendations for Test Parameter Standardization.

Authors:  F Elkholy; S Schmidt; M Amirkhani; F Schmidt; B G Lapatki
Journal:  J Healthc Eng       Date:  2019-05-29       Impact factor: 2.682

Review 8.  Clear Aligners: Between Evolution and Efficiency-A Scoping Review.

Authors:  Alessandra Putrino; Ersilia Barbato; Gabriella Galluccio
Journal:  Int J Environ Res Public Health       Date:  2021-03-11       Impact factor: 3.390

9.  Influence of a bleaching agent on surface and mechanical properties of orthodontic thermoplastic retainer materials : An in vitro study.

Authors:  Neda Babanouri; Nazanin Ahmadi; Hamid Reza Pakshir; Shabnam Ajami; Raha Habibagahi
Journal:  J Orofac Orthop       Date:  2021-06-25       Impact factor: 2.341

10.  Thickness of orthodontic clear aligners after thermoforming and after 10 days of intraoral exposure: a prospective clinical study.

Authors:  Rosaria Bucci; Roberto Rongo; Carmine Levatè; Ambrosina Michelotti; Sandro Barone; Armando Viviano Razionale; Vincenzo D'Antò
Journal:  Prog Orthod       Date:  2019-09-09       Impact factor: 2.750

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.