Su Yeon Jo1, Ju Mi Lee1, Hye Lim Kim2, Kyeong Hwa Sin2, Hyeon Ji Lee1, Chulhun Ludgerus Chang1, Hyung Hoi Kim3.
Abstract
BACKGROUND: ABO blood typing in pre-transfusion testing is a major component of the high workload in blood banks that therefore requires automation. We often experienced discrepant results from an automated system, especially weak serum reactions. We evaluated the discrepant results by the reference manual method to confirm ABO blood typing.Entities:
Keywords: ABO typing; AutoVue system; Automated system; Discrepancy; Manual method
Mesh:
Substances:
Year: 2017 PMID: 28028997 PMCID: PMC5203988 DOI: 10.3343/alm.2017.37.2.117
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Results of ABO blood typing from the AutoVue system
| Forward typing | Reverse typing | ABO blood type | N of cases (%) | ||
|---|---|---|---|---|---|
| Anti-A | Anti-B | A cell | B cell | ||
| ≥ 3+ | - | - | ≥ 2+ | A | 4,407 (33.6) |
| - | ≥ 3+ | ≥ 2+ | - | B | 3,416 (26.0) |
| - | - | ≥ 2+ | ≥ 2+ | O | 3,537 (27.0) |
| ≥ 3+ | ≥ 3+ | - | - | AB | 1,456 (11.1) |
| < 3+ | < 3+ | < 2+ | < 2+ | Not classified | 144 (1.1) |
| Error messages at any column | 153 (1.2) | ||||
| Total cases of not classified results | 297 (2.3) | ||||
| Total | 13,113 (100.0) | ||||
Forward typing, ABO blood typing with antibody reagent; Reverse typing, ABO blood typing with red cell; A cell, A1 phenotype red cell reagent; B cell, B phenotype red cell reagent; ≥3+, ≥3+ reaction grade; ≥2+, ≥2+ reaction grade; <3+, <3+ reaction grade; <2+, <2+ reaction grade; -, no reaction; Error messages at columns, the AutoVue system's specific error messages are presented at any column of the cassette (the cassette are composed of anti-A, anti-B, anti-D, A cell and B cell column); Not classified, the samples with discrepancies or error messages at any column of the cassette are not classified into ABO blood group.
Abbreviations: Anti-A, anti-A reagent; Anti-B, anti-B reagent.
Results of ABO blood typing from the manual method
| Forward typing | Reverse typing | ABO blood type | N of cases (%) | ||
|---|---|---|---|---|---|
| Anti-A | Anti-B | A cell | B cell | ||
| ≥ 3+ | - | - | ≥ 2+ | A | 4,546 (34.7) |
| - | ≥ 3+ | ≥ 2+ | - | B | 3,465 (26.4) |
| - | - | ≥ 2+ | ≥ 2+ | O | 3,552 (27.1) |
| ≥ 3+ | ≥ 3+ | - | - | AB | 1,470 (11.2) |
| < 3+ | < 3+ | < 2+ | < 2+ | Not classified | 80 (0.6) |
| Total | 13,113 (100.0) | ||||
Forward typing, ABO blood typing with antibody reagent; Reverse typing, ABO blood typing with red cell; A cell, A1 phenotype red cell reagent; B cell, B phenotype red cell reagent; ≥3+, ≥3+ reaction grade; ≥2+, ≥2+ reaction grade; <3+, <3+ reaction grade; <2+, <2+ reaction grade; -, no reaction; Error messages at columns, the AutoVue system's specific error messages are presented at any column of the cassette (the cassette are composed of anti-A, anti-B, anti-D, A cell and B cell column); Not classified, the samples with discrepancies at any column of the cassette are not classified into ABO blood group; numbers in parentheses indicate the % of cases among the total cases.
Abbreviations: Anti-A, anti-A reagent; Anti-B, anti-B reagent.
