Literature DB >> 28079300

Assessing the accuracy of the OMRON HEM-907XL oscillometric blood pressure measurement device in patients with nondialytic chronic kidney disease.

Jordana B Cohen1, Tiffany C Wong2, Bruce S Alpert3, Raymond R Townsend1.   

Abstract

The OMRON HEM-907XL is a commercial oscillometric blood pressure (BP) monitor that was used in the Systolic Blood Pressure Intervention Trial (SPRINT), in which 28% of participants had chronic kidney disease (CKD). This study examined the accuracy of the monitor in nondialytic patients with CKD. Eighty-seven patients met inclusion criteria. The authors used a modified Association for the Advancement of Medical Instrumentation (AAMI) protocol, with one observer recording measurements from the monitor and two blinded physicians obtaining simultaneous aneroid values by auscultation. Using AAMI method 1, there was a 2.5±9.5 mm Hg difference in OMRON and aneroid systolic BP, and a -1.6±6.5 mm Hg difference in diastolic BP. Using AAMI method 2, there was a 5.1±7.4 mm Hg difference in systolic BP and a -0.2±5.4 mm Hg difference in diastolic BP. In patients with CKD, the OMRON HEM-907XL appears to be accurate for measuring diastolic BP, but did not perform as well for systolic BP. ©2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  ambulatory blood pressure/home blood pressure monitor; blood pressure monitoring; chronic kidney disease; clinical management of high blood pressure; hypertension

Mesh:

Year:  2017        PMID: 28079300      PMCID: PMC5340575          DOI: 10.1111/jch.12961

Source DB:  PubMed          Journal:  J Clin Hypertens (Greenwich)        ISSN: 1524-6175            Impact factor:   3.738


  28 in total

1.  Validation of the Omron HEM-907 device for blood pressure measurement.

Authors:  Mohamed A El Assaad; Jirar A Topouchian; Bernadette M Darné; Roland G Asmar
Journal:  Blood Press Monit       Date:  2002-08       Impact factor: 1.444

2.  Accuracy of oscillometric blood pressure monitoring with concurrent auscultatory blood pressure in hemodialysis patients.

Authors:  Merfake Semret; Miriam Zidehsarai; Rajiv Agarwal
Journal:  Blood Press Monit       Date:  2005-10       Impact factor: 1.444

Review 3.  Oscillometric Blood Pressure Estimation: Past, Present, and Future.

Authors:  Mohamad Forouzanfar; Hilmi R Dajani; Voicu Z Groza; Miodrag Bolic; Sreeraman Rajan; Izmail Batkin
Journal:  IEEE Rev Biomed Eng       Date:  2015-05-15

Review 4.  Oscillometric blood pressure: a review for clinicians.

Authors:  Bruce S Alpert; David Quinn; David Gallick
Journal:  J Am Soc Hypertens       Date:  2014-09-02

5.  Endothelial function and cardiovascular events in chronic kidney disease.

Authors:  Yoshihiro Hirata; Seigo Sugiyama; Eiichiro Yamamoto; Yasushi Matsuzawa; Eiichi Akiyama; Hiroaki Kusaka; Koichiro Fujisue; Hirofumi Kurokawa; Junichi Matsubara; Koichi Sugamura; Hirofumi Maeda; Satomi Iwashita; Hideaki Jinnouchi; Kunihiko Matsui; Hisao Ogawa
Journal:  Int J Cardiol       Date:  2014-03-20       Impact factor: 4.164

6.  Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer.

Authors:  N M van Popele; W J Bos; N A de Beer; D A van Der Kuip; A Hofman; D E Grobbee; J C Witteman
Journal:  Hypertension       Date:  2000-10       Impact factor: 10.190

7.  Determining the accuracy of blood pressure measurement by the Omron HEM-907 before and after hemodialysis.

Authors:  Marek Czarkowski; Monika Staszków; Kacper Kostyra; Zuhier Shebani; Stanisław Niemczyk; Joanna Matuszkiewicz-Rowińska
Journal:  Blood Press Monit       Date:  2009-10       Impact factor: 1.444

8.  Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension.

Authors:  Stefano Omboni; Isabella Riva; Alessia Giglio; Gianluca Caldara; Antonella Groppelli; Gianfranco Parati
Journal:  Blood Press Monit       Date:  2007-08       Impact factor: 1.444

Review 9.  Chronic kidney disease and cardiovascular risk: epidemiology, mechanisms, and prevention.

