Noninvasive methods for haemoglobin screening in prospective blood donors.

A Belardinelli1, M Benni, P L Tazzari, P Pagliaro.   

Abstract

BACKGROUND AND OBJECTIVES: The haemoglobin level of prospective blood donors is usually performed on blood obtained by from the finger pulp by fingerstick with a lancet and filling a capillary tube with a sample. New noninvasive methods are now available for rapid, noninvasive predonation haemoglobin screening.
MATERIALS AND METHODS: Prospective blood donors at our blood centre were tested, in two different trials, as follows: by the NBM 200 (OrSense) test (n = 445 donors) and by the Pronto-7 (Masimo) test (n = 463 donors). The haemoglobin values of each trial and the haemoglobin of finger pulp blood obtained by fingerstick with a lancet (HemoCue) were compared with the haemoglobin values obtained from a venous sample on a Cell Counter (Beckman Coulter).
RESULTS: Comparison of Beckman Coulter Cell Counter and OrSense and results showed a bias of 0.29 g/dl, the standard deviation of the differences (SDD) of 0.98 and 95% limits of agreement from -1.64 to 2.21, using Bland and Altman statistical methodology. Comparison of Masimo and Beckman Coulter Cell Counter results showed a bias of -0.53 g/dl, SDD of 1.04 and 95% limits of agreement from -2.57 to 1.51. Cumulative analysis of all 908 donors, as tested by the usual fingerstick test showed a bias of 0.83 g/dl, SDD of 0.70 and 95% limits of agreement from -0.54 to 2.20 compared with the Coulter Cell Counter. Compared with the Coulter Counter, the specificity of the methods was 99.5% for fingerstick, 97% for OrSense and 83% for Massimo, and the sensitivity was 99, 98 and 93%, respectively.
CONCLUSIONS: Analysis of finger pulp blood by either direct sampling by fingerstick and Hemocue, or by noninvasive haemoglobin tests does not replicate the results of cell counter analysis of venous samples. Compared with fingerstick, noninvasive haemoglobin tests eliminate pain and reduce stress, but have a lower level of specificity and sensitivity.
© 2013 International Society of Blood Transfusion.

Entities:  

Keywords:  blood collection; blood donation testing; donors; motivation/recruitment

Mesh:

Substances:

Year:  2013        PMID: 23600766     DOI: 10.1111/vox.12033

Source DB:  PubMed          Journal:  Vox Sang        ISSN: 0042-9007            Impact factor:   2.144


  13 in total

1.  Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors.

Authors:  Abhay Singh; Anju Dubey; Atul Sonker; Rajendra Chaudhary
Journal:  Blood Transfus       Date:  2014-10-29       Impact factor: 3.443

2.  Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Authors:  Matthias L Riess; Paul S Pagel
Journal:  J Cardiothorac Vasc Anesth       Date:  2016-03-24       Impact factor: 2.628

3.  Point of care hemoglobin testing in plateletpheresis donors: Noninvasive versus invasive methods.

Authors:  R S Mallhi; Amit Pawar; Neerja Kushwaha; Sudeep Kumar; Ujjwal Dimri
Journal:  Med J Armed Forces India       Date:  2016-08-11

Review 4.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

5.  A non-invasive strategy for haemoglobin screening of blood donors.

Authors:  Pasqualepaolo Pagliaro; Annarita Belardinelli; Viana Boko; Petar Salamon; Silvia Manfroi; Pier Luigi Tazzari
Journal:  Blood Transfus       Date:  2014-06-05       Impact factor: 3.443

6.  Noninvasive Assessment of Excessive Erythrocytosis as a Screening Method for Chronic Mountain Sickness at High Altitude.

Authors:  Kaetan J Vyas; David Danz; Robert H Gilman; Robert A Wise; Fabiola León-Velarde; J Jaime Miranda; William Checkley
Journal:  High Alt Med Biol       Date:  2015-05-14       Impact factor: 1.981

Review 7.  Emerging point-of-care technologies for anemia detection.

Authors:  Ran An; Yuning Huang; Yuncheng Man; Russell W Valentine; Erdem Kucukal; Utku Goreke; Zoe Sekyonda; Connie Piccone; Amma Owusu-Ansah; Sanjay Ahuja; Jane A Little; Umut A Gurkan
Journal:  Lab Chip       Date:  2021-05-18       Impact factor: 6.799

8.  An instantaneous low-cost point-of-care anemia detection device.

Authors:  Jaime Punter-Villagrasa; Joan Cid; Cristina Páez-Avilés; Ivón Rodríguez-Villarreal; Esteve Juanola-Feliu; Jordi Colomer-Farrarons; Pere Ll Miribel-Català
Journal:  Sensors (Basel)       Date:  2015-02-16       Impact factor: 3.576

Review 9.  Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions.

Authors:  Rajendra Chaudhary; Anju Dubey; Atul Sonker
Journal:  J Blood Med       Date:  2017-07-03

10.  Noninvasive hematocrit assessment for cardiovascular magnetic resonance extracellular volume quantification using a point-of-care device and synthetic derivation.

Authors:  Sean Robison; Gauri Rani Karur; Rachel M Wald; Paaladinesh Thavendiranathan; Andrew M Crean; Kate Hanneman
Journal:  J Cardiovasc Magn Reson       Date:  2018-03-15       Impact factor: 5.364