The discrepant results of ABO blood typing from the AutoVue system and manual method
| Typing method | AutoVue | Manual | ||
|---|---|---|---|---|
| Forward | Weak red cell reaction | 12 (4.0%) | ABO subgroup | 3 (3.8%) |
| SCT | 9 (11.4%) | |||
| Mixed red cell reaction | 8 (2.7%) | ABO subgroup | 4 (5.1%) | |
| SCT | 4 (5.1%) | |||
| Total | 20 (6.7%) | Total | 20 (25.4%) | |
| Reverse | Weak serum reaction | 98 (33.0%) | Weak serum reaction | 28 (35.4%) |
| Extra serum reaction | 26 (8.8%) | Alloantibodies | 11 (13.9%) | |
| Autoantibodies | 5 (6.3%) | |||
| Cold agglutinins | 15 (19.0%) | |||
| Total | 124 (41.8%) | Total | 59 (74.6%) | |
| Error messages at columns | 153 (51.5%) | None | ||
| Total | 297 (100.0%) | 79 (100.0%) | ||
Typing method, ABO blood typing method; Forward, Forward typing; Reverse, Reverse typing; Error messages at columns, the AutoVue system's specific error messages are presented at any column of the cassette (the cassette are composed of anti-A, anti-B, anti-D, A cell and B cell column); Results, discrepant results; Weak red cell reaction, samples with <3+ reaction grade; Mixed red cell reaction, samples with mixed field reaction; Weak serum reaction, samples with <2+ reaction grade; Extra serum reaction, samples with unpredicted extra reaction at reverwse typing; AutoVue, the AutoVue system; Manual, the manual method; SCT, samples of the patients who had a hematopoietic stem cell transplant; ABO subgroup, samples defined as ABO blood subgroup by additional methods; Alloantibodies, samples with alloantibodies in serum; Autoantibodies, samples with autoantibodies in serum; Cold agglutinins, samples with cold agglutinins in serum; Numbers in parentheses indicate the % of cases showing discrepant results among the total discrepant cases respectively.
Abbreviation: SCT, stem cell transplanted.
Results showing a weak serum reaction from the AutoVue system
| A cell | B cell | ||
|---|---|---|---|
| AutoVue | Manual | AutoVue | Manual |
| 1+ (10) | 3+ (1) | 1+ (41) | 3+ (10) |
| 2+ (5) | 2+ (23) | ||
| 1+ (4) | 1+ (8) | ||
| 0.5+ (0) | 0.5+ (0) | ||
| - (0) | - (0) | ||
| 0.5+ (7) | 3+ (4) | 0.5+ (38) | 3+ (8) |
| 2+ (2) | 2+ (16) | ||
| 1+ (1) | 1+ (7) | ||
| 0.5+ (0) | 0.5+ (7) | ||
| - (0) | - (0) | ||
| - (0) | 2+ (0) | - (2) | 2+ (1) |
| 1+ (0) | 1+ (1) | ||
| 0.5+ (0) | 0.5+ (0) | ||
| - (0) | - (0) | ||
AutoVue, the AutoVue system; Manual, the reference manual method; A cells, A1 phenotype red cell reagent; B cell, B phenotype red cell reagent; 3+, 3+ reaction grade; 2+, 2+ reaction grade; 1+, 1+ reaction grade; 0.5+, 0.5+ reaction grade; -, no reaction; numbers in parentheses indicate the number of cases showing a specific reaction grade.
Results showing a 2+ serum reaction grade in the AutoVue
| A cell | B cell | ||
|---|---|---|---|
| AutoVue | Manual | AutoVue | Manual |
| 2+ (190) | 4+ (10) | 2+ (339) | 4+ (6) |
| 3+ (87) | 3+ (173) | ||
| 2+ (93) | 2+ (160) | ||
| 1+ (0) | 1+ (0) | ||
| 0.5+ (0) | 0.5+ (0) | ||
| - (0) | - (0) | ||
AutoVue, the AutoVue system; Manual, the reference manual method; A cells, A1 phenotype red cell reagent; B cell, B phenotype red cell reagent; 4+, 4+ reaction grade; 3+, 3+ reaction grade; 2+, 2+ reaction grade; 1+, 1+ reaction grade; 0.5+, 0.5+ reaction grade; -, no reaction; numbers in parentheses indicate the number of cases showing a specific reaction grade.