Authors:  Ron T Gansevoort; Ricardo Correa-Rotter; Brenda R Hemmelgarn; Tazeen H Jafar; Hiddo J Lambers Heerspink; Johannes F Mann; Kunihiro Matsushita; Chi Pang Wen
Journal:  Lancet       Date:  2013-05-31       Impact factor: 79.321

10.  A new equation to estimate glomerular filtration rate.

Authors:  Andrew S Levey; Lesley A Stevens; Christopher H Schmid; Yaping Lucy Zhang; Alejandro F Castro; Harold I Feldman; John W Kusek; Paul Eggers; Frederick Van Lente; Tom Greene; Josef Coresh
Journal:  Ann Intern Med       Date:  2009-05-05       Impact factor: 25.391

View more
  11 in total

Review 1.  Automated Office-Based Blood Pressure Measurement: an Overview and Guidance for Implementation in Primary Care.

Authors:  Romsai T Boonyasai; Erika L McCannon; Joseph E Landavaso
Journal:  Curr Hypertens Rep       Date:  2019-04-04       Impact factor: 5.369

2.  History and Justification of a National Blood Pressure Measurement Validated Device Listing.

Authors:  Jordana B Cohen; Raj S Padwal; Michael Gutkin; Beverly B Green; Michael J Bloch; F Wilford Germino; Domenic A Sica; Anthony J Viera; Benjamin M Bluml; William B White; Sandra J Taler; Steven Yarows; Daichi Shimbo; Raymond R Townsend
Journal:  Hypertension       Date:  2019-02       Impact factor: 10.190

3.  Working to Increase Stability through Exercise (WISE): Study protocol for a pragmatic randomized controlled trial of a coached exercise program to reduce serious fall-related injuries.

Authors:  Christopher Sciamanna; Noel H Ballentine; Melissa Bopp; Jennifer S Brach; Vernon M Chinchilli; Joseph T Ciccolo; Molly B Conroy; Abigail Fisher; Edward J Fox; Susan L Greenspan; M Jan De Beur Suzanne; Kalen Kearcher; Jennifer L Kraschnewski; Kathleen M McTigue; Edward McAuley; Natalia E Morone; Anuradha Paranjape; Sol Rodriguez-Colon; Andrew Rosenzweig; Joshua M Smyth; Kerry J Stewart; Heather L Stuckey
Journal:  Contemp Clin Trials       Date:  2018-09-24       Impact factor: 2.226

4.  Partnering with Churches to Conduct a Wide-Scale Health Screening of an Urban, Segregated Community.

Authors:  Elizabeth B Lynch; Joselyn Williams; Elizabeth Avery; Melissa M Crane; Brittney Lange-Maia; Christy Tangney; LaDawne Jenkins; Sheila A Dugan; Erin E Emery-Tiburcio; Steve M Epting
Journal:  J Community Health       Date:  2020-02

5.  Validation of Blood Pressure Device Accuracy: When the Bottom Line Is Not Enough.

Authors:  Jordana B Cohen; Tammy M Brady
Journal:  Circulation       Date:  2022-01-10       Impact factor: 29.690

Review 6.  Blood Pressure Goals in Patients with CKD: A Review of Evidence and Guidelines.

Authors:  Alex R Chang; Meghan Lóser; Rakesh Malhotra; Lawrence J Appel
Journal:  Clin J Am Soc Nephrol       Date:  2018-11-19       Impact factor: 8.237

7.  Assessing the accuracy of the OMRON HEM-907XL oscillometric blood pressure measurement device in patients with nondialytic chronic kidney disease.

Authors:  Jordana B Cohen; Tiffany C Wong; Bruce S Alpert; Raymond R Townsend
Journal:  J Clin Hypertens (Greenwich)       Date:  2017-01-12       Impact factor: 3.738

8.  Blood pressure is blood pressure is blood pressure: Or is it?

Authors:  Rajiv Agarwal
Journal:  J Clin Hypertens (Greenwich)       Date:  2017-01-13       Impact factor: 3.738

9.  The ongoing saga of poor blood pressure measurement: Past, present, and future perspectives.

Authors:  Jennifer Ringrose; Raj Padwal
Journal:  J Clin Hypertens (Greenwich)       Date:  2017-05-02       Impact factor: 3.738

10.  Changes in blood pressure according to stature in Mexican adults.

Authors:  Miguel A Perez-Sastre; Luis Ortiz-Hernandez
Journal:  Rev Saude Publica       Date:  2021-12-01       Impact factor: 2.106

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.