k-t accelerated aortic 4D flow MRI in under two minutes: Feasibility and impact of resolution, k-space sampling patterns, and respiratory navigator gating on hemodynamic measurements.

Emilie Bollache1, Alex J Barker1, Ryan Scott Dolan1, James C Carr1, Pim van Ooij2, Rouzbeh Ahmadian1, Alex Powell1, Jeremy D Collins1, Julia Geiger1,3, Michael Markl1,4.   

Abstract

PURPOSE: To assess the performance of highly accelerated free-breathing aortic four-dimensional (4D) flow MRI acquired in under 2 minutes compared to conventional respiratory gated 4D flow.
METHODS: Eight k-t accelerated nongated 4D flow MRI (parallel MRI with extended and averaged generalized autocalibrating partially parallel acquisition kernels [PEAK GRAPPA], R = 5, TRes = 67.2 ms) using four ky -kz Cartesian sampling patterns (linear, center-out, out-center-out, random) and two spatial resolutions (SRes1 = 3.5 × 2.3 × 2.6 mm3 , SRes2 = 4.5 × 2.3 × 2.6 mm3 ) were compared in vitro (aortic coarctation flow phantom) and in 10 healthy volunteers, to conventional 4D flow (16 mm-navigator acceptance window; R = 2; TRes = 39.2 ms; SRes = 3.2 × 2.3 × 2.4 mm3 ). The best k-t accelerated approach was further assessed in 10 patients with aortic disease.
RESULTS: The k-t accelerated in vitro aortic peak flow (Qmax), net flow (Qnet), and peak velocity (Vmax) were lower than conventional 4D flow indices by ≤4.7%, ≤ 11%, and ≤22%, respectively. In vivo k-t accelerated acquisitions were significantly shorter but showed a trend to lower image quality compared to conventional 4D flow. Hemodynamic indices for linear and out-center-out k-space samplings were in agreement with conventional 4D flow (Qmax ≤ 13%, Qnet ≤ 13%, Vmax ≤ 17%, P > 0.05).
CONCLUSION: Aortic 4D flow MRI in under 2 minutes is feasible with moderate underestimation of flow indices. Differences in k-space sampling patterns suggest an opportunity to mitigate image artifacts by an optimal trade-off between scan time, acceleration, and k-space sampling. Magn Reson Med 79:195-207, 2018.
© 2018 International Society for Magnetic Resonance in Medicine. © 2017 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  4D flow magnetic resonance imaging; aortic hemodynamics; k-space sampling; k-t acceleration; navigator gating; short scan time

Mesh:

Year:  2017        PMID: 28266062      PMCID: PMC5589472          DOI: 10.1002/mrm.26661

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  35 in total

1.  Closed circuit MR compatible pulsatile pump system using a ventricular assist device and pressure control unit.

Authors:  R Lorenz; C Benk; J Bock; A F Stalder; J G Korvink; J Hennig; M Markl
Journal:  Magn Reson Med       Date:  2011-05-31       Impact factor: 4.668

2.  Generalized autocalibrating partially parallel acquisitions (GRAPPA).

Authors:  Mark A Griswold; Peter M Jakob; Robin M Heidemann; Mathias Nittka; Vladimir Jellus; Jianmin Wang; Berthold Kiefer; Axel Haase
Journal:  Magn Reson Med       Date:  2002-06       Impact factor: 4.668

3.  Highly undersampled phase-contrast flow measurements using compartment-based k-t principal component analysis.

Authors:  Daniel Giese; Tobias Schaeffter; Sebastian Kozerke
Journal:  Magn Reson Med       Date:  2012-04-23       Impact factor: 4.668

Review 4.  Phase contrast cine magnetic resonance imaging.

Authors:  N J Pelc; R J Herfkens; A Shimakawa; D R Enzmann
Journal:  Magn Reson Q       Date:  1991-10

5.  Parallel MRI with extended and averaged GRAPPA kernels (PEAK-GRAPPA): optimized spatiotemporal dynamic imaging.

Authors:  Bernd Jung; Peter Ullmann; Matthias Honal; Simon Bauer; Jürgen Hennig; Michael Markl
Journal:  J Magn Reson Imaging       Date:  2008-11       Impact factor: 4.813

6.  Efficient method for volumetric assessment of peak blood flow velocity using 4D flow MRI.

Authors:  Michael J Rose; Kelly Jarvis; Varun Chowdhary; Alex J Barker; Bradley D Allen; Joshua D Robinson; Michael Markl; Cynthia K Rigsby; Susanne Schnell
Journal:  J Magn Reson Imaging       Date:  2016-05-18       Impact factor: 4.813

7.  3.0T, time-resolved, 3D flow-sensitive MR in the thoracic aorta: Impact of k-t BLAST acceleration using 8- versus 32-channel coil arrays.

Authors:  Arshad Zaman; Manish Motwani; James J Oliver; Gerard Crelier; Laura E Dobson; David M Higgins; Sven Plein; John P Greenwood
Journal:  J Magn Reson Imaging       Date:  2014-12-02       Impact factor: 4.813

8.  Whole-heart four-dimensional flow can be acquired with preserved quality without respiratory gating, facilitating clinical use: a head-to-head comparison.

