An experience of the introduction of a blood bank automation system (Ortho AutoVue Innova) in a regional acute hospital.

Yuk Wah Cheng1, Jenny M Wilkinson2.   

Abstract

This paper reports on an evaluation of the introduction of a blood bank automation system (Ortho AutoVue(®) Innova) in a hospital blood bank by considering the performance and workflow as compared with manual methods. The turnaround time was found to be 45% faster than the manual method. The concordance rate was found to be 100% for both ABO/Rh(D) typing and antibody screening in both of the systems and there was no significant difference in detection sensitivity for clinically significant antibodies. The Ortho AutoVue(®) Innova automated blood banking system streamlined the routine pre-transfusion testing in hospital blood bank with high throughput, equivalent sensitivity and reliability as compared with conventional manual method.
Copyright © 2015 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Blood bank automation; Hospital blood bank; Ortho AutoVue(®) Innova; Type & screen

Mesh:

Substances:

Year:  2015        PMID: 25863409     DOI: 10.1016/j.transci.2015.03.007

Source DB:  PubMed          Journal:  Transfus Apher Sci        ISSN: 1473-0502            Impact factor:   1.764


  2 in total

1.  Performance Evaluation of Automated Immunohematology Analyzer IH-500 for Blood Bank Testing.

Authors:  Sang Hyuk Park; Jaewook Kim; Ji-Hun Lim; Joseph Jeong; Seon-Ho Lee
Journal:  Indian J Hematol Blood Transfus       Date:  2019-05-02       Impact factor: 0.900

2.  Comparative Analysis of Clinical Samples Showing Weak Serum Reaction on AutoVue System Causing ABO Blood Typing Discrepancies.

Authors:  Su Yeon Jo; Ju Mi Lee; Hye Lim Kim; Kyeong Hwa Sin; Hyeon Ji Lee; Chulhun Ludgerus Chang; Hyung Hoi Kim
Journal:  Ann Lab Med       Date:  2017-03       Impact factor: 3.464

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.