Comparison of three noninvasive methods for hemoglobin screening of blood donors.

Sergey Ardin1, Melanie Störmer, Stela Radojska, Larissa Oustianskaia, Moritz Hahn, Birgit S Gathof.   

Abstract

BACKGROUND: To prevent phlebotomy of anemic individuals and to ensure hemoglobin (Hb) content of the blood units, Hb screening of blood donors before donation is essential. Hb values are mostly evaluated by measurement of capillary blood obtained from fingerstick. Rapid noninvasive methods have recently become available and may be preferred by donors and staff. The aim of this study was to evaluate for the first time all different noninvasive methods for Hb screening. STUDY DESIGN AND METHODS: Blood donors were screened for Hb levels in three different trials using three different noninvasive methods (Haemospect [MBR Optical Systems GmbH & Co. KG], NBM 200 [LMB Technology GmbH], Pronto-7 [Masimo Europe Ltd]) in comparison to the established fingerstick method (CompoLab Hb [Fresenius Kabi GmbH]) and to levels obtained from venous samples on a cell counter (Sysmex [Sysmex Europe GmbH]) as reference. The usability of the noninvasive methods was assessed with an especially developed survey.
RESULTS: Technical failures occurred by using the Pronto-7 due to nail polish, skin color, or ambient light. The NBM 200 also showed a high sensitivity to ambient light and noticeably lower Hb levels for women than obtained from the Sysmex. The statistical analysis showed the following bias and standard deviation of differences of all methods in comparison to the venous results: Haemospect, -0.22 ± 1.24; NBM, 200 -0.12 ± 1.14; Pronto-7, -0.50 ± 0.99; and CompoLab Hb, -0.53 ± 0.81.
CONCLUSION: Noninvasive Hb tests represent an attractive alternative by eliminating pain and reducing risks of blood contamination. The main problem for generating reliable results seems to be preanalytical variability in sampling. Despite the sensitivity to environmental stress, all methods are suitable for Hb measurement.
© 2014 AABB.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25123051     DOI: 10.1111/trf.12819

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  12 in total

1.  Comparison of noninvasive and invasive point-of-care testing methods with reference method for hemoglobin measurement.

Authors:  Gamze Avcioglu; Cemil Nural; Fatma Meriç Yilmaz; Pervin Baran; Özcan Erel; Gülsen Yilmaz
Journal:  J Clin Lab Anal       Date:  2017-08-23       Impact factor: 2.352

2.  Point of care hemoglobin testing in plateletpheresis donors: Noninvasive versus invasive methods.

Authors:  R S Mallhi; Amit Pawar; Neerja Kushwaha; Sudeep Kumar; Ujjwal Dimri
Journal:  Med J Armed Forces India       Date:  2016-08-11

Review 3.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

4.  Poor Validity of Noninvasive Hemoglobin Measurements by Pulse Oximetry Compared with Conventional Absorptiometry in Children in Côte d'Ivoire.

Authors:  Isaac I Bogoch; Jean T Coulibaly; Jason Rajchgot; Jason R Andrews; Jana Kovac; Jürg Utzinger; Gordana Panic; Jennifer Keiser
Journal:  Am J Trop Med Hyg       Date:  2016-11-14       Impact factor: 2.345

Review 5.  Emerging point-of-care technologies for anemia detection.

Authors:  Ran An; Yuning Huang; Yuncheng Man; Russell W Valentine; Erdem Kucukal; Utku Goreke; Zoe Sekyonda; Connie Piccone; Amma Owusu-Ansah; Sanjay Ahuja; Jane A Little; Umut A Gurkan
Journal:  Lab Chip       Date:  2021-05-18       Impact factor: 6.799

6.  An evaluation of hemoglobin measurement tools and their accuracy and reliability when screening for child anemia in Rwanda: A randomized study.

Authors:  Megan Parker; Zhen Han; Elizabeth Abu-Haydar; Eric Matsiko; Damien Iyakaremye; Lisine Tuyisenge; Amalia Magaret; Alexandre Lyambabaje
Journal:  PLoS One       Date:  2018-01-04       Impact factor: 3.240

Review 7.  Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions.

Authors:  Rajendra Chaudhary; Anju Dubey; Atul Sonker
Journal:  J Blood Med       Date:  2017-07-03

8.  Accuracy of Capillary Hemoglobin Measurements for the Detection of Anemia among U.S. Low-Income Toddlers and Pregnant Women.

Authors:  Safia Boghani; Zuguo Mei; Geraldine S Perry; Gary M Brittenham; Mary E Cogswell
Journal:  Nutrients       Date:  2017-03-09       Impact factor: 5.717

9.  Noninvasive hematocrit assessment for cardiovascular magnetic resonance extracellular volume quantification using a point-of-care device and synthetic derivation.

Authors:  Sean Robison; Gauri Rani Karur; Rachel M Wald; Paaladinesh Thavendiranathan; Andrew M Crean; Kate Hanneman
Journal:  J Cardiovasc Magn Reson       Date:  2018-03-15       Impact factor: 5.364

10.  Evaluation of new non-invasive & conventional invasive methods of haemoglobin estimation in blood donors.

Authors:  Diptiranjan Rout; Suchet Sachdev; Neelam Marwaha
Journal:  Indian J Med Res       Date:  2019-06       Impact factor: 2.375

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.