| Literature DB >> 27491453 |
Anastasia Lapkovska1, Anja M Palmowski-Wolfe1, Margarita G Todorova2.
Abstract
BACKGROUND: In infant ERG recordings skin electrodes frequently result in a better compliance. In order to assess the quality of such recordings, we compared the recording characteristics of DTL microfiber and Neuroline surface electrodes using a modified ISCEV protocol in the Mini Ganzfeld ERG.Entities:
Keywords: DTL microfiber electrode; Full-field ERG; ISCEV protocol; Mini Ganzfeld ERG; Neuroline surface electrode
Mesh:
Year: 2016 PMID: 27491453 PMCID: PMC4974799 DOI: 10.1186/s12886-016-0311-4
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1a The microfiber (DTL Plus) electrode is placed across the cornea over the lower eyelid (right eye). b The Neuroline skin electrode is placed on the surface of the left lower lid (left eye)
Fig. 2Depicts for comparison representative ERG traces recorded with both electrode types as follows: a scotopic, b standard flash cone, c light-adapted 3.0 flicker, d red cone
Fig. 3Box plots of the scotopic a-wave amplitudes (a), b-wave amplitudes (c), a-wave (b) and b-wave implicit times (d). With increasing luminance the a-wave, as well as the b-wave amplitude got larger resulting in an increase of the DTL: Neuroline amplitude-relationship from 1:1 in lower stimulus intensities to almost 1:4 in higher stimulus intensities. For all box plots, recordings obtained with a DTL electrode are plotted to the left, while recordings obtained with a Neuroline skin electrode are plotted to the right. The box length is the inter-quartile range. The line in bold depicts the median. Values between 1.3 and 3.0 box length represent outliers (open circles) and those with values more than 3.0 box length represent extremes (arrow heads). The respective p-values are shown on the top right hand side of the graphs (multivariate ANOVA)
represents the mean, standard deviation, 95 % confidence interval, as well as their respective p-values (ANOVA based on mixed effects model) of the a- and b-wave amplitudes, as well as the corresponding implicit times of the scotopic ERG for both electrode types
| ERG Recordings | Electrodes | Mean ± SD | 95 % confidence interval | P-values between eyes | P-values between Elctrodes |
|---|---|---|---|---|---|
| A-wave amplitudes | |||||
| 0.01 cd.s/m2 | DTL | −29.27 ± 37.15 | −43.04,−15.50 | 0.811 | 0.384 |
| 0.03 cd.s/m2 | DTL | −38.64 ± 41.64 | −49.64,−25.01 | 0.347 | 0.019* |
| 0.1 cd.s/m2 | DTL | −60.76 ± 45.91 | −71.38,−46.41 | 0.289 | <0.001* |
| 0.3 cd.s/m2 | DTL | −91.18 ± 50.84 | −102.79,−75.34 | 0.630 | <0.001* |
| 1.0 cd.s/m2 | DTL | −135.63 ± 79.56 | −155.86,−111.57 | 0.739 | <0.001* |
| 3.0 cd.s/m2 | DTL | −197.21 ± 96.78 | −219.68,−169.12 | 0.120 | <0.001* |
| B-wave amplitudes | |||||
| 0.01 cd.s/m2 | DTL | 197.49 ± 92.05 | 167.65, 227.33 | 0.757 | <0.001* |
| 0.03 cd.s/m2 | DTL | 254.35 ± 108.02 | 220.85, 278.81 | 0.306 | <0.001* |
| 0.1 cd.s/m2 | DTL | 246.02 ± 117.42 | 209.02, 272.11 | 0.380 | <0.001* |
| 0.3 cd.s/m2 | DTL | 308.60 ± 105.02 | 274.70, 330.52 | 0.260 | <0.001* |
| 1.0 cd.s/m2 | DTL | 371.47 ± 92.31 | 341.58, 392.78 | 0.213 | <0.001* |
| 3.0 cd.s/m2 | DTL | 394.34 ± 133.71 | 354.93, 428.03 | 0.460 | <0.001* |
| A-wave implicit times | |||||
| 0.01 cd.s/m2 | DTL | 38.98 ± 7.75 | 34.73, 43.22 | 0.335 | 0.176 |
| 0.03 cd.s/m2 | DTL | 34.17 ± 9.15 | 31.07, 37.62 | 0.640 | 0.645 |
| 0.1 cd.s/m2 | DTL | 28.14 ± 2.09 | 26.80, 29.35 | 0.011* | 0.221 |
| 0.3 cd.s/m2 | DTL | 25.30 ± 0.87 | 23.87, 26.71 | 0.407 | 0.756 |
| 1.0 cd.s/m2 | DTL | 22.97 ± 2.13 | 21.63, 24.31 | 0.986 | 0.124 |
| 3.0 cd.s/m2 | DTL | 17.81 ± 3.59 | 16.31, 19.45 | 0.416 | 0.853 |
| B-wave implicit times | |||||
| 0.01 cd.s/m2 | DTL | 86.43 ± 12.00 | 80.37, 92.50 | 0.576 | 0.052 |
| 0.03 cd.s/m2 | DTL | 73.17 ± 12.93 | 68.61, 78.10 | 0.985 | 0.048* |
| 0.1 cd.s/m2 | DTL | 61.48 ± 10.18 | 57.69, 66.00 | 0.671 | 0.377 |
| 0.3 cd.s/m2 | DTL | 56.69 ± 14.24 | 51.71, 58.42 | 0.237 | 0.193 |
| 1.0 cd.s/m2 | DTL | 55.08 ± 8.43 | 51.82, 58.34 | 0.954 | 0.291 |
| 3.0 cd.s/m2 | DTL | 51.64 ± 5.62 | 48.28, 54.94 | 0.197 | 0.196 |
Statistically significant p-values (<0.05) are labeled with asteriks
