Scanning electron microscopy evaluation of aligner fit on teeth.

Edoardo Mantovani, Enrico Castroflorio, Gabriele Rossini, Francesco Garino, Giovanni Cugliari, Andrea Deregibus, Tommaso Castroflorio.   

Abstract

OBJECTIVES: The fitting of aligners on anchorage teeth is a crucial factor in clear aligner orthodontics. The purpose of this experimental study was to evaluate the fitting of two aligner systems, Invisalign and CA-Clear Aligner, using scanning electron microscopy (SEM).
MATERIALS AND METHODS: Passive aligners (Invisalign and CA-Clear Aligner) were adapted on resin casts obtained by stereolithography (STL) files of a patient, and then sectioned buccolingually. Upper and lower central incisors, upper and lower first premolars, and upper and lower first molars were the regions analyzed. Representative microphotographs of sections were taken with a scanning electron microscope (SEM); a total of 160 micrometric measurements were obtained and analyzed with ANOVA tests.
RESULTS: Invisalign provided an overall better fit on lower incisors ( F = 11.48, P = .0095) and on lower molars ( F = 19.93, P = .0012). Considering the different regions, Invisalign provided better fit at the gingival edge of the buccal aspect on lower incisors ( F = 11.33, P = 0.0056) and at the gingival edge of the lingual aspect on upper premolars ( F =5.34, P = 0.0047). On the upper molars, Invisalign provided better fit at the gingival edge of the buccal aspect, while CA-Clear Aligner provided better fit at the buccal maximum convexity, on the buccal cusp, on the occlusal groove and at the palatal cusp. On lower molars, Invisalign showed a more accurate fit at the buccal aspect points.
CONCLUSIONS: Invisalign and CA-Clear Aligner exhibited comparable fit on anchorage teeth. Invisalign provided better fit at the gingival edges of aligners, while the CA-Clear Aligner provided better fit on complex occlusal surfaces.

Entities:  

Keywords:  Aligners; CA-Clear Aligner; Fitting; Invisalign; Microscopy; SEM

Mesh:

Year:  2018        PMID: 29911907      PMCID: PMC8183126          DOI: 10.2319/120417-827.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  18 in total

1.  Torquing an upper central incisor with aligners--acting forces and biomechanical principles.

Authors:  Wolfram Hahn; Antonia Zapf; Henning Dathe; Julia Fialka-Fricke; Susanne Fricke-Zech; Rudolf Gruber; Dietmar Kubein-Meesenburg; Reza Sadat-Khonsari
Journal:  Eur J Orthod       Date:  2010-05-12       Impact factor: 3.075

2.  Bone Tissue Engineering with Multilayered Scaffolds-Part I: An Approach for Vascularizing Engineered Constructs In Vivo.

Authors:  Binulal Nelson Sathy; Ullas Mony; Deepthy Menon; V K Baskaran; Antonios G Mikos; Shantikumar Nair
Journal:  Tissue Eng Part A       Date:  2015-10       Impact factor: 3.845

Review 3.  Diagnostic accuracy and measurement sensitivity of digital models for orthodontic purposes: A systematic review.

Authors:  Gabriele Rossini; Simone Parrini; Tommaso Castroflorio; Andrea Deregibus; Cesare L Debernardi
Journal:  Am J Orthod Dentofacial Orthop       Date:  2016-02       Impact factor: 2.650

Review 4.  Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review.

Authors:  Gabriele Rossini; Simone Parrini; Tommaso Castroflorio; Andrea Deregibus; Cesare L Debernardi
Journal:  Angle Orthod       Date:  2014-11-20       Impact factor: 2.079

5.  A Three-Dimensional Finite Element Analysis of Upper-Canine Distalization with Clear Aligners, Composite Attachments, and Class II Elastics.

Authors:  Benedetta Comba; Simone Parrini; Gabriele Rossini; Tommaso Castroflorio; Andrea Deregibus
Journal:  J Clin Orthod       Date:  2017-01

6.  Effects of variable attachment shapes and aligner material on aligner retention.

Authors:  Hiltrud Dasy; Andreas Dasy; Greg Asatrian; Noémi Rózsa; Hao-Fu Lee; Jin Hee Kwak
Journal:  Angle Orthod       Date:  2015-11       Impact factor: 2.079

7.  The iTero intraoral scanner in Invisalign treatment: a two-year report.

Authors:  Francesco Garino; G Battista Garino; Tommaso Castroflorio
Journal:  J Clin Orthod       Date:  2014-02

8.  Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine : An in vitro study.

Authors:  F Elkholy; B Mikhaiel; F Schmidt; B G Lapatki
Journal:  J Orofac Orthop       Date:  2017-09       Impact factor: 1.938

9.  A randomized clinical trial comparing mandibular incisor proclination produced by fixed labial appliances and clear aligners.

Authors:  Joe Hennessy; Thérèse Garvey; Ebrahim A Al-Awadhi
Journal:  Angle Orthod       Date:  2016-09       Impact factor: 2.079

10.  Effects of mechanical properties of thermoplastic materials on the initial force of thermoplastic appliances.

Authors:  Naohisa Kohda; Masahiro Iijima; Takeshi Muguruma; William A Brantley; Karamdeep S Ahluwalia; Itaru Mizoguchi
Journal:  Angle Orthod       Date:  2012-10-04       Impact factor: 2.079

View more
  9 in total

1.  Micro computed tomography evaluation of Invisalign aligner thickness homogeneity.

Authors:  Edoardo Mantovani; Simone Parrini; Emanuele Coda; Giovanni Cugliari; Nicola Scotti; Damiano Pasqualini; Andrea Deregibus; Tommaso Castroflorio
Journal:  Angle Orthod       Date:  2021-05-01       Impact factor: 2.079

2.  Adaptational changes in clear aligner fit with time.

Authors:  Amal I Linjawi; Amal M Abushal
Journal:  Angle Orthod       Date:  2022-03-01       Impact factor: 2.079

3.  Three-dimensional analysis of aligner gaps and thickness distributions, using hard x-ray tomography with micrometer resolution.

Authors:  Rémi Ammann; Christine Tanner; Georg Schulz; Bekim Osmani; Prasad Nalabothu; Tino Töpper; Bert Müller
Journal:  J Med Imaging (Bellingham)       Date:  2022-06-16

4.  Accuracy of 3D Tooth Movements in the Fabrication of Manual Setup Models for Aligner Therapy.

Authors:  Hisham Sabbagh; Sebastian Marcus Heger; Thomas Stocker; Uwe Baumert; Andrea Wichelhaus; Lea Hoffmann
Journal:  Materials (Basel)       Date:  2022-05-28       Impact factor: 3.748

5.  Risk Factors of Composite Attachment Loss in Orthodontic Patients during Orthodontic Clear Aligner Therapy: A Prospective Study.

Authors:  Chen Yaosen; A M Mohamed; Wang Jinbo; Zheng Ziwei; Maher Al-Balaa; Yang Yan
Journal:  Biomed Res Int       Date:  2021-01-23       Impact factor: 3.411

6.  Micro-computed tomography evaluation of general trends in aligner thickness and gap width after thermoforming procedures involving six commercial clear aligners: An in vitro study.

Authors:  Mario Palone; Mattia Longo; Niki Arveda; Michele Nacucchi; Fabio De Pascalis; Giorgio Alfredo Spedicato; Giuseppe Siciliani; Luca Lombardo
Journal:  Korean J Orthod       Date:  2021-03-25       Impact factor: 1.372

7.  Thermo-mechanical properties of 3D printed photocurable shape memory resin for clear aligners.

Authors:  Se Yeon Lee; Hoon Kim; Hyun-Joong Kim; Chooryung J Chung; Yoon Jeong Choi; Su-Jung Kim; Jung-Yul Cha
Journal:  Sci Rep       Date:  2022-04-15       Impact factor: 4.996

8.  Comparison of dimensional accuracy between direct-printed and thermoformed aligners.

Authors:  Nickolas Koenig; Jin-Young Choi; Julie McCray; Andrew Hayes; Patricia Schneider; Ki Beom Kim
Journal:  Korean J Orthod       Date:  2022-04-22       Impact factor: 1.361

9.  SEM characterization and ageing analysis on two generation of invisible aligners.

Authors:  Roberta Condò; Gianluca Mampieri; Aldo Giancotti; Loredana Cerroni; Guido Pasquantonio; Andrea Divizia; Annalisa Convertino; Barbara Mecheri; Luca Maiolo
Journal:  BMC Oral Health       Date:  2021-06-23       Impact factor: 2.757

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.