| Literature DB >> 28219323 |
Ban Leong Sng1,2, Farida Binte Ithnin3, Deepak Mathur3, Eileen Lew3, Nian-Lin Reena Han4, Alex Tiong-Heng Sia3,5.
Abstract
BACKGROUND: The LMA Protector™ is the latest CE marked single use supraglottic airway device. This airway device provides access and functional separation of the respiratory and digestive tracts. There are two ports (male, female ports) to provide suction in the laryngeal region and insertion of the gastric tube. The aim of our study is to assess the ease of use, airway quality, device positioning, airway leak and complications associated with initial clinical experience in LMA Protector™ usage.Entities:
Keywords: Airway management; Anaesthesia; Laryngeal mask
Mesh:
Year: 2017 PMID: 28219323 PMCID: PMC5319052 DOI: 10.1186/s12871-017-0323-5
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Fig. 1The LMA Protector™ contains two separate drain channels
Baseline and demographic characteristics of subjects recruited
| Parametres | Number of patients | Mean (SD)/Median [Range]/ Percentage |
|---|---|---|
| Age; years | 26 | 43.3 (11.5) |
| Body Weight; kg | 26 | 60.5 (9.4) |
| Body Mass Index; kg/m2 | 26 | 24.9 (3.4) |
| ASA Status: | ||
| Grade 1 | 16 | 61.5% |
| Grade 2 | 10 | 38.5% |
| Modified Mallampati Classification | ||
| 1 | 0 | 0% |
| 2 | 26 | 100% |
| 3 | 0 | 0% |
| 4 | 0 | 0% |
| Duration of Procedures; min | 26 | 19 [12-32] |
The data are represented as mean (SD), median [range] or number (% percentage)
Airway and ventilation profile
| Successful insertion LMA, | |
| 1st attempt; | 23 (88.5%) |
| 2nd attempt; | 3 (11.5%) |
| Resistance on insertion of airway, | |
| No or minimal; | 25 (96.2) |
| Moderate; | 1 (3.8) |
| Time to successful airway placement; sec | 19 (17-21 [14-58]) |
| Volume of air to achieve intra-cuff pressure of 60 cm H2O; ml | 15 (15-17 [13-18]) |
| Vocal cord visibility under fibre-optic inspection, | |
| Grade 1; | 26(100%) |
| Grade 2; | 0 (0%) |
| Grade 3; | 0 (0%) |
| Successful gastric tube insertion, | |
| 1st attempt; | 24 (92.4%) |
| 2nd attempt; | 1 (3.8%) |
| Not possible; | 1 (3.8%) |
| Suprasternal notch test, | |
| Positive; | 26 (100%) |
| Negative; | 0 (0%) |
| Oropharyngeal leak pressure; cm H2O | 25.2 (23-29 [21-30]) |
The data are represented as number (% percentage) or median (IQR [range])
Side effects associated with the LMA protector
| Blood or excessive airway secretion on airway device; % | 0 (0%) |
| Sore throat after 1 day surgery; % | 6 (23.1%) |
| Dysphagia or hoarseness of voice; % | 0 (0%) |