Fig. 4Box plots of the light-adapted 3.0 responses. a depicts box plots of the a-, and b-wave amplitudes, whereas b depict box plots of the a-, and b-wave implicit times recorded with both electrode types. For all box plots, recordings obtained with a DTL electrode are plotted to the left, while recordings obtained with a Neuroline skin electrode are plotted to the right. The box length is the inter-quartile range. The line in bold depicts the median. Values between 1.3 and 3.0 box length represent outliers (open circles) and those with values more than 3.0 box length represent extremes (arrow heads). The respective p-values are shown on the top right hand side of the graphs (multivariate ANOVA)
Fig. 6Respective box plots for a-wave amplitudes (a), b-wave amplitudes (c), a-wave (b) and b-wave implicit times (d) of the red cone response. In analogy to the scotopic response, with increasing stimulus intensity the a- as well as the b-red cone response got larger in the DTL electrode recordings compared to the Neuroline skin electrode recordings. A novel finding is the implicit time delay of the red a- and b-waves with the DTL electrode when compared to the recordings with the Neuroline skin electrode (Table 4). For all box plots, recordings obtained with a DTL electrode are plotted to the left, while recordings obtained with a Neuroline skin electrode are plotted to the right. The box length is the inter-quartile range. The line in bold depicts the median. Values between 1.3 and 3.0 box length represent outliers (open circles) and those with values more than 3.0 box length represent extremes (arrow heads). The respective p-values are shown on the top right hand side of the graphs (multivariate ANOVA)
Fig. 5Shows the light-adapted 3.0 flicker box plots. The trough- and peak-amplitudes are presented in a. The data of the corresponding implicit times are shown in b. For all box plots, recordings obtained with a DTL electrode are plotted to the left, while recordings obtained with a Neuroline skin electrode are plotted to the right. The box length is the inter-quartile range. The line in bold depicts the median. Values between 1.3 and 3.0 box length represent outliers (open circles) and those with values more than 3.0 box length represent extremes (arrow heads). The respective p-values are shown on the top right hand side of the graphs (multivariate ANOVA)
shows the mean, standard deviation, 95 % confidence interval, as well as their respective p-values (ANOVA based on mixed effects model) of the a- and b-wave amplitudes, as well as the corresponding implicit times of the photopic flash ERG for both electrode types
| ERG Recordings | Electrodes | Mean ± SD | 95 % confidence interval | P-values between eyes | P-values between elctrodes |
|---|---|---|---|---|---|
| A-wave amplitudes | |||||
| DTL | −42.85 ± 24.33 | −49.06,−35.39 | 0.069 | <0.001* | |
| B-wave amplitudes | |||||
| DTL | 124.95 ± 32.90 | 114.68, 137.45 | 0.868 | <0.001* | |
| A-wave implicit times | |||||
| DTL | 15.03 ± 1.89 | 13.47, 16.53 | 0.735 | 0.427 | |
| B-wave implicit times | |||||
| DTL | 29.76 ± 4.28 | 28.05, 31.35 | 0.588 | 0.369 | |
Statistically significant p-values (<0.05) are labeled with asteriks
presents the mean, standard deviation, 95 % confidence interval, as well as their respective p-values (ANOVA based on mixed effects model) of the peak- and trough- amplitudes, as well as the corresponding implicit times of the 30Hz ERG for both electrode types
| ERG Recordings | Electrodes | Mean ± SD | 95 % confidence interval | P-values between eyes | P-values between elctrodes |
|---|---|---|---|---|---|
| T-wave amplitudes | |||||
| DTL | −42.54 ± 20.92 | −47.99,−36.20 | 0.091 | <0.001* | |
| P-wave amplitudes | |||||
| DTL | 97.64 ± 26.46 | 89.54, 104.85 | 0.184 | <0.001* | |
| T-wave implicit times | |||||
| DTL | 14.75 ± 2.80 | 13.73, 16.01 | 0.333 | 0.903 | |
| P-wave implicit times | |||||
| DTL | 27.58 ± 1.64 | 26.93, 28.32 | 0.633 | 0.865 | |
Statistically significant p-values (<0.05) are labeled with asteriks
shows the mean, standard deviation, 95 % confidence interval, as well as their respective p-values (ANOVA based on mixed