The influence of occlusal forces on force delivery properties of aligners during rotation of an upper central incisor.

Wolfram Hahn1, Benjamin Engelke, Klaus Jung, Henning Dathe, Franz-Joseph Kramer, Tina Rödig, Dietmar Kubein-Meesenburg, Rudolf M Gruber.   

Abstract

OBJECTIVE: To determine the forces and moments delivered to a maxillary central incisor during rotation with aligners when a simulated occlusal force generated during swallowing acts on the appliance.
MATERIALS AND METHODS: Five identical appliances were manufactured from four different starting materials (Erkodur 0.8 mm and 1.0 mm; Biolon 0.75 mm and 1.0 mm). An upper central incisor fixed in a measuring device was rotated around its central axis in 0.5-degree steps up to ±10 degrees with the appliance fixed in place. An occlusal force of 30 N generated during swallowing was simulated with a weight positioned on the appliance. For statistical analysis, the moments Tz (rotation) and forces Fz (intrusion) at a deflection of ±0.34 mm to the incisor edge (±5 degrees rotation) were tested. Means and standard deviations for Tz and median and 25% and 75% quartiles for Fz were calculated. An analysis of variance was performed.
RESULTS: The simulated occlusal force increased the measured intrusive force Fz (maximum with a weight, -3.7 N [-3.7, -2.4]; minimum without a weight, -1.3 N [-1.4, -1.1]) and the rotary moment Tz (maximum with a weight, -50.8 Nmm [±0.8]; minimum without a weight, 18.2 Nmm [±0.9]) significantly in all cases (P < .01). This was found for all materials measured and for both directions of rotation.
CONCLUSION: During rotation with aligners, a simulated occlusal force increases the intrusive force and the rotary moment. The biological adverse side effects of these phenomena remain unclear, especially in patients with periods of bruxism.

Entities:  

Mesh:

Year:  2011        PMID: 21612314      PMCID: PMC8903843          DOI: 10.2319/013111-62.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  20 in total

1.  THE FREQUENCY OF DEGLUTITION IN MAN.

Authors:  C S LEAR; J B FLANAGAN; C F MOORREES
Journal:  Arch Oral Biol       Date:  1965 Jan-Feb       Impact factor: 2.633

2.  Torquing an upper central incisor with aligners--acting forces and biomechanical principles.

Authors:  Wolfram Hahn; Antonia Zapf; Henning Dathe; Julia Fialka-Fricke; Susanne Fricke-Zech; Rudolf Gruber; Dietmar Kubein-Meesenburg; Reza Sadat-Khonsari
Journal:  Eur J Orthod       Date:  2010-05-12       Impact factor: 3.075

3.  Initial forces generated by three types of thermoplastic appliances on an upper central incisor during tipping.

Authors:  Wolfram Hahn; Julia Fialka-Fricke; Henning Dathe; Susanne Fricke-Zech; Antonia Zapf; Rudolf Gruber; Dietmar Kubein-Meesenburg; Reza Sadat-Khonsari
Journal:  Eur J Orthod       Date:  2009-06-12       Impact factor: 3.075

4.  Occlusal contacts: vectorial analysis of forces transmitted to temporomandibular joint and teeth.

Authors:  J dos Santos Júnior; W G de Rijk
Journal:  Cranio       Date:  1993-04       Impact factor: 2.020

5.  Cellular response to orthodontic force.

Authors:  W E Roberts; W C Goodwin; S R Heiner
Journal:  Dent Clin North Am       Date:  1981-01

6.  Biomechanical principles of tooth movement.

Authors:  D K Yoshikawa
Journal:  Dent Clin North Am       Date:  1981-01

7.  Occlusal forces in normal- and long-face adults.

Authors:  W R Proffit; H W Fields; W L Nixon
Journal:  J Dent Res       Date:  1983-05       Impact factor: 6.116

8.  Esthetic orthodontic treatment using the invisalign appliance for moderate to complex malocclusions.

Authors:  Robert L Boyd
Journal:  J Dent Educ       Date:  2008-08       Impact factor: 2.264

9.  Physical properties of root cementum: Part 10. Comparison of the effects of invisible removable thermoplastic appliances with light and heavy orthodontic forces on premolar cementum. A microcomputed-tomography study.

