Literature DB >> 19413827

Evaluation of the LMA Supreme in 100 non-paralysed patients.

T M Cook1, J J Gatward, J Handel, R Hardy, C Thompson, R Srivastava, P A Clarke.   

Abstract

We studied the LMA Supreme in 100 elective, anaesthetised, healthy patients assessing: ease of use, airway quality, anatomical and functional positioning, airway leak and complications. Insertion was successful on first, second or third attempt in 90, nine and one patient respectively. Thirty manipulations were required in 22 patients to achieve a clear airway. Median [interquartile (range)] insertion time was 18 [10-25 (5-120)] s. During ventilation, an expired tidal volume of 7 ml x kg(-1) was achieved in all patients. Median [interquartile (range)] airway leak pressure was 24 [20-28 (13-40)] cmH(2)O. On fibreoptic examination via the device, vocal cords were visible in 83 patients (85%). During maintenance, five patients (5%) required 13 airway manipulations. There was one episode of minor regurgitation, without aspiration. Other complications and patient side-effects were mild and few. The LMA Supreme is easily and rapidly inserted, providing a reliable airway and good airway seal. Further studies are indicated to assess safety and performance compared to other supraglottic airway devices.

Entities:  

Mesh:

Year:  2009        PMID: 19413827     DOI: 10.1111/j.1365-2044.2008.05824.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  22 in total

1.  Comparison of Supreme(®) and Soft Seal(®) laryngeal masks for airway management during cardiopulmonary resuscitation in novice doctors: a manikin study.

Authors:  Hanako Kohama; Nobuyasu Komasawa; Ryusuke Ueki; Aoi Samma; Masashi Nakagawa; Shin-ichi Nishi; Yoshiroh Kaminoh
Journal:  J Anesth       Date:  2010-12-01       Impact factor: 2.078

2.  Comparison of LM-Supreme™ and endotracheal tube in patients undergoing gynecological laparoscopic surgery.

Authors:  Bahar Kuvaki; Şule Özbilgin; Sakize Ferim Günenç; Burcu Ataseven Küçük
Journal:  J Clin Monit Comput       Date:  2019-04-09       Impact factor: 2.502

3.  Comparison of Laryngeal Mask Supreme® and Soft Seal® for airway management in several positions.

Authors:  Nobuyasu Komasawa; Ryusuke Ueki; Akari Fujii; Aoi Samma; Masashi Nakagawa; Shin-ich Nishi; Yoshiroh Kaminoh
Journal:  J Anesth       Date:  2011-04-12       Impact factor: 2.078

4.  A novel modified laryngeal mask airway allowing full separation of the digestive and respiratory tracts, along with double-catheter ventilation, diagnosis, and treatment.

Authors:  Jun Yin; Shao-Lin Wang; Xiao-Bin Liu
Journal:  Int Surg       Date:  2014 Mar-Apr

5.  Positive end-expiratory pressure increases arterial oxygenation in elderly patients undergoing urological surgery using laryngeal mask airway in lithotomy position.

Authors:  Doo-Hwan Kim; Jun-Young Park; Jihion Yu; Gi-Ho Koh; Eunkyul Kim; Jai-Hyun Hwang; Young-Kug Kim
Journal:  J Clin Monit Comput       Date:  2019-02-20       Impact factor: 2.502

6.  Comparison of Size 2 LMA-ProSeal™ and LMA-Supreme™ in Spontaneously Breathing Children: a Randomised Clinical Trial.

Authors:  Zehra İpek Arslan; Canan Balcı; Duygu Akalın Oysu; Mehmet Yılmaz; Necla Gürbüz; Zekeriya Ilce
Journal:  Balkan Med J       Date:  2013-03-01       Impact factor: 2.021

7.  Does a 4 diagram manual enable laypersons to operate the Laryngeal Mask Supreme®? A pilot study in the manikin.

Authors:  Gereon Schälte; Christian Stoppe; Rolf Rossaint; Laura Gilles; Maike Heuser; Steffen Rex; Mark Coburn; Norbert Zoremba; Annette Rieg
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2012-03-27       Impact factor: 2.953

8.  A comparison of the performance of the I-gel™ vs. the LMA-S™during anesthesia: a meta-analysis of randomized controlled trials.

Authors:  Xiaoguang Chen; Jinghua Jiao; Xuefeng Cong; Lei Liu; Xiaomei Wu
Journal:  PLoS One       Date:  2013-08-12       Impact factor: 3.240

9.  Postoperative pharyngolaryngeal adverse events with laryngeal mask airway (LMA Supreme) in laparoscopic surgical procedures with cuff pressure limiting 25 cmH₂O: prospective, blind, and randomised study.

Authors:  Joo-Eun Kang; Chung-Sik Oh; Jae Won Choi; Il Soon Son; Seong-Hyop Kim
Journal:  ScientificWorldJournal       Date:  2014-03-17

10.  Extraglottic airway devices: A review.

Authors:  Ramesh Ramaiah; Debasmita Das; Sanjay M Bhananker; Aaron M Joffe
Journal:  Int J Crit Illn Inj Sci       Date:  2014-01