BACKGROUND: Evaluation of set up error detection by a transperineal ultrasound in comparison with a cone beam CT (CBCT) based system in external beam radiation therapy (EBRT) of prostate cancer.Entities:
Keywords: Cone beam CT; IGRT; Prostate cancer; Setup verification; Transperineal ultrasound
Mesh:
Year: 2016 PMID: 27825386 PMCID: PMC5101794 DOI: 10.1186/s13014-016-0722-7
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1Schematic illustration of clinical workflow with setup verification using Conebeam CT and transperineal US
Fig. 2Schematic patient setup for transperineal ultrasound
Summary of setup errors: mean, standard deviation, median and range
| Dimension | Mean ± STD | Median | Range |
|---|---|---|---|
| Lateral | |||
| US imaging | −0.4 ± 3.8 | −0.3 | −12.6 → 10.3 |
| CBCT imaging | 1.1 ± 3.6 | 1.4 | −7.3 → 12.1 |
| Absolute difference of US and CBCT | 2.7 ± 2.3 | 2.5 | 0 → 10.3 |
| Vertical | |||
| US imaging | 0.8 ± 4.1 | 1.0 | −12.6 → 11.0 |
| CBCT imaging | 0.3 ± 3.3 | 1.2 | −12.0 → 7.7 |
| Absolute difference of US and CBCT | 3.0 ± 2.4 | 2.5 | 0 → 11.8 |
| Longitudinal | |||
| US imaging | −0.5 ± 4.4 | −0.8 | −11.9 → 11.3 |
| CBCT imaging | 0.9 ± 3.2 | 0.6 | −10.3 → 9.7 |
| Absolute difference of US and CBCT | 3.2 ± 2.7 | 2.6 | 0 → 16.5 |
| 3D | |||
| US imaging | 6.0 ± 2.9 | 6.3 | 0.8 → 16.9 |
| CBCT imaging | 5.4 ± 2.7 | 4.9 | 0.6 → 13.3 |
| Absolute difference of US and CBCT | 6.0 ± 3.1 | 5.6 | 1.1 → 19.2 |
In units of mm
Fig. 3Setup errors measured with US and CBCT imaging shown in correlation plots with line of equality (a, b, c) and Bland-Altmann plots (d, e, f)
Fig. 4Boxplots of setup errors for (a) US, (b) CBCT imaging and (c) CT-US in lateral, vertical and longitudinal dimension