Point of care hemoglobin testing in plateletpheresis donors: Noninvasive versus invasive methods.

R S Mallhi1, Amit Pawar2, Neerja Kushwaha3, Sudeep Kumar3, Ujjwal Dimri2.   

Abstract

BACKGROUND: Predonation hemoglobin (Hb) for plateletpheresis donors is estimated by presently available invasive methods. Venous samples of only those donors whose Hb is more than 12.5 g/dl are screened for complete blood count and transfusion transmissible infections. There is a pressing need to substitute this invasive Hb test with noninvasive one to reduce donor discomfort and avoid further pricking the donor. We therefore went ahead with the aim of comparing a noninvasive Hb estimation method NBM 200 with the invasive method - Hemocue, taking SYSMEX KX-21 as a gold standard.
METHODS: 500 voluntary consenting plateletpheresis donors qualifying the laid down criteria for platelet donation were included in the study. Hb readings obtained by the NBM 200 and Hemocue were compared to those obtained from the fully automated hematology analyzer SYSMEX KX-21.
RESULTS: Coefficients of correlation were found to be statistically significant at <0.0001 level of confidence. Results of Friedman's test on the three methods also showed significant difference in means. Bland-Altman plots and mountain plots also confirm the same. NBM 200 was found to be more sensitive, specific, and precise than Hemocue in detecting ineligible donors.
CONCLUSION: NBM 200 was found to be more sensitive, specific, and precise as compared to Hemocue for predonation screening of Hb in plateletpheresis donors and the prime benefit it offers is that it is 'noninvasive' thereby assisting in stemming the platelet donor pool. The onus lies on the blood transfusion services to make use of appropriately validated gadgets that reduce the donor discomfort.

Entities:  

Keywords:  Hemocue; Hemoglobin estimation; Invasive versus noninvasive; NBM 200; SYSMEX KX-21

Year:  2016        PMID: 27843180      PMCID: PMC5099441          DOI: 10.1016/j.mjafi.2016.06.013

Source DB:  PubMed          Journal:  Med J Armed Forces India        ISSN: 0377-1237


  10 in total

1.  Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors.

Authors:  Abhay Singh; Anju Dubey; Atul Sonker; Rajendra Chaudhary
Journal:  Blood Transfus       Date:  2014-10-29       Impact factor: 3.443

2.  The difference between fingerstick and venous hemoglobin and hematocrit varies by sex and iron stores.

Authors:  Ritchard G Cable; Whitney R Steele; Russell S Melmed; Bryce Johnson; Alan E Mast; Patricia M Carey; Joseph E Kiss; Steven H Kleinman; David J Wright
Journal:  Transfusion       Date:  2011-10-20       Impact factor: 3.157

3.  The new noninvasive occlusion spectroscopy hemoglobin measurement method: a reliable and easy anemia screening test for blood donors.

Authors:  Márcio Pinto; Maria Lourdes Barjas-Castro; Simone Nascimento; Mônica Almeida Falconi; Roberto Zulli; Vagner Castro
Journal:  Transfusion       Date:  2012-07-15       Impact factor: 3.157

4.  Noninvasive methods for haemoglobin screening in prospective blood donors.

Authors:  A Belardinelli; M Benni; P L Tazzari; P Pagliaro
Journal:  Vox Sang       Date:  2013-04-19       Impact factor: 2.144

5.  Comparison of three noninvasive methods for hemoglobin screening of blood donors.

Authors:  Sergey Ardin; Melanie Störmer; Stela Radojska; Larissa Oustianskaia; Moritz Hahn; Birgit S Gathof
Journal:  Transfusion       Date:  2014-08-15       Impact factor: 3.157

6.  A review of blood donor motivation and recruitment.

Authors:  R M Oswalt
Journal:  Transfusion       Date:  1977 Mar-Apr       Impact factor: 3.157

7.  Continuous noninvasive hemoglobin monitoring during complex spine surgery.

Authors:  Lauren Berkow; Stephanie Rotolo; Erin Mirski
Journal:  Anesth Analg       Date:  2011-09-29       Impact factor: 5.108

8.  Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender.

Authors:  John R Feiner; John W Severinghaus; Philip E Bickler
Journal:  Anesth Analg       Date:  2007-12       Impact factor: 5.108

9.  Comparison of the accuracy of noninvasive hemoglobin sensor (NBM-200) and portable hemoglobinometer (HemoCue) with an automated hematology analyzer (LH500) in blood donor screening.

Authors:  Moon Jung Kim; Quehn Park; Myung Hee Kim; Jeong Won Shin; Hyun Ok Kim
Journal:  Ann Lab Med       Date:  2013-06-24       Impact factor: 3.464

10.  Analysis of efforts to maintain safe donor in main donor pool after completion of temporary deferral period.

Authors:  Ripal Shah; Sunita Tulsiani; V Harimoorthy; Ankit Mathur; Nabajyoti Choudhury
Journal:  Asian J Transfus Sci       Date:  2013-01
  10 in total
  1 in total

1.  Reproducibility and validity of portable haemoglobinometer for the diagnosis of anaemia in children under the age of 5 years.

Authors:  Alessandra da Silva Pereira; Inês Rugani Ribeiro de Castro; Flávia Fioruci Bezerra; José Firmino Nogueira Neto; Ana Carolina Feldenheimer da Silva
Journal:  J Nutr Sci       Date:  2020-01-20
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.