Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement.

Jarkko Harju1, Antti Vehkaoja2, Pekka Kumpulainen2, Stefano Campadello3, Ville Lindroos4, Arvi Yli-Hankala5,4, Niku Oksala4,6.   

Abstract

Intermittent non-invasive blood pressure measurement with tourniquets is slow, can cause nerve and skin damage, and interferes with other measurements. Invasive measurement cannot be safely used in all conditions. Modified arterial tonometry may be an alternative for fast and continuous measurement. Our aim was to compare arterial tonometry sensor (BPro®) with invasive blood pressure measurement to clarify whether it could be utilized in the postoperative setting. 28 patients who underwent elective surgery requiring arterial cannulation were analyzed. Patients were monitored post-operatively for 2 h with standard invasive monitoring and with a study device comprising an arterial tonometry sensor (BPro®) added with a three-dimensional accelerometer to investigate the potential impact of movement. Recordings were collected electronically. The results revealed inaccurate readings in method comparison between the devices based on recommendations by Association for the Advancement of Medical Instrumentation (AAMI). On a Bland-Altman plot, the bias and precision between these two methods was 19.8 ± 16.7 (Limits of agreement - 20.1 to 59.6) mmHg, Spearman correlation coefficient r = 0.61. For diastolic pressure, the difference was 4.8 ± 7.7 (LoA - 14.1 to 23.6) mmHg (r = 0.72), and for mean arterial pressure it was 11.18 ± 11.1 (LoA - 12.1 to 34.2) mmHg (r = 0.642). Our study revealed inaccurate agreement (AAMI) between the two methods when measuring systolic and mean blood pressures during post-operative care. The readings for diastolic pressures were inside the limits recommended by AAMI. Movement increased the failure rate significantly (p < 0.001). Thus, arterial tonometry is not an appropriate replacement for invasive blood pressure measurement in these patients.

Entities:  

Keywords:  Arterial pressure; Blood pressure monitors; Monitoring, intraoperative

Mesh:

Year:  2017        PMID: 28105538     DOI: 10.1007/s10877-017-9984-3

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  27 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  Comparison of two measuring instruments, B-pro and SphygmoCor system as reference, to evaluate central systolic blood pressure and radial augmentation index.

Authors:  Luis Garcia-Ortiz; José I Recio-Rodríguez; Joan J Canales-Reina; Alfredo Cabrejas-Sánchez; Amparo Gomez-Arranz; Jose F Magdalena-Belio; Nahia Guenaga-Saenz; Cristina Agudo-Conde; Manuel A Gomez-Marcos
Journal:  Hypertens Res       Date:  2012-02-02       Impact factor: 3.872

3.  Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research.

Authors:  Thomas G Pickering; John E Hall; Lawrence J Appel; Bonita E Falkner; John Graves; Martha N Hill; Daniel W Jones; Theodore Kurtz; Sheldon G Sheps; Edward J Roccella
Journal:  Hypertension       Date:  2004-12-20       Impact factor: 10.190

4.  Blood pressure cuff compression injury of the radial nerve.

Authors:  C C Lin; B Jawan; M V de Villa; F C Chen; P P Liu
Journal:  J Clin Anesth       Date:  2001-06       Impact factor: 9.452

5.  Radial artery applanation tonometry for continuous non-invasive arterial pressure monitoring in intensive care unit patients: comparison with invasively assessed radial arterial pressure.

Authors:  A S Meidert; W Huber; J N Müller; M Schöfthaler; A Hapfelmeier; N Langwieser; J Y Wagner; F Eyer; R M Schmid; B Saugel
Journal:  Br J Anaesth       Date:  2013-12-18       Impact factor: 9.166

6.  Confidence interval estimation for the Bland-Altman limits of agreement with multiple observations per individual.

Authors:  G Y Zou
Journal:  Stat Methods Med Res       Date:  2011-06-24       Impact factor: 3.021

7.  Tracking Changes in Cardiac Output: Statistical Considerations on the 4-Quadrant Plot and the Polar Plot Methodology.

Authors:  Bernd Saugel; Oliver Grothe; Julia Y Wagner
Journal:  Anesth Analg       Date:  2015-08       Impact factor: 5.108

Review 8.  Accurate, reproducible measurement of blood pressure.

Authors:  N R Campbell; A Chockalingam; J G Fodor; D W McKay
Journal:  CMAJ       Date:  1990-07-01       Impact factor: 8.262

9.  Ambulatory tonometric blood pressure measurements in patients with diabetes.

Authors:  Simone Theilade; Christel Joergensen; Frederik Persson; Maria Lajer; Peter Rossing
Journal:  Diabetes Technol Ther       Date:  2012-04-23       Impact factor: 6.118

10.  Skin avulsion during oscillometry.

Authors:  G S Bause; A C Weintraub; G E Tanner
Journal:  J Clin Monit       Date:  1986-10
View more
  3 in total

1.  A novel non-invasive blood pressure waveform measuring system compared to Millar applanation tonometry.

Authors:  Sándor Földi; Tamás Horváth; Flóra Zieger; Péter Sótonyi; György Cserey
Journal:  J Clin Monit Comput       Date:  2017-10-04       Impact factor: 2.502

Review 2.  Journal of clinical monitoring and computing end of year summary 2018: hemodynamic monitoring and management.

Authors:  Bernd Saugel; Moritz Flick; Karim Bendjelid; Lester A H Critchley; Simon T Vistisen; Thomas W L Scheeren
Journal:  J Clin Monit Comput       Date:  2019-03-07       Impact factor: 2.502

Review 3.  Recent Advances in Non-Invasive Blood Pressure Monitoring and Prediction Using a Machine Learning Approach.

Authors:  Siti Nor Ashikin Ismail; Nazrul Anuar Nayan; Rosmina Jaafar; Zazilah May
Journal:  Sensors (Basel)       Date:  2022-08-18       Impact factor: 3.847

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.