| Literature DB >> 32042411 |
Alessandra da Silva Pereira1, Inês Rugani Ribeiro de Castro2, Flávia Fioruci Bezerra3, José Firmino Nogueira Neto4, Ana Carolina Feldenheimer da Silva2.
Abstract
Portable haemoglobinometers have been used in order to estimate the prevalence of anaemia in diverse settings. However, few studies have been conducted to evaluate their performance in children of different age groups in distinct epidemiological contexts. To evaluate the reproducibility and reliability of a portable haemoglobinometer for the diagnosis of anaemia in children <5 years Hb was measured in the venous blood of 351 children <5 years by an automated system (standard method) and in three capillary blood samples, using a portable haemoglobinometer (HemoCue®; test method). The reproducibility of the device and of the test method was evaluated using the intraclass correlation coefficient (ICC) (Hb in its continuous form), κ and prevalence-adjusted bias-adjusted κ (PABAK) (categorised variable: anaemia: yes/no). For test method validation, Bland-Altman analyses were performed and sensitivity, specificity, accuracy rate, positive predictive value (PPV) and negative predictive values (NPV) were calculated. The haemoglobinometer presented good device reproducibility (ICC = 0·79) and reasonable method reproducibility (puncture, collection and reading) (ICC = 0·71). Superficial and fair agreement (κ) and good agreement (PABAK) were observed among the diagnoses obtained through the test method. The prevalence of anaemia was 19·1 and 19·7 % using the standard and the test method, respectively, with no statistically significant differences. The test method presented higher specificity (87·7 %) and NPV (88·3 %) than sensitivity (50·7 %) and PPV (49·3 %), and intermediary accuracy rate (57·8 %). HemoCue® showed good device reproducibility and reasonable method reproducibility, as well as good performance in estimating the prevalence of anaemia. Nevertheless, it showed a fair reliability and low individual diagnostic accuracy.Entities:
Keywords: Accuracy; Anaemia; Children; ICC, intraclass correlation coefficient; NPV, negative predictive value; PABAK, prevalence-adjusted bias-adjusted κ; PPV, positive predictive value; Reproducibility
Mesh:
Substances:
Year: 2020 PMID: 32042411 PMCID: PMC6984124 DOI: 10.1017/jns.2019.43
Source DB: PubMed Journal: J Nutr Sci ISSN: 2048-6790
Fig. 1.Distribution of Hb concentration of three capillary blood samples obtained through a portable haemoglobinometer in children aged 6 to 59 months (Rio de Janeiro, 2014). n 357 (sample test 1 (––) v. sample test 2 (---)) and n 355 (sample test 1 v. sample test 3 (– –)). Sample test 1: first drop of capillary blood, extracted from the first puncture, read on a portable haemoglobinometer. Sample test 2: second drop of capillary blood, extracted from the first puncture, read on a portable haemoglobinometer. Sample test 3: third drop of capillary blood, extracted from the second puncture, read on a portable haemoglobinometer. (a) Total group; (b) < 2 years old; (c) ≥ 2 years old. * To convert Hb in g/dl to mmol/l, multiply by 0·6206.
Hb concentration, intraclass correlation coefficient, κ and adjusted κ for Hb determination and diagnosis of anaemia by portable haemoglobinometer in children aged 6 to 59 months by age group (Rio de Janeiro, 2014)*
| Hb (g/dl) | Intraclass correlation coefficient | ||||
|---|---|---|---|---|---|
| Test method | Mean | κ | Adjusted κ | ||
| Children <2 years | |||||
| Sample test 1 | 11·70 | 1·06 | – | – | – |
| Sample test 2 | 11·57 | 1·16 | 0·81 | 0·46 | 0·54 |
| Sample test 3 | 11·67 | 1·15 | 0·68 | 0·42 | 0·61 |
| Children ≥2 years | |||||
| Sample test 1 | 12·31 | 1·08 | |||
| Sample test 2 | 12·18 | 1·06 | 0·75 | 0·28 | 0·69 |
| Sample test 3 | 12·28 | 0·98 | 0·68 | 0·18 | 0·67 |
| Total | |||||
| Sample test 1 | 12·04 | 1·11 | – | – | – |
| Sample test 2 | 11·70 | 1·15 | 0·79 | 0·43 | 0·62 |
| Sample test 3 | 12·00 | 1·10 | 0·71 | 0·36 | 0·63 |
n 357 (test 1 v. test 2) and 355 (tests 1 v. 3). Test method: Hb concentration in capillary blood sample analysed by portable haemoglobinometer (HemoCue® Hb 301).
