A fully automated blood typing system for hospital transfusion services. ABS2000 Study Group.

S G Sandler1, A Langeberg, N Avery, P D Mintz.   

Abstract

BACKGROUND: The results of routine blood bank testing by a fully automated blood typing system (ABS2000) were compared with those obtained by standard manual methods in six hospital transfusion services. STUDY DESIGN AND METHODS: The ABS2000 system uses microtiter plates for determining ABO and D types, solid-phase red cell adherence (SPRCA) assays for antibody detection, and modified SPRCA plates for IgG crossmatches. The transfusion services used their standard manual test tube methods.
RESULTS: Of 3779 donors' samples tested for ABO types (red cell typings only), 3.0 percent could not be interpreted by the ABS2000 system's neural network, because of clots, hemolysis, or lipemic samples. The results for ABO types were concordant for 99.8 percent of the remaining samples. Of 3779 donors' samples tested for D types, the results were concordant for 98.7 percent. Of 7580 patients' samples tested for ABO types (red cell and plasma typings), 5.8 percent could not be interpreted by the ABS2000 system. There was 100-percent concordance of ABO typing results for the remaining 7140 samples. There was 99. 7-percent concordance of results for patients' D types. The results of 96.7 percent of antibody detection tests and 98.8 percent of crossmatches were concordant. Neither method failed to detect a serologically incompatible crossmatch that was associated with a specific, clinically significant alloantibody. The ABS2000 system performed 45 confirmatory donor ABO and D types in 115 minutes, 22 antibody detection tests in 116 minutes, 16 patients' ABO/D types in 149 minutes, and 40 crossmatches in 140 minutes.
CONCLUSION: The ABS2000 blood typing system automates routine blood bank tests with accuracy comparable to that of hospital transfusion services' standard manual methods.

Entities:  

Mesh:

Year:  2000        PMID: 10686004     DOI: 10.1046/j.1537-2995.2000.40020201.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  4 in total

1.  Blood-typing and irregular antibody screening through multi-channel microfluidic discs with surface antifouling modification.

Authors:  Yan-Wen Chen; Wen-Tyng Li; Yung Chang; Rong-Ho Lee; Ging-Ho Hsiue
Journal:  Biomicrofluidics       Date:  2019-05-13       Impact factor: 2.800

2.  Performance Evaluation of Automated Immunohematology Analyzer IH-500 for Blood Bank Testing.

Authors:  Sang Hyuk Park; Jaewook Kim; Ji-Hun Lim; Joseph Jeong; Seon-Ho Lee
Journal:  Indian J Hematol Blood Transfus       Date:  2019-05-02       Impact factor: 0.900

3.  Comparative Analysis of Clinical Samples Showing Weak Serum Reaction on AutoVue System Causing ABO Blood Typing Discrepancies.

Authors:  Su Yeon Jo; Ju Mi Lee; Hye Lim Kim; Kyeong Hwa Sin; Hyeon Ji Lee; Chulhun Ludgerus Chang; Hyung Hoi Kim
Journal:  Ann Lab Med       Date:  2017-03       Impact factor: 3.464

4.  Comparison Between Conventional and Automated Techniques for Blood Grouping and Crossmatching: Experience from a Tertiary Care Centre.

Authors:  Swarupa Nikhil Bhagwat; Jayashree H Sharma; Julie Jose; Charusmita J Modi
Journal:  J Lab Physicians       Date:  2015 Jul-Dec
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.