Multi-centre evaluation of pre-transfusional routine tests using 8-column format gel cards (DG Gel®).

J Taylor1, J Hyare, P Stelfox, M Williams, R Lees, M Maley.   

Abstract

BACKGROUND: Advances in immunohaematology laboratory practice to improve performance, cost-effectiveness and patient safety are desirable.
OBJECTIVES: To perform a multi-centre evaluation of the 8-column Grifols DG Gel(®) cards and reagent system to assess its performance, suitability and adaptability to the daily blood transfusion laboratory routine in the United Kingdom. METHODS/MATERIALS: A total of 4281 immunohematological analyses {1825 ABO/D grouping, 1921 antibody screening, 75 Rh phenotyping and K antigen determination, 361 antibody identification and 99 neonates [ABO/D and DAT (direct anti-globulin test)]} were performed on 2255 specimens. All cases were run in parallel with the reference method of each laboratory (DiaMed-ID(®) cards or conventional tube technique in some cases).
RESULTS: Concordant results between Grifols DG Gel(®) system and the reference method were obtained in 97·7% of tests. For ABO grouping by the Grifols DG Gel(®) system, sensitivity was 99·95%, specificity was 99·96%, predictive positive value (PPV) was 99·89% and predictive negative value (PNV) was 99·98%. For D grouping, sensitivity was 99·78%, specificity was 100%, PPV was 100% and PNV was 99·78%. For antibody screening, sensitivity was 90·63%, specificity was 99·94%, PPV was 99·32% and PNV was 99·15%. Of the Rh subgroups and K types, results were 100% concordant. For antibody specificity detection, accuracy was 96·95% for Grifols DG Gel(®) system and 95·29% for DiaMed-ID(®) system. For the newborn tests, concordant results were obtained in 100% of ABO/D grouping and in 89·9% of DAT.
CONCLUSION: The Grifols DG Gel(®) 8-column system is reliable and safe for routine tests performed in the immunohaematology laboratory.
© 2010 The Authors. Transfusion Medicine © 2010 British Blood Transfusion Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21118316     DOI: 10.1111/j.1365-3148.2010.01054.x

Source DB:  PubMed          Journal:  Transfus Med        ISSN: 0958-7578            Impact factor:   2.019


  3 in total

1.  Performance Evaluation of Automated Immunohematology Analyzer IH-500 for Blood Bank Testing.

Authors:  Sang Hyuk Park; Jaewook Kim; Ji-Hun Lim; Joseph Jeong; Seon-Ho Lee
Journal:  Indian J Hematol Blood Transfus       Date:  2019-05-02       Impact factor: 0.900

2.  Comparative Analysis of Clinical Samples Showing Weak Serum Reaction on AutoVue System Causing ABO Blood Typing Discrepancies.

Authors:  Su Yeon Jo; Ju Mi Lee; Hye Lim Kim; Kyeong Hwa Sin; Hyeon Ji Lee; Chulhun Ludgerus Chang; Hyung Hoi Kim
Journal:  Ann Lab Med       Date:  2017-03       Impact factor: 3.464

3.  A comparison of three column agglutination tests for red blood cell alloantibody identification.

Authors:  Siska Blomme; Emilie De Maertelaere; Eline Verhoye
Journal:  BMC Res Notes       Date:  2020-03-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.