Dual mobility cups provide biomechanical advantages in situations at risk for dislocation: a finite element analysis.

Alexandre Terrier1, Adeliya Latypova1, Maika Guillemin1, Valérie Parvex1, Olivier Guyen2.   

Abstract

PURPOSE: Constrained devices, standard implants with large heads, and dual mobility systems have become popular options to manage instability after total hip arthroplasty (THA). Clinical results with these options have shown variable success rates and significant higher rates of aseptic loosening and mechanical failures with constrained implants. Literature suggests potential advantages of dual mobility, however little is known about its biomechanics. We present a comparative biomechanical study of a standard implant, a constrained implant, and a dual mobility system.
METHODS: A finite element analysis was developed to assess and compare these acetabular options with regard to the range of motion (ROM) to impingement, the angle of dislocation, the resistive torque, the volume of polyethylene (PE) with a stress above 80% of the elastic limit, and the interfacial cup/bone stress.
RESULTS: Dual mobility implants provided the greatest ROM to impingement and allowed delaying subluxation and dislocation when compared to standard and constrained implants. Dual mobility also demonstrated the lowest resistive torque at subluxation while the constrained implant provided the greatest one. The lowest critical PE volume was observed with the dual mobility implant, and the highest stress at the interfaces was observed with the constrained implant.
CONCLUSION: This study highlights the biomechanical advantages of dual mobility systems over constrained and standard implants, and is supported by the clinical results reported. Therefore, the use of dual mobility systems in situations at risk for instability should be advocated and constrained implants should be restricted to salvage situations.

Entities:  

Keywords:  Biomechanics; Constrained implants; Dislocation; Dual mobility; Instability; Subluxation

Mesh:

Substances:

Year:  2017        PMID: 28070611     DOI: 10.1007/s00264-016-3368-z

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  29 in total

1.  Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept.

Authors:  O K Muratoglu; C R Bragdon; D O'Connor; R S Perinchief; D M Estok; M Jasty; W H Harris
Journal:  J Arthroplasty       Date:  2001-12       Impact factor: 4.757

2.  The use of a constrained acetabular component for recurrent dislocation.

Authors:  Gary S Shapiro; Daniel E Weiland; David C Markel; Douglas E Padgett; Thomas P Sculco; Paul M Pellicci
Journal:  J Arthroplasty       Date:  2003-04       Impact factor: 4.757

3.  Do large heads enhance stability and restore native anatomy in primary total hip arthroplasty?

Authors:  Adolph V Lombardi; Michael D Skeels; Keith R Berend; Joanne B Adams; Orlando J Franchi
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

Review 4.  Results of reoperation for hip dislocation: the big picture.

Authors:  Bernard F Morrey
Journal:  Clin Orthop Relat Res       Date:  2004-12       Impact factor: 4.176

5.  Unconstrained tripolar hip implants: effect on hip stability.

Authors:  Olivier Guyen; Qing Shan Chen; Jacques Bejui-Hugues; Daniel J Berry; Kai-Nan An
Journal:  Clin Orthop Relat Res       Date:  2007-02       Impact factor: 4.176

6.  Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants.

Authors:  Olivier Guyen; David G Lewallen; Miguel E Cabanela
Journal:  J Bone Joint Surg Am       Date:  2008-07       Impact factor: 5.284

Review 7.  Large diameter femoral heads: is bigger always better?

Authors:  H J Cooper; C J Della Valle
Journal:  Bone Joint J       Date:  2014-11       Impact factor: 5.082

8.  Impact of individual anatomy on the benefit of cup medialisation in total hip arthroplasty.

Authors:  Alexandre Terrier; Valérie Parvex; Hannes A Rüdiger
Journal:  Hip Int       Date:  2016-10-08       Impact factor: 2.135

9.  Predicting the compressive mechanical behavior of bone.

Authors:  T S Keller
Journal:  J Biomech       Date:  1994-09       Impact factor: 2.712

10.  Lack of early dislocation following total hip arthroplasty with a new dual mobility acetabular design.

Authors:  Jonathan M Vigdorchik; Michele R D'Apuzzo; David C Markel; Arthur L Malkani; Stephen Raterman; Kipling P Sharpe; Charles N Cornell; Geoffrey H Westrich
Journal:  Hip Int       Date:  2015-02-03       Impact factor: 2.135

View more
  9 in total

1.  Long-term clinical results of bipolar hemiarthroplasty for osteoarthritis and rheumatoid arthritis of the hip: A retrospective study.

Authors:  Tetsutaro Abe; Nobuhiro Kaku; Tomonori Tabata; Hiroaki Tagomori; Hiroshi Tsumura
Journal:  J Orthop       Date:  2019-08-12

2.  Dual-mobility cup: a new French revolution.

Authors:  Jacques Henri Caton; André Ferreira
Journal:  Int Orthop       Date:  2017-02-14       Impact factor: 3.075

3.  Salvage of a monoblock metal-on-metal cup using a dual mobility liner: a two-year MRI follow-up study.

Authors:  Maximilian F Kasparek; Lisa Renner; Martin Faschingbauer; Wenzel Waldstein; Kilian Rueckl; Friedrich Boettner
Journal:  Int Orthop       Date:  2017-09-19       Impact factor: 3.075

Review 4.  Dual Mobility in Total Hip Arthroplasty: Biomechanics, Indications and Complications-Current Concepts.

Authors:  Nilesh Patil; Prashant Deshmane; Ajit Deshmukh; Christopher Mow
Journal:  Indian J Orthop       Date:  2021-10-13       Impact factor: 1.033

5.  Outcomes of dual mobility articulation total hip arthroplasty in ipsilateral residual poliomyelitis.

Authors:  Teng-Feng Zhuang; Song-Wei Huan; Si-Min Luo; Guo-Rong She; Wen-Rui Wu; Jun-Yuan Chen; Ning Liu; Zhen-Gang Zha
Journal:  Int Orthop       Date:  2021-09-27       Impact factor: 3.075

6.  Polyethylene wear of dual mobility cups: a comparative analysis based on patient-specific finite element modeling.

Authors:  Julien Wegrzyn; Alexander Antoniadis; Ehsan Sarshari; Matthieu Boubat; Alexandre Terrier
Journal:  Int Orthop       Date:  2022-01-12       Impact factor: 3.075

7.  Dual mobility in hip arthroplasty: What evidence do we need?

Authors:  S Horriat; F S Haddad
Journal:  Bone Joint Res       Date:  2018-09-15       Impact factor: 5.853

8.  Postero-posterolateral approach in total hip arthroplasty.

Authors:  Mokrane Ait Mokhtar
Journal:  Int Orthop       Date:  2020-07-17       Impact factor: 3.075

9.  Implant survival of 662 dual-mobility cups and 727 constrained liners in primary THA: small femoral head size increases the cumulative incidence of revision.

Authors:  Oskari Pakarinen; Olli Lainiala; Aleksi Reito; Perttu Neuvonen; Keijo Mäkelä; Antti Eskelinen
Journal:  Acta Orthop       Date:  2021-07-09       Impact factor: 3.717

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.