Justin T Sick1, Nicholas J Rancilio2, Caroline V Fulkerson2, Jeannie M Plantenga1,2,3, Deborah W Knapp2,3, Keith M Stantz1,4.
Abstract
BACKGROUND ANDEntities:
Keywords: 3D printing; Agar-based tissue equivalent phantom; Bladder cancer; Multimodal registration; Ultrasound guided radiation therapy
Year: 2019 PMID: 33458289 PMCID: PMC7807639 DOI: 10.1016/j.phro.2019.10.003
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Platform Design. (A) Overview of the platform design, (B) Different possible discrete angles of the transducer, (C) Longitudinal and lateral range, and (D) different coordinate systems of the US guidance platform and the RT treatment bed.
Fig. 2Images of the urological phantom. (A) CT slice of phantom, (B) corresponding US slice.
Physical Properties of the Tissue-Equivalent Bladder Phantom.
| Recipe | Measured | ||||
|---|---|---|---|---|---|
| Glycerine (%) | Agar (%) | CT Number (HU) | Density (g/cm3) | Speed of Sound (mm/μs) | |
| Water | 0 | 0 | −1.2 ± 2.6 | 0.99 | 1.49 |
| Skin | 50 | 2 | 107.3 ± 3.5 | 1.10 | 1.70 |
| Fat | 0 | 2 | 2.6 ± 2.6 | 0.99 | 1.49 |
| Muscle | 25 | 2 | 69.6 ± 3.9 | 1.06 | 1.59 |
| Urine | 0 | 0 | 41.5 ± 0.8 | 1.03 | 1.55 |
Level of user intervention required to achieve a quality registration.
| CT-to-CT | US-to-US | US-to-CT | |
|---|---|---|---|
| Wedge Only | Automatic | Automatic | Manual |
| 5 | Automatic | Automatic | Manual |
| 15 | Automatic | Initial manual alignment | Manual |
| 45 | Automatic | Initial manual alignment | Manual |
*Automatic, The automatic registration successfully registered the datasets every time.
**Initial manual alignment, the user had to only roughly shift and rotate the target dataset before applying the automatic registration. Once a manual rough alignment was done, the automatic registration would be successful.
***Manual, the user had to apply multiple techniques to register the datasets. The automatic registration was never successful.
Fig. 3Displayed are the CT-to-CT, US-to-CT, and US-to-US registration results of the bladder tissue-equivalent phantom after displacement (translation and rotation). The first column are the images prior to registration, the second column represents the displaced image in blue; the final column the images after registration. (A-C) CT-CT registration between the phantom rotated to 45° and the flat position. (D–F) US-CT registration between the phantom rotated to 45° and the flat position. (G-I) US-US registration between the phantom rotated to 45° and the flat position. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Comparison of the Phantom Physical Properties to those in the Literature.
| Density (g/cm3) | Speed of Sound (mm/μs) | |||
|---|---|---|---|---|
| Measured | Literature | Measured | Literature | |
| Water | 0.99 | 0.998 | 1.49 | 1.483 |
| Skin | 1.10 | 1.110–1.119 | 1.70 | 1.590–2.170 |
| Fat | 0.99 | 0.916 | 1.49 | 1.412–1.487 |
| Muscle | 1.06 | 1.038–1.056 | 1.59 | 1.589–1.603 |
| Urine | 1.03 | 1.001–1.050 | 1.55 | 1.554 |