Accuracy of oscillometric blood pressure monitoring with concurrent auscultatory blood pressure in hemodialysis patients.

Merfake Semret1, Miriam Zidehsarai, Rajiv Agarwal.   

Abstract

BACKGROUND: Oscillometric devices are commonly used to measure blood pressure and their validation entails sequential measurements of auscultated and oscillometric blood pressures. It is unknown whether simultaneous measurement of auscultated blood pressure and concurrent digitization and recording of the sounds can improve assessment of such devices. The aim of this study was to develop a technique of simultaneous measurement of blood pressure using oscillometric and auscultated measurements in the same deflation. Using such a device, we validated the Omron HEM-907 blood pressure measuring device in hemodialysis patients.
METHODS: Twenty non-hypertensive subjects and 20 hemodialysis patients were studied. Six blood pressure readings were obtained in each participant; three readings were obtained using the automatic, oscillometric mode of Omron HEM-907 and three readings were obtained using the manual mode of the Omron HEM-907. In each situation, simultaneous digitized Korotkoff sounds and manometric pressures were recorded. The grading scale indicated by the British Hypertension Society protocol was used to assess the device.
RESULTS: Systolic blood pressure measured simultaneously by digitized sound and pressures agreed closely for systolic pressure (Grade A for both hemodialysis and normal controls) but not for diastolic pressure (Grade C for both hemodialysis and controls). Sequential comparison of oscillometric readings with auscultated systolic and diastolic pressure yielded a Grade B for both hemodialysis patients and controls. When concurrent digitized systolic readings were used, however, Grade A was achieved for both populations for systolic readings, but Grade C for diastolic readings. The mean differences (SD) between the oscillometric and auscultatory blood pressure reading in normal controls were 4.3 (8.9) and 0.6 (8.7) for systolic and diastolic blood pressures, respectively. In hemodialysis patients, the mean differences (SD) were 2.7 (9.3) and 0.4 (7.0) for systolic and diastolic blood pressures, respectively.
CONCLUSION: Simultaneous measurement of digitized Korotkoff sounds and pressure can improve the assessment of devices for systolic but not diastolic blood pressure. The Omron HEM-907 device can be recommended for use in hemodialysis patients.

Entities:  

Mesh:

Year:  2005        PMID: 16205443     DOI: 10.1097/01.mbp.0000172713.28029.84

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  9 in total

Review 1.  Managing hypertension using home blood pressure monitoring among haemodialysis patients--a call to action.

Authors:  Rajiv Agarwal
Journal:  Nephrol Dial Transplant       Date:  2010-03-28       Impact factor: 5.992

2.  Arm circumference: its importance for dialysis patients in the obesity era.

Authors:  Tekin Akpolat; Coşkun Kaya; Cengiz Utaş; Turgay Arinsoy; Hülya Taşkapan; Emre Erdem; M Emin Yilmaz; Rezzan Ataman; Semra Bozfakioğlu; Çetin Özener; Ibrahim Karayaylali; Rümeyza Kazancioğlu; Taner Çamsari; Mahmut Yavuz; Fevzi Ersoy; Soner Duman; Kenan Ateş
Journal:  Int Urol Nephrol       Date:  2012-06-30       Impact factor: 2.370

3.  Call to action on use and reimbursement for home blood pressure monitoring: a joint scientific statement from the American Heart Association, American Society Of Hypertension, and Preventive Cardiovascular Nurses Association.

Authors:  Thomas G Pickering; Nancy Houston Miller; Gbenga Ogedegbe; Lawrence R Krakoff; Nancy T Artinian; David Goff
Journal:  Hypertension       Date:  2008-05-22       Impact factor: 10.190

4.  Assessing the accuracy of the OMRON HEM-907XL oscillometric blood pressure measurement device in patients with nondialytic chronic kidney disease.

Authors:  Jordana B Cohen; Tiffany C Wong; Bruce S Alpert; Raymond R Townsend
Journal:  J Clin Hypertens (Greenwich)       Date:  2017-01-12       Impact factor: 3.738

5.  Blood pressure is blood pressure is blood pressure: Or is it?

Authors:  Rajiv Agarwal
Journal:  J Clin Hypertens (Greenwich)       Date:  2017-01-13       Impact factor: 3.738

6.  Limitations of the of the oscillometric method for blood pressure measurements in dialyzed patients.

Authors:  Marek Czarkowski; Monika Staszkow; Kacper Kostyra; Zuhier Shebani; Krzysztof Rozanowski; Joanna Matuszkiewicz-Rowinska; Stanislaw Niemczyk
Journal:  Med Sci Monit       Date:  2011-04

Review 7.  Sources of inaccuracy in the measurement of adult patients' resting blood pressure in clinical settings: a systematic review.

Authors:  Noa Kallioinen; Andrew Hill; Mark S Horswill; Helen E Ward; Marcus O Watson
Journal:  J Hypertens       Date:  2017-03       Impact factor: 4.844

8.  QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease: A Validation Study According to the European Society of Hypertension International Protocol Revision 2010.

Authors:  Victoria Mazoteras-Pardo; Ricardo Becerro-De-Bengoa-Vallejo; Marta Elena Losa-Iglesias; Daniel López-López; David Rodríguez-Sanz; Israel Casado-Hernández; Cesar Calvo-Lobo; Patricia Palomo-López
Journal:  J Med Internet Res       Date:  2019-12-02       Impact factor: 5.428

9.  An Automated Blood Pressure Display for Self-Measurement in Patients With Chronic Kidney Disease (iHealth Track): Device Validation Study.

Authors:  Victoria Mazoteras-Pardo; Ricardo Becerro-De-Bengoa-Vallejo; Marta Elena Losa-Iglesias; Daniel López-López; César Calvo-Lobo; David Rodríguez-Sanz; Eva María Martínez-Jiménez; Patricia Palomo-López
Journal:  JMIR Mhealth Uhealth       Date:  2020-04-02       Impact factor: 4.773

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.