Literature DB >> 24355832

Radial artery applanation tonometry for continuous non-invasive arterial pressure monitoring in intensive care unit patients: comparison with invasively assessed radial arterial pressure.

A S Meidert1, W Huber, J N Müller, M Schöfthaler, A Hapfelmeier, N Langwieser, J Y Wagner, F Eyer, R M Schmid, B Saugel.   

Abstract

BACKGROUND: Radial artery applanation tonometry technology can be used for continuous non-invasive measurement of arterial pressure (AP). The purpose of this study was to evaluate this AP monitoring technology in intensive care unit (ICU) patients in comparison with invasive AP monitoring using a radial arterial catheter.
METHODS: In 24 ICU patients (German university hospital), AP values were simultaneously recorded on a beat-to-beat basis using radial artery applanation tonometry (T-Line system; Tensys Medical, San Diego, CA, USA) and a radial arterial catheter (contralateral arm). The primary endpoint of the study was to investigate the accuracy and precision of the non-invasively assessed AP measurements with the Bland-Altman method based on averaged 10 beat AP epochs (n=2993 10 beat epochs).
RESULTS: For mean AP (MAP), systolic AP (SAP), and diastolic AP (DAP), we observed a bias (±standard deviation of the bias; 95% limits of agreement; percentage error) of +2 mm Hg (±6; -11 to +15 mm Hg; 15%), -3 mm Hg (±15; -33 to +27 mm Hg; 23%), and +5 mm Hg (±7; -9 to +19 mm Hg; 22%), respectively.
CONCLUSIONS: In ICU patients, MAP and DAP measurements obtained using radial artery applanation tonometry show clinically acceptable agreement with invasive AP determination with a radial arterial catheter. While the radial artery applanation tonometry technology also allows SAP measurements with high accuracy, its precision for SAP measurements needs to be further improved.

Entities:  

Keywords:  T-Line; arterial pressure, measurement; measurement techniques, arterial pressure; monitoring, intensive care

Mesh:

Year:  2013        PMID: 24355832     DOI: 10.1093/bja/aet400

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  17 in total

1.  Continuous noninvasive arterial pressure measurement using the volume clamp method: an evaluation of the CNAP device in intensive care unit patients.

Authors:  Julia Y Wagner; Ileana Negulescu; Miriam Schöfthaler; Alexander Hapfelmeier; Agnes S Meidert; Wolfgang Huber; Roland M Schmid; Bernd Saugel
Journal:  J Clin Monit Comput       Date:  2015-03-01       Impact factor: 2.502

2.  The future of intraoperative blood pressure management.

Authors:  Frederic Michard; Ngai Liu; Andrea Kurz
Journal:  J Clin Monit Comput       Date:  2017-02-07       Impact factor: 2.502

3.  Comparison of blood pressure monitoring by applanation tonometry and invasively assessed blood pressure in cardiological patients.

Authors:  G Greiwe; S Hoffmann; L Herich; M S Winkler; C J Trepte; C R Behem; M Petzoldt; D A Reuter; S A Haas
Journal:  J Clin Monit Comput       Date:  2017-12-04       Impact factor: 2.502

4.  Continuous blood pressure monitoring via non-invasive radial artery applanation tonometry and invasive arterial catheter demonstrates good agreement in patients undergoing colon carcinoma surgery.

Authors:  Jing Sun; Hanjian Chen; Jun Zheng; Bin Mao; Shengmei Zhu; Jingyi Feng
Journal:  J Clin Monit Comput       Date:  2016-12-20       Impact factor: 2.502

5.  Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement.

Authors:  Jarkko Harju; Antti Vehkaoja; Pekka Kumpulainen; Stefano Campadello; Ville Lindroos; Arvi Yli-Hankala; Niku Oksala
Journal:  J Clin Monit Comput       Date:  2017-01-19       Impact factor: 2.502

6.  Intraocular pressure with rebound tonometry and effects of topical intraocular pressure reducing medications in guinea pigs.

Authors:  Yue Di; Xiu-Mei Luo; Tong Qiao; Na Lu
Journal:  Int J Ophthalmol       Date:  2017-02-18       Impact factor: 1.779

7.  Radial artery applanation tonometry for continuous noninvasive arterial blood pressure monitoring in the cardiac intensive care unit.

Authors:  Nicolas Langwieser; Luisa Prechtl; Agnes S Meidert; Alexander Hapfelmeier; Christian Bradaric; Tareq Ibrahim; Karl-Ludwig Laugwitz; Roland M Schmid; Julia Y Wagner; Bernd Saugel
Journal:  Clin Res Cardiol       Date:  2015-01-25       Impact factor: 5.460

Review 8.  Preeclampsia and Vascular Function: A Window to Future Cardiovascular Disease Risk.

Authors:  Davaasambuu Enkhmaa; Danielle Wall; Puja K Mehta; Jennifer J Stuart; Janet Wilson Rich-Edwards; C Noel Bairey Merz; Chrisandra Shufelt
Journal:  J Womens Health (Larchmt)       Date:  2016-01-18       Impact factor: 2.681

Review 9.  Assessment of volume status and fluid responsiveness in the emergency department: a systematic approach.

Authors:  C Maurer; J Y Wagner; R M Schmid; B Saugel
Journal:  Med Klin Intensivmed Notfmed       Date:  2015-12-16       Impact factor: 0.840

Review 10.  Hemodynamic Monitoring in the Critically Ill Patient - Current Status and Perspective.

Authors:  Samir G Sakka
Journal:  Front Med (Lausanne)       Date:  2015-08-03