Validation of the Omron HEM-907 device for blood pressure measurement.

Mohamed A El Assaad1, Jirar A Topouchian, Bernadette M Darné, Roland G Asmar.   

Abstract

BACKGROUND: The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol.
DESIGN: The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 more subjects are selected making a total number of 33 subjects on which the final validation is performed.
METHODS: For each subject, BP measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with the Omron HEM-907 device. In all, 99 measurements were obtained for comparison. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure.
RESULTS: The difference between the two observers was -1 +/- 2 mmHg for the systolic BP (SBP) and for the diastolic BP (DBP). The Omron HEM-907 passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were -1 +/- 7 and -5 +/- 6 mmHg for SBP and DBP respectively. Readings for the HEM-907 device differed by less than 5 mmHg for 61 of the systolic readings and 52 of the diastolic readings; by less than 10 mmHg for 85 of the systolic readings and 85 of the diastolic readings; and by less than 15 mmHg for 94 of the systolic readings and 96 of the diastolic readings.
CONCLUSIONS: The Omron HEM-907 device passes the two phases of the international validation protocol. Copyright 2002 Lippincott Williams & Wilkins

Entities:  

Mesh:

Year:  2002        PMID: 12198340     DOI: 10.1097/00126097-200208000-00006

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  84 in total

1.  The effects of pre-ejection period on post-exercise systolic blood pressure estimation using the pulse arrival time technique.

Authors:  Mico Yee Man Wong; Emma Pickwell-MacPherson; Yuan Ting Zhang; Jack C Y Cheng
Journal:  Eur J Appl Physiol       Date:  2010-09-08       Impact factor: 3.078

2.  Waiting a few extra minutes before measuring blood pressure has potentially important clinical and research ramifications.

Authors:  S B Nikolic; W P Abhayaratna; R Leano; M Stowasser; J E Sharman
Journal:  J Hum Hypertens       Date:  2013-05-30       Impact factor: 3.012

3.  Serum 25-hydroxyvitamin D level and kidney function decline in a Swiss general adult population.

Authors:  Idris Guessous; William McClellan; David Kleinbaum; Viola Vaccarino; Henry Hugues; Olivier Boulat; Pedro Marques-Vidal; Fred Paccaud; Jean-Marc Theler; Jean-Michel Gaspoz; Michel Burnier; Gérard Waeber; Peter Vollenweider; Murielle Bochud
Journal:  Clin J Am Soc Nephrol       Date:  2015-04-21       Impact factor: 8.237

4.  Early pregnancy bisphenol and phthalate metabolite levels, maternal hemodynamics and gestational hypertensive disorders.

Authors:  Elise M Philips; Leonardo Trasande; Linda G Kahn; Romy Gaillard; Eric A P Steegers; Vincent W V Jaddoe
Journal:  Hum Reprod       Date:  2019-02-01       Impact factor: 6.918

5.  The acute effects of running on blood pressure estimation using pulse transit time in normotensive subjects.

Authors:  Mico Yee-Man Wong; Emma Pickwell-MacPherson; Yuan-Ting Zhang
Journal:  Eur J Appl Physiol       Date:  2009-06-20       Impact factor: 3.078

Review 6.  Automated Office-Based Blood Pressure Measurement: an Overview and Guidance for Implementation in Primary Care.

Authors:  Romsai T Boonyasai; Erika L McCannon; Joseph E Landavaso
Journal:  Curr Hypertens Rep       Date:  2019-04-04       Impact factor: 5.369

7.  Effect of Three Hypopnea Scoring Criteria on OSA Prevalence and Associated Comorbidities in the General Population.

Authors:  Camila Hirotsu; Jose Haba-Rubio; Daniela Andries; Nadia Tobback; Pedro Marques-Vidal; Peter Vollenweider; Gérard Waeber; Raphael Heinzer
Journal:  J Clin Sleep Med       Date:  2019-02-15       Impact factor: 4.062

8.  Aortic-to-brachial artery stiffness gradient is not blood pressure independent.

Authors:  Matthew K Armstrong; Martin G Schultz; Dean S Picone; James E Sharman
Journal:  J Hum Hypertens       Date:  2019-01-10       Impact factor: 3.012

9.  The impact of arm position and pulse pressure on the validation of a wrist-cuff blood pressure measurement device in a high risk population.

Authors:  Ali Reza Khoshdel; Shane Carney; Alastair Gillies
Journal:  Int J Gen Med       Date:  2010-04-08

Review 10.  Determining which automatic digital blood pressure device performs adequately: a systematic review.

Authors:  Y Wan; C Heneghan; R Stevens; R J McManus; A Ward; R Perera; M Thompson; L Tarassenko; D Mant
Journal:  J Hum Hypertens       Date:  2010-04-08       Impact factor: 3.012

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.