Ambulatory tonometric blood pressure measurements in patients with diabetes.

Simone Theilade1, Christel Joergensen, Frederik Persson, Maria Lajer, Peter Rossing.   

Abstract

BACKGROUND: Arterial tonometry is a novel technique for measuring ambulatory blood pressure (AMBP). The watch-like device BPro(®) (HealthSTATS International, Singapore) captures radial pulsewave reflection and calculates brachial blood pressure (BP). In this study we investigate if arterial tonometry is applicable and reliable in patients with diabetes. SUBJECTS AND METHODS: We compared tonometric (BPro) to cuff-based oscillometric and auscultatoric BPs (Takeda model TM2421, A&D Medical, Tokyo, Japan) in 25 Caucasian patients with type 1 or type 2 diabetes. Patients were seen twice within 2 weeks. At visit 1, a 15-min rest was followed by the recording of three cuff-based BPs and 2-min continuous tonometric BPs. At both visits 24-h AMBP measurements were recorded with the BPro device.
RESULTS: At Visit 1, auscultatoric BP (mean±SD) was 136±19/72±8 mm Hg versus 138±19/78±8 mm Hg with the tonometric device. Visit 1 AMBP was 131±20/76±9 mm Hg versus 131±12/75±9 mm Hg at Visit 2. Mean 24-h AMBP, daytime BP, nighttime BP, and dipping at the two visits were similar (P>0.40). Linear and intraclass correlations coefficients between auscultatoric and tonometric systolic and diastolic BP were r=0.86 and 0.65, respectively (P<0.001 for both), and r=0.83 and 0.77, respectively (P<0.001 for both). The mean differences between devices were 1.9±10 and 5.5±6.6 mm Hg for systolic and diastolic BP, respectively.
CONCLUSIONS: In patients with diabetes tonometric and cuff-based BPs are comparable, and tonometric AMBPs are reproducible and feasible.

Entities:  

Mesh:

Year:  2012        PMID: 22524631     DOI: 10.1089/dia.2012.0006

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  13 in total

Review 1.  Blood Pressure Assessment in Adults in Clinical Practice and Clinic-Based Research: JACC Scientific Expert Panel.

Authors:  Paul Muntner; Paula T Einhorn; William C Cushman; Paul K Whelton; Natalie A Bello; Paul E Drawz; Beverly B Green; Daniel W Jones; Stephen P Juraschek; Karen L Margolis; Edgar R Miller; Ann Marie Navar; Yechiam Ostchega; Michael K Rakotz; Bernard Rosner; Joseph E Schwartz; Daichi Shimbo; George S Stergiou; Raymond R Townsend; Jeff D Williamson; Jackson T Wright; Lawrence J Appel
Journal:  J Am Coll Cardiol       Date:  2019-01-29       Impact factor: 24.094

2.  Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement.

Authors:  Jarkko Harju; Antti Vehkaoja; Pekka Kumpulainen; Stefano Campadello; Ville Lindroos; Arvi Yli-Hankala; Niku Oksala
Journal:  J Clin Monit Comput       Date:  2017-01-19       Impact factor: 2.502

3.  Cardiac (82)Rb PET/CT for fast and non-invasive assessment of microvascular function and structure in asymptomatic patients with type 2 diabetes.

Authors:  Bernt J von Scholten; Philip Hasbak; Thomas E Christensen; Adam A Ghotbi; Andreas Kjaer; Peter Rossing; Tine W Hansen
Journal:  Diabetologia       Date:  2015-11-02       Impact factor: 10.122

4.  Non-invasive assessment of temporal changes in myocardial microvascular function in persons with type 2 diabetes and healthy controls.

Authors:  Ida K B Rasmussen; Philip Hasbak; Bernt J von Scholten; Jens C Laursen; Emilie H Zobel; Lars Jorge Diaz; Lene Holmvang; Rasmus S Ripa; Peter Rossing; Andreas Kjaer; Tine W Hansen
Journal:  Diabet Med       Date:  2021-01-25       Impact factor: 4.359

5.  Effect of large weight reductions on measured and estimated kidney function.

Authors:  Bernt Johan von Scholten; Frederik Persson; Maria S Svane; Tine W Hansen; Sten Madsbad; Peter Rossing
Journal:  BMC Nephrol       Date:  2017-02-06       Impact factor: 2.388

6.  QardioArm Blood Pressure Monitoring in a Population With Type 2 Diabetes: Validation Study.

Authors:  Daniel López-López; Victoria Mazoteras-Pardo; Ricardo Becerro-De-Bengoa-Vallejo; Marta Elena Losa-Iglesias; Eva María Martínez-Jiménez; César Calvo-Lobo; Carlos Romero-Morales; Patricia Palomo-López
Journal:  J Med Internet Res       Date:  2020-07-24       Impact factor: 5.428

7.  Discrepancy between tonometric ambulatory and cuff-based office blood pressure measurements in patients with type 1 diabetes.

Authors:  Simone Theilade; Maria Lajer; Christel Joergensen; Frederik Persson; Peter Rossing
Journal:  J Clin Hypertens (Greenwich)       Date:  2012-08-03       Impact factor: 3.738

8.  Arterial stiffness is associated with cardiovascular, renal, retinal, and autonomic disease in type 1 diabetes.

Authors:  Simone Theilade; Maria Lajer; Frederik Persson; Christel Joergensen; Peter Rossing
Journal:  Diabetes Care       Date:  2012-11-27       Impact factor: 19.112

9.  Validation of BP devices QardioArm® in the general population and Omron M6 Comfort® in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP).

Authors:  Mirna N Chahine; Jirar Topouchian; Parounak Zelveian; Zoya Hakobyan; Arevik Melkonyan; Alaa Azaki; Reem Diab; Aya Harb; Roland Asmar
Journal:  Med Devices (Auckl)       Date:  2017-12-27

10.  Circulating matrix metalloproteinases are associated with arterial stiffness in patients with type 1 diabetes: pooled analysis of three cohort studies.

Authors:  Stijn A Peeters; Lian Engelen; Jacqueline Buijs; Nish Chaturvedi; John H Fuller; Anders Jorsal; Hans-Henrik Parving; Lise Tarnow; Simone Theilade; Peter Rossing; Casper G Schalkwijk; Coen D A Stehouwer
Journal:  Cardiovasc Diabetol       Date:  2017-10-25       Impact factor: 9.951

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.