To convert Hb in g/dl to mmol/l, multiply by 0·6206.
Sample test 1: 1st drop of capillary blood, extracted from the first puncture, read on a portable haemoglobinometer.
Sample test 2: 2nd drop of capillary blood, extracted from the first puncture, read on a portable haemoglobinometer.
Sample test 3: 3rd drop of capillary blood, extracted from the second puncture, read on a portable haemoglobinometer.
Fig. 2.Distribution of Hb concentration according to the standard method (––) and the test method (---) carried out on children aged 6 to 59 months by age group (Rio de Janeiro, 2014). n 351. Standard method: Hb analysis in venous blood samples by automated system (XS1000i Sysmex®). Test method: Hb concentration in capillary blood sample analysed by portable haemoglobinometer (HemoCue® Hb 301). (a) Total group; (b) < 2 years old; (c) ≥ 2 years old. * To convert Hb in g/dl to mmol/l, multiply by 0·6206.
Hb concentration and prevalence of anaemia according to the standard method and the test method in children aged 6 to 59 months (n 351) by age group (Rio de Janeiro, 2014)
| Method | Hb (g/dl) | Prevalence of anaemia (%) | ||||
|---|---|---|---|---|---|---|
| Mean | Median | Variance | Interquartile range | |||
| Children <2 years | ||||||
| Standard | 11·52 | 0·99 | 11·50 | 1·00 | 1·40 | 32·9 |
| Test | 11·70 | 1·06 | 11·70 | 1·13 | 1·40 | 29·1 |
| Children ≥2 years | ||||||
| Standard | 12·14 | 0·87 | 12·10 | 0·75 | 1·20 | 7·8 |
| Test | 12·31 | 1·08 | 12·40 | 1·18 | 1·60 | 11·9 |
| Total | ||||||
| Standard | 11·86 | 0·97 | 11·80 | 0·95 | 1·30 | 19·1 |
| Test | 12·04 | 1·11 | 12·00 | 1·25 | 1·50 | 19·7 |
To convert Hb in g/dl to mmol/l, multiply by 0·6206.
Standard method: Hb concentration in venous blood sample analysed by automated system (XS1000i Sysmex®).
No statistically signficant difference (P > 0·05) between prevalences of anaemia obtained by the standard method and the test method (McNemar's χ2 test).
Test method: Hb concentration in capillary blood sample analysed by portable haemoglobinometer (HemoCue® Hb 301).
Statistically significant difference (P < 0·05) between the mean Hb concentration obtained by the standard method and the test method (Student's t test for paired samples).
Fig. 3.Bland−Altman plot showing agreement in Hb concentration of children aged 6 to 59 months of Rio de Janeiro assessed by test method (Hb concentration in capillary blood sample analysed by portable haemoglobinometer: HemoCue® Hb 301) and standard method (Hb analysis in venous blood samples by automated system: XS1000i Sysmex®). * To convert Hb in g/dl to mmol/l, multiply by 0·6206.
Sensitivity, specificity, accuracy index, positive predictive value and negative predictive value of the portable haemoglobinometer compared with the automated system for estimating anaemia in children aged 6 to 59 months according to age group (Rio de Janeiro, 2014)
| Age group | Sensitivity (%) | Specificity (%) | Accuracy index (%) | Positive predictive value (%) | Negative predictive value (%) |
|---|---|---|---|---|---|
| Children <2 years ( | 56·5 | 76·7 | 73·6 | 50·0 | 81·1 |
| Children ≥2 years ( | 34·8 | 95·9 | 36·3 | 53·3 | 91·6 |
| Total ( | 50·7 | 87·7 | 57·8 | 49·3 | 88·3 |