Authors:  Laura J Barbagallo; Allan S Jones; Peter Petocz; M Ali Darendeliler
Journal:  Am J Orthod Dentofacial Orthop       Date:  2008-02       Impact factor: 2.650

10.  Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor.

Authors:  Wolfram Hahn; Benjamin Engelke; Klaus Jung; Henning Dathe; Julia Fialka-Fricke; Dietmar Kubein-Meesenburg; Reza Sadat-Khonsari
Journal:  Angle Orthod       Date:  2010-03       Impact factor: 2.079

View more
  9 in total

1.  Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation.

Authors:  Fayez Elkholy; Thanapon Panchaphongsaphak; Fatih Kilic; Falko Schmidt; Bernd G Lapatki
Journal:  J Orofac Orthop       Date:  2015-11       Impact factor: 1.938

2.  Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine : An in vitro study.

Authors:  F Elkholy; B Mikhaiel; F Schmidt; B G Lapatki
Journal:  J Orofac Orthop       Date:  2017-09       Impact factor: 1.938

3.  Force decay of polyethylene terephthalate glycol aligner materials during simulation of typical clinical loading/unloading scenarios.

Authors:  Fayez Elkholy; Silva Schmidt; Falko Schmidt; Masoud Amirkhani; Bernd G Lapatki
Journal:  J Orofac Orthop       Date:  2021-12-09       Impact factor: 1.938

4.  A comparative study of two different clear aligner systems.

Authors:  Federica Ercoli; Michele Tepedino; Vincenzo Parziale; Cesare Luzi
Journal:  Prog Orthod       Date:  2014-05-02       Impact factor: 2.750

5.  Removable thermoplastic appliances modified by incisal cuts show altered biomechanical properties during tipping of a maxillary central incisor.

Authors:  Phillipp Brockmeyer; Katharina Kramer; Florian Böhrnsen; Rudolf Matthias Gruber; Sarah Batschkus; Tina Rödig; Wolfram Hahn
Journal:  Prog Orthod       Date:  2017-08-28       Impact factor: 2.750

6.  Efficacy of self-designed intraoral appliances in prevention of cheek, lip and tongue bite after local anesthesia administration in pediatric patients.

Authors:  Wala A Alghamidi; Sondos B Alghamdi; Jawaher-Ahmad Assiri; Abeer A Almathami; Zuhair-Motlak Alkahtani; Rafi A Togoo
Journal:  J Clin Exp Dent       Date:  2019-04-01

7.  Mechanical Characterization of Thermoplastic Aligner Materials: Recommendations for Test Parameter Standardization.

Authors:  F Elkholy; S Schmidt; M Amirkhani; F Schmidt; B G Lapatki
Journal:  J Healthc Eng       Date:  2019-05-29       Impact factor: 2.682

8.  Force delivery modification of removable thermoplastic appliances using Hilliard precision thermopliers for tipping an upper central incisor.

Authors:  Bernhard Wiechens; Phillipp Brockmeyer; Teresa Erfurth-Jach; Wolfram Hahn
Journal:  Clin Oral Investig       Date:  2022-05-31       Impact factor: 3.606

9.  Mechanical evaluation of newly developed mouthpiece using polyethylene terephthalate glycol for transoral robotic surgery.

Authors:  Kazunori Fujiwara; Takahiro Fukuhara; Koji Niimi; Takahiro Sato; Hideyuki Kataoka; Hiroya Kitano; Hiromi Takeuchi
Journal:  J Robot Surg       Date:  2015-11-04
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.