Masahiko Goya1, Michio Nagashima1, Ken-Ichi Hiroshima1, Kentaro Hayashi1, Yu Makihara1, Masato Fukunaga1, Yoshimori An1, Masatsugu Ohe1, So-Ichiro Yamazato1, Ko-Ichiro Sonoda1, Kennosuke Yamashita1, Kouji Katayama2, Tomoaki Ito2, Harushi Niu2, Kenji Ando1, Hiroyoshi Yokoi1, Masashi Iwabuchi1.
Abstract
BACKGROUND: Lead extraction using laser sheaths is performed mainly for cardiac implantable electronic device (CIED) infections. However, there are few reports concerning the management of CIED infections in Japan. METHODS ANDEntities:
Keywords: Defibrillator; Infection; Laser; Lead extraction; Pacemaker
Year: 2016 PMID: 27588155 PMCID: PMC4996845 DOI: 10.1016/j.joa.2016.02.004
Source DB: PubMed Journal: J Arrhythm ISSN: 1880-4276
Summary data of the baseline characteristics of the study patients.
| Summary data of the patients | |
|---|---|
| No. of patients | 183 |
| Gender | Male: 131, female: 52 |
| Male, | 131 (72) |
| Age (years) | 72.2±14.3 |
| BMI (kg/m2) | 22.1±3.8 |
| WBC | 5970±1620 |
| Cr (mg/dl) | 1.1±1.4 |
| CRP (mg/dl) | 1.1±2.8 |
| Device type, n (%) | |
| Pacemaker | 136 (74.3) |
| CRT-P | 2 (1.1) |
| ICD | 28 (15.3) |
| CRT-D | 17 (9.3) |
BMI: body mass index, WBC: white blood cell, Cr: serum creatinine, CRP: C-reactive protein, CRT: cardiac resynchronization therapy, P: pacing, ICD: implantable cardioverter defibrillator, D: defibrillator
Summary data of 450 extracted leads.
| No. of extracted leads | 450 | 88.5±77.6 |
| Atrial leads | 170 | 92.3±76.4 |
| Active fixation atrial lead | 53 | |
| Passive fixation atrial lead | 113 | |
| Others | 4 | |
| Ventricular leads (except ICD) | 181 | 102.5±90.1 |
| Active fixation ventricular lead | 46 | |
| Passive fixation ventricular lead | 106 | |
| Others | 29 | |
| Coronary sinus lead | 20 | 34.9±27.6 |
| ICD lead | 79 | 62.4±35.5 |
| Dual coil active fixation ICD lead | 63 | |
| Dual coil passive fixation ICD lead | 10 | |
| Single coil active fixation ICD lead | 4 | |
| Single coil passive fixation ICD lead | 2 |
Mo: months.
Summary data of the pathogens isolated from the study patients.
| Organism | No. | % |
|---|---|---|
| Coagulase-negative | 55 | 30.1 |
| Methicillin-sensitive | 48 | 26.2 |
| Methicillin-resistant | 20 | 10.9 |
| | 1 | 0.5 |
| | 1 | 0.5 |
| | 8 | 4.4 |
| | 3 | 1.6 |
| | 3 | 1.6 |
| | 2 | 1.1 |
| | 1 | 0.5 |
| 4 | 2.2 | |
| | 2 | 1.1 |
| | 2 | 1.1 |
| 33 | 18 |
Fig. 1Left: A vegetation (2.5 cm) attached to the leads was detected by transesophageal echo. Right: A vegetation removed by open heart surgery is shown.
Fig. 2Microbiology of the cardiac implantable electronic device infections (n=183).