Determining the accuracy of blood pressure measurement by the Omron HEM-907 before and after hemodialysis.

Marek Czarkowski1, Monika Staszków, Kacper Kostyra, Zuhier Shebani, Stanisław Niemczyk, Joanna Matuszkiewicz-Rowińska.   

Abstract

BACKGROUND: The accuracy of oscillometric blood pressure (BP) measuring devices determined in resting conditions may be distorted in those under hemodialysis (HD).
METHODS: The evaluation of accuracy of the Omron HEM-907 was performed in accordance with the International Protocol for validation of BP measuring devices (IP). Out of a group of 49 patients with end-stage renal disease (ESRD) (56.7 +/- 15.6 years), four groups of 33 patients each were selected to ascertain their BP before and after HD.
RESULTS: The Omron HEM-907 fulfilled the IP accuracy requirements for systolic BP measurements before and after HD. In the case of diastolic BP before HD, the Omron HEM-907 did not fulfill both requirements for phase 2.2 IP. In only 21 out of 33 patients, the difference between the mean of two observers and the Omron HEM-907 device (Diff) was not more than 5 mmHg, while the requirement was 22 patients. In seven patients, all three Diff were over 5 mmHg, while the accepted limit is three. After HD, the Omron HEM-907 did not fulfill one out of two criteria of phase 2.2 IP - in five people, all three Diff were over 5 mmHg.
CONCLUSION: The Omron HEM-907 oscillometric method of BP measurement should not be used to measure BP before and after HD in ESRD patients. In ESRD patients, an alternative method of BP measurement, in the event of a lack of a mercury sphygmomanometer, could be a classical auscultatory method together with the Omron HEM-907, which allows for this type of BP measurement.

Entities:  

Mesh:

Year:  2009        PMID: 19938337     DOI: 10.1097/mbp.0b013e328331d5b5

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  9 in total

Review 1.  Managing hypertension using home blood pressure monitoring among haemodialysis patients--a call to action.

Authors:  Rajiv Agarwal
Journal:  Nephrol Dial Transplant       Date:  2010-03-28       Impact factor: 5.992

Review 2.  Automated Office-Based Blood Pressure Measurement: an Overview and Guidance for Implementation in Primary Care.

Authors:  Romsai T Boonyasai; Erika L McCannon; Joseph E Landavaso
Journal:  Curr Hypertens Rep       Date:  2019-04-04       Impact factor: 5.369

3.  Different expression of 11β-hydroxylase and aldosterone synthase between aldosterone-producing microadenomas and macroadenomas.

Authors:  Yoshikiyo Ono; Yasuhiro Nakamura; Takashi Maekawa; Saulo J A Felizola; Ryo Morimoto; Yoshitsugu Iwakura; Masataka Kudo; Kazumasa Seiji; Kei Takase; Yoichi Arai; Celso E Gomez-Sanchez; Sadayoshi Ito; Hironobu Sasano; Fumitoshi Satoh
Journal:  Hypertension       Date:  2014-05-19       Impact factor: 10.190

4.  Arm circumference: its importance for dialysis patients in the obesity era.

Authors:  Tekin Akpolat; Coşkun Kaya; Cengiz Utaş; Turgay Arinsoy; Hülya Taşkapan; Emre Erdem; M Emin Yilmaz; Rezzan Ataman; Semra Bozfakioğlu; Çetin Özener; Ibrahim Karayaylali; Rümeyza Kazancioğlu; Taner Çamsari; Mahmut Yavuz; Fevzi Ersoy; Soner Duman; Kenan Ateş
Journal:  Int Urol Nephrol       Date:  2012-06-30       Impact factor: 2.370

Review 5.  Hypertension in hemodialysis patients: an opinion-based update.

Authors:  Tariq Shafi; Sana Waheed; Philip G Zager
Journal:  Semin Dial       Date:  2014-02-05       Impact factor: 3.455

6.  Assessing the accuracy of the OMRON HEM-907XL oscillometric blood pressure measurement device in patients with nondialytic chronic kidney disease.

Authors:  Jordana B Cohen; Tiffany C Wong; Bruce S Alpert; Raymond R Townsend
Journal:  J Clin Hypertens (Greenwich)       Date:  2017-01-12       Impact factor: 3.738

7.  Limitations of the of the oscillometric method for blood pressure measurements in dialyzed patients.

Authors:  Marek Czarkowski; Monika Staszkow; Kacper Kostyra; Zuhier Shebani; Krzysztof Rozanowski; Joanna Matuszkiewicz-Rowinska; Stanislaw Niemczyk
Journal:  Med Sci Monit       Date:  2011-04

8.  QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease: A Validation Study According to the European Society of Hypertension International Protocol Revision 2010.

Authors:  Victoria Mazoteras-Pardo; Ricardo Becerro-De-Bengoa-Vallejo; Marta Elena Losa-Iglesias; Daniel López-López; David Rodríguez-Sanz; Israel Casado-Hernández; Cesar Calvo-Lobo; Patricia Palomo-López
Journal:  J Med Internet Res       Date:  2019-12-02       Impact factor: 5.428

9.  An Automated Blood Pressure Display for Self-Measurement in Patients With Chronic Kidney Disease (iHealth Track): Device Validation Study.

Authors:  Victoria Mazoteras-Pardo; Ricardo Becerro-De-Bengoa-Vallejo; Marta Elena Losa-Iglesias; Daniel López-López; César Calvo-Lobo; David Rodríguez-Sanz; Eva María Martínez-Jiménez; Patricia Palomo-López
Journal:  JMIR Mhealth Uhealth       Date:  2020-04-02       Impact factor: 4.773

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.