Ultrasound tracking for intra-fractional motion compensation in radiation therapy.

J Schwaab1, M Prall2, C Sarti3, R Kaderka2, C Bert2, C Kurz4, K Parodi5, M Günther6, J Jenne6.   

Abstract

Modern techniques as ion beam therapy or 4D imaging require precise target position information. However, target motion particularly in the abdomen due to respiration or patient movement is still a challenge and demands methods that detect and compensate this motion. Ultrasound represents a non-invasive, dose-free and model-independent alternative to fluoroscopy, respiration belt or optical tracking of the patient surface. Thus, ultrasound based motion tracking was integrated into irradiation with actively scanned heavy ions. In a first in vitro experiment, the ultrasound tracking system was used to compensate diverse sinusoidal target motions in two dimensions. A time delay of ∼200 ms between target motion and reported position data was compensated by a prediction algorithm (artificial neural network). The irradiated films proved feasibility of the proposed method. Furthermore, a practicable and reliable calibration workflow was developed to enable the transformation of ultrasound tracking data to the coordinates of the treatment delivery or imaging system - even if the ultrasound probe moves due to respiration. A first proof of principle experiment was performed during time-resolved positron emission tomography (4DPET) to test the calibration workflow and to show the accuracy of an ultrasound based motion tracking in vitro. The results showed that optical ultrasound tracking can reach acceptable accuracies and encourage further research.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Motion correction; Particle therapy; Ultrasound

Mesh:

Year:  2014        PMID: 24695273     DOI: 10.1016/j.ejmp.2014.03.003

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  8 in total

1.  [Ultrasound motion tracking for radiation therapy].

Authors:  J Jenne; J Schwaab
Journal:  Radiologe       Date:  2015-11       Impact factor: 0.635

2.  Analysis and optimization of the robot setup for robotic-ultrasound-guided radiation therapy.

Authors:  Matthias Schlüter; Stefan Gerlach; Christoph Fürweger; Alexander Schlaefer
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-06-06       Impact factor: 2.924

3.  Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology.

Authors:  Edward D Brandner; Indrin J Chetty; Tawfik G Giaddui; Ying Xiao; M Saiful Huq
Journal:  Med Phys       Date:  2017-04-20       Impact factor: 4.071

4.  Therapeutic Ultrasound Effects on Human Induced Pluripotent Stem Cell Cardiomyocytes Measured Optically and with Spectral Ultrasound.

Authors:  Andrew W Chen; George Saab; Aleksandar Jeremic; Vesna Zderic
Journal:  Ultrasound Med Biol       Date:  2022-03-15       Impact factor: 3.694

Review 5.  Particle therapy of moving targets-the strategies for tumour motion monitoring and moving targets irradiation.

Authors:  Tomasz Kubiak
Journal:  Br J Radiol       Date:  2016-07-19       Impact factor: 3.039

6.  Evaluation of daily patient positioning for radiotherapy with a commercial 3D surface-imaging system (Catalyst™).

Authors:  F Walter; P Freislederer; C Belka; C Heinz; M Söhn; F Roeder
Journal:  Radiat Oncol       Date:  2016-11-24       Impact factor: 3.481

7.  AI-based optimization for US-guided radiation therapy of the prostate.

Authors:  Stefan Gerlach; Theresa Hofmann; Christoph Fürweger; Alexander Schlaefer
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-05-20       Impact factor: 3.421

8.  Initial results for patient setup verification using transperineal ultrasound and cone beam CT in external beam radiation therapy of prostate cancer.

Authors:  Anne Richter; Bülent Polat; Ingulf Lawrenz; Stefan Weick; Otto Sauer; Michael Flentje; Frederick Mantel
Journal:  Radiat Oncol       Date:  2016-11-08       Impact factor: 3.481