Evaluation of two automated instruments for pre-transfusion testing: AutoVue Innova and Techno TwinStation.

So Youn Shin1, Kye Chul Kwon, Sun Hoe Koo, Jong Woo Park, Chi Seon Ko, Jeong Hoon Song, Ji Youn Sung.   

Abstract

BACKGROUND: Despite the advances in total laboratory automation, a considerable amount of work in blood banks is still done using outdated manual methods. Some automated pre-transfusion testing instruments have recently been developed. Of these, we evaluated and compared the AutoVue Innova (Ortho, USA) and the Techno TwinStation (DiaMed AG, Switzerland).
METHODS: Forward and reverse ABO/Rh typing and unexpected antibody screening and identification tests were performed on 4,628 samples using the manual method and the two automated instruments. Two different anticoagulants (EDTA and citrate) were compared in ABO/Rh typing and unexpected antibody screening tests. Titrating studies were conducted on the following 7 dilutions using 5 samples of irregular antibodies with anti-E, anti-E & -c, anti-D, and anti-Le(a) with anti-Fy(a): 1:2, 1:4, 1:8, 1:16, 1:32, 1:64, and 1:128. The test throughput per hour, the time required to perform 1 and 100 tests, and a simulation test for total events occurring in 1 day were also measured.
RESULTS: No erroneous results were reported between the two instruments and the manual method. Discrepancies observed in 10 cases (0.4%) of ABO/Rh typing were of higher intensity with AutoVue Innova than with the manual method. AutoVue Innova exhibited the highest sensitivity in the titrating study and throughput performance compared with the manual method and the Techno TwinStation. Especially in the throughput and time required to complete 100 antibody screening tests, AutoVue Innova had a 3.3- and 3.5-fold higher performance, respectively, than Techno TwinStation.
CONCLUSIONS: Because both of the two fully automated instruments (AutoVue Innova and Techno TwinStation) had high levels of accuracy and performance, it is expected that use of fully automated instruments will reduce human labor, turnaround time, and operator error in the blood bank.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18594174     DOI: 10.3343/kjlm.2008.28.3.214

Source DB:  PubMed          Journal:  Korean J Lab Med        ISSN: 1598-6535


  7 in total

1.  Performance Evaluation of Automated Immunohematology Analyzer IH-500 for Blood Bank Testing.

Authors:  Sang Hyuk Park; Jaewook Kim; Ji-Hun Lim; Joseph Jeong; Seon-Ho Lee
Journal:  Indian J Hematol Blood Transfus       Date:  2019-05-02       Impact factor: 0.900

2.  Ebola Virus Inactivation by Detergents Is Annulled in Serum.

Authors:  Jeroen J A van Kampen; Andrei Tintu; Henk Russcher; Pieter L A Fraaij; Chantal B E M Reusken; Mikel Rijken; Jaap J van Hellemond; Perry J J van Genderen; Rob Koelewijn; Menno D de Jong; Elaine Haddock; Robert J Fischer; Vincent J Munster; Marion P G Koopmans
Journal:  J Infect Dis       Date:  2017-10-17       Impact factor: 5.226

3.  Economic and workflow analysis of a blood bank automated system.

Authors:  Kyung-Hwa Shin; Hyung Hoi Kim; Chulhun L Chang; Eun Yup Lee
Journal:  Ann Lab Med       Date:  2013-06-24       Impact factor: 3.464

4.  Comparative Analysis of Clinical Samples Showing Weak Serum Reaction on AutoVue System Causing ABO Blood Typing Discrepancies.

Authors:  Su Yeon Jo; Ju Mi Lee; Hye Lim Kim; Kyeong Hwa Sin; Hyeon Ji Lee; Chulhun Ludgerus Chang; Hyung Hoi Kim
Journal:  Ann Lab Med       Date:  2017-03       Impact factor: 3.464

5.  A comparison of three column agglutination tests for red blood cell alloantibody identification.

Authors:  Siska Blomme; Emilie De Maertelaere; Eline Verhoye
Journal:  BMC Res Notes       Date:  2020-03-04

6.  Detection of ABO Discrepancy in a Case of Coronary Artery Disease by Conventional Tube Technique: A Miss by Column Agglutination Technology.

Authors:  Suman Sudha Routray; Satya Prakash; Gopal Krushna Ray; Somnath Mukherjee; Ansuman Sahu
Journal:  J Lab Physicians       Date:  2022-01-18

7.  Comparison Between Conventional and Automated Techniques for Blood Grouping and Crossmatching: Experience from a Tertiary Care Centre.

Authors:  Swarupa Nikhil Bhagwat; Jayashree H Sharma; Julie Jose; Charusmita J Modi
Journal:  J Lab Physicians       Date:  2015 Jul-Dec
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.