Authors:  Mikael Kanski; Johannes Töger; Katarina Steding-Ehrenborg; Christos Xanthis; Karin Markenroth Bloch; Einar Heiberg; Marcus Carlsson; Håkan Arheden
Journal:  BMC Med Imaging       Date:  2015-06-18       Impact factor: 1.930

9.  Comparison of respiratory motion suppression techniques for 4D flow MRI.

Authors:  Petter Dyverfeldt; Tino Ebbers
Journal:  Magn Reson Med       Date:  2017-01-11       Impact factor: 4.668

10.  Phase-contrast MRI volume flow--a comparison of breath held and navigator based acquisitions.

Authors:  Charlotta Andersson; Johan Kihlberg; Tino Ebbers; Lena Lindström; Carl-Johan Carlhäll; Jan E Engvall
Journal:  BMC Med Imaging       Date:  2016-03-28       Impact factor: 1.930

View more
  15 in total

1.  Aortic 4D flow MRI in 2 minutes using compressed sensing, respiratory controlled adaptive k-space reordering, and inline reconstruction.

Authors:  Liliana E Ma; Michael Markl; Kelvin Chow; Hyungkyu Huh; Christoph Forman; Alireza Vali; Andreas Greiser; James Carr; Susanne Schnell; Alex J Barker; Ning Jin
Journal:  Magn Reson Med       Date:  2019-02-25       Impact factor: 4.668

Review 2.  Intraoperative Imaging and Image Fusion for Venous Interventions.

Authors:  Ponraj Chinnadurai; Jean Bismuth
Journal:  Methodist Debakey Cardiovasc J       Date:  2018 Jul-Sep

Review 3.  Noninvasive imaging assessment of portal hypertension.

Authors:  Paul Kennedy; Octavia Bane; Stefanie J Hectors; Aaron Fischman; Thomas Schiano; Sara Lewis; Bachir Taouli
Journal:  Abdom Radiol (NY)       Date:  2020-09-14

4.  Two-Minute k-Space and Time-accelerated Aortic Four-dimensional Flow MRI: Dual-Center Study of Feasibility and Impact on Velocity and Wall Shear Stress Quantification.

Authors:  Emilie Bollache; Kristopher D Knott; Kelly Jarvis; Redha Boubertakh; Ryan Scott Dolan; Claudia Camaioni; Louise Collins; Paul Scully; Sydney Rabin; Thomas Treibel; James C Carr; Pim van Ooij; Jeremy D Collins; Julia Geiger; James C Moon; Alex J Barker; Steffen E Petersen; Michael Markl
Journal:  Radiol Cardiothorac Imaging       Date:  2019-06-27

5.  Altered Aortic Hemodynamics and Relative Pressure in Patients with Dilated Cardiomyopathy.

Authors:  David Marlevi; Jorge Mariscal-Harana; Nicholas S Burris; Julio Sotelo; Bram Ruijsink; Myrianthi Hadjicharalambous; Liya Asner; Eva Sammut; Radomir Chabiniok; Sergio Uribe; Reidar Winter; Pablo Lamata; Jordi Alastruey; David Nordsletten
Journal:  J Cardiovasc Transl Res       Date:  2021-12-09       Impact factor: 4.132

Review 6.  4D flow MRI applications in congenital heart disease.

Authors:  Judy Rizk
Journal:  Eur Radiol       Date:  2020-09-01       Impact factor: 5.315

7.  Multipoint 5D flow cardiovascular magnetic resonance - accelerated cardiac- and respiratory-motion resolved mapping of mean and turbulent velocities.

Authors:  Jonas Walheim; Hannes Dillinger; Sebastian Kozerke
Journal:  J Cardiovasc Magn Reson       Date:  2019-07-22       Impact factor: 5.364

8.  Fully self-gated whole-heart 4D flow imaging from a 5-minute scan.

Authors:  Aaron Pruitt; Adam Rich; Yingmin Liu; Ning Jin; Lee Potter; Matthew Tong; Saurabh Rajpal; Orlando Simonetti; Rizwan Ahmad
Journal:  Magn Reson Med       Date:  2020-09-30       Impact factor: 4.668

Review 9.  Additional value and new insights by four-dimensional flow magnetic resonance imaging in congenital heart disease: application in neonates and young children.

Authors:  Julia Geiger; Fraser M Callaghan; Barbara E U Burkhardt; Emanuela R Valsangiacomo Buechel; Christian J Kellenberger
Journal:  Pediatr Radiol       Date:  2020-12-11

10.  Pseudo-spiral sampling and compressed sensing reconstruction provides flexibility of temporal resolution in accelerated aortic 4D flow MRI: A comparison with k-t principal component analysis.

Authors:  Lukas M Gottwald; Eva S Peper; Qinwei Zhang; Bram F Coolen; Gustav J Strijkers; Aart J Nederveen; Pim van Ooij
Journal:  NMR Biomed       Date:  2020-01-20       Impact factor: 4.044

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.