Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization.

Mareike Simon1, Ludger Keilig2, Jörg Schwarze3, Britta A Jung4, Christoph Bourauel5.   

Abstract

INTRODUCTION: The exact force systems as well as their progressions generated by removable thermoplastic appliances have not been investigated. Thus, the purposes of this experimental study were to quantify the forces and moments delivered by a single aligner and a series of aligners (Invisalign; Align Technology, Santa Clara, Calif) and to investigate the influence of attachments and power ridges on the force transfer.
METHODS: We studied 970 aligners of the Invisalign system (60 series of aligners). The aligners came from 30 consecutive patients, of which 3 tooth movements (incisor torque, premolar derotation, molar distalization) with 20 movements each were analyzed. The 3 movement groups were subdivided so that 10 movements were supported with an attachment and 10 were not. The patients' ClinCheck (Align Technology, Santa Clara, Calif) was planned so that the movements to be investigated were performed in isolation in the respective quadrant. Resin replicas of the patients' intraoral situation before the start of the investigated movement were taken and mounted in a biomechanical measurement system. An aligner was put on the model, the force systems were measured, and the calculated movements were experimentally performed until no further forces or moments were generated. Subsequently, the next aligners were installed, and the measurements were repeated.
RESULTS: The initial mean moments were about 7.3 N·mm for maxillary incisor torque and about 1.0 N for distalization. Significant differences in the generated moments were measured in the premolar derotation group, whether they were supported with an attachment (8.8 N·mm) or not (1.2 N·mm). All measurements showed an exponential force change.
CONCLUSIONS: Apart from a few maximal initial force systems, the forces and moments generated by aligners of the Invisalign system are within the range of orthodontic forces. The force change is exponential while a patient is wearing removable thermoplastic appliances.
Copyright © 2014 American Association of Orthodontists. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 24880843     DOI: 10.1016/j.ajodo.2014.03.015

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  32 in total

1.  Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation.

Authors:  Fayez Elkholy; Thanapon Panchaphongsaphak; Fatih Kilic; Falko Schmidt; Bernd G Lapatki
Journal:  J Orofac Orthop       Date:  2015-11       Impact factor: 1.938

2.  Force changes associated with different intrusion strategies for deep-bite correction by clear aligners.

Authors:  Yang Liu; Wei Hu
Journal:  Angle Orthod       Date:  2018-07-23       Impact factor: 2.079

3.  Clear aligner orthodontic therapy of rotated mandibular round-shaped teeth: A finite element study.

Authors:  Andrea Cortona; Gabriele Rossini; Simone Parrini; Andrea Deregibus; Tommaso Castroflorio
Journal:  Angle Orthod       Date:  2019-08-30       Impact factor: 2.079

4.  Effect of in vitro aging by water immersion and thermocycling on the mechanical properties of PETG aligner material.

Authors:  Benjamin A Ihssen; Jan H Willmann; Amr Nimer; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2019-10-02       Impact factor: 1.938

5.  Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine : An in vitro study.

Authors:  F Elkholy; B Mikhaiel; F Schmidt; B G Lapatki
Journal:  J Orofac Orthop       Date:  2017-09       Impact factor: 1.938

6.  Comparison of cytokine level changes in gingival crevicular fluid between the aligner and pendulum appliance during early molar distalization : A single-center, prospective, observational study.

Authors:  Huijuan Chen; Liuhui Liu; Yuan Li; Lingyun Guo; Dongmei Sun
Journal:  J Orofac Orthop       Date:  2021-10-22       Impact factor: 1.938

7.  Variation of the modulus of elasticity of aligner foil sheet materials due to thermoforming.

Authors:  Bijan Golkhani; Anna Weber; Ludger Keilig; Susanne Reimann; Christoph Bourauel
Journal:  J Orofac Orthop       Date:  2021-08-19       Impact factor: 2.341

8.  Staging Orthodontic Aligners for Complex Orthodontic Tooth Movement.

Authors:  Shivam Mehta; Dolly Patel; Sumit Yadav
Journal:  Turk J Orthod       Date:  2021-09

9.  Effectiveness of Invisalign® aligners in the treatment of severe gingival recession: A case report.

Authors:  Marcio Antonio de Figueiredo; Fábio Lourenço Romano; Murilo Fernando Neuppmann Feres; Maria Bernadete Sasso Stuani; Ana Carla Raphaelli Nahás-Scocate; Mírian Aiko Nakane Matsumoto
Journal:  Korean J Orthod       Date:  2021-07-25       Impact factor: 1.372

10.  SEM characterization and ageing analysis on two generation of invisible aligners.

Authors:  Roberta Condò; Gianluca Mampieri; Aldo Giancotti; Loredana Cerroni; Guido Pasquantonio; Andrea Divizia; Annalisa Convertino; Barbara Mecheri; Luca Maiolo
Journal:  BMC Oral Health       Date:  2021-06-23       Impact factor: 2.757

View more