Richard A P Takx1, Radko Krissak2,3, Christian Fink2,4, Valentin Bachmann2, Thomas Henzler2, Mathias Meyer2, John W Nance5, Stefan O Schoenberg2, Paul Apfaltrer2,6.
Abstract
OBJECTIVES: To investigate the relationship between image quality and patient size at 100 kilovoltage (kV) compared to 120 kV ECG-gated Triple-Rule-Out CT angiography (TRO-CTA).Entities:
Keywords: Body mass index; Chest pain; Image enhancement; Multidetector computed tomography; Radiation dosage
Mesh:
Year: 2016 PMID: 27686566 PMCID: PMC5408040 DOI: 10.1007/s00330-016-4607-9
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Example image of measurements (a) ROI in ascending aorta and pulmonary artery. b Left main coronary artery and epicardial fat. c left and right ventricle. d anthropomorphic measures at level of the origin of the left main coronary artery
Patient characteristics and radiation dose
| 100 kV ( | 120 kV ( |
| |
|---|---|---|---|
| BMI (kg/m2) | 26.0 ± 4.6 | 28.0 ± 6.7 | 0.138 |
| <25 | 19 | 14 | |
| 25.1 – 30 | 17 | 11 | |
| >30.1 | 4 | 8 | |
| CTDI (mGy) | 19.2 ± 6.9 | 41.3 ± 21.5 | <0.001 |
| mAs | 162.4 ± 55.6 | 192.9 ± 70.8 | 0.044 |
| DLP (mGy cm) | 630 ± 240 | 1215 ± 626 | <0.001 |
Fig. 2Correlation plot between the various anthropomorphic measures. a PC and BMI, b TSA and BMI, and c PC and TSA
Mean attenuation and noise
| 100 kV ( | 120 kV ( |
| p-value adjusted for BMI | p-value adjusted for PC | p-value adjusted for TSA | |
|---|---|---|---|---|---|---|
| Attenuation | ||||||
| Pulmonary artery | 459.8 ± 107.2 | 348.6 ± 85.3 | <0.001 | <0.001 | <0.001 | <0.001 |
| Ascending Aorta | 435.7 ± 102.4 | 353.6 ± 66.9 | <0.001 | <0.001 | <0.001 | <0.001 |
| Left ventricle | 407.9 ± 97.6 | 326.2 ± 68.9 | <0.001 | <0.001 | <0.001 | <0.001 |
| Right ventricle | 405.5 ± 135.3 | 335.3 ± 118.9 | 0.024 | 0.040 | 0.029 | 0.021 |
| Left coronary artery | 369.5 ± 117.1 | 299.4 ± 101.3 | 0.009 | 0.015 | 0.011 | 0.009 |
| Epicardial fat | 99.8 ± 25.5 | 93.9 ± 25.6 | 0.335 | 0.404 | 0.367 | 0.336 |
| Noise | ||||||
| Pulmonary artery | 30.3 ± 10.9 | 22.3 ± 8.5 | 0.001 | <0.001 | <0.001 | 0.001 |
| Ascending Aorta | 27.4 ± 9.2 | 22.3 ± 9.1 | 0.057 | 0.061 | 0.049 | 0.060 |
| Left ventricle | 36.6 ± 15.7 | 31.9 ± 22.9 | 0.302 | 0.188 | 0.212 | 0.269 |
| Right ventricle | 46.3 ± 21.7 | 40.8 ± 24.7 | 0.315 | 0.231 | 0.258 | 0.303 |
| Left main coronary artery | 67.9 ± 28.5 | 48.3 ± 21.9 | 0.002 | 0.001 | 0.002 | 0.002 |
| Epicardial fat | 33.9 ± 14.3 | 32.0 ± 20.0 | 0.649 | 0.657 | 0.594 | 0.645 |
| CNR aorta | 12.6 ± 4.7 | 13.8 ± 8.9 | 0.463 | 0.419 | 0.391 | 0.453 |
| SNR aorta | 16.9 ± 6.9 | 17.7 ± 6.9 | 0.623 | 0.460 | 0.529 | 0.666 |
Fig. 3The right coronary artery in patients of similar BMI scanned with Triple-Rule-Out CT angiography at 100 kV (a) and 120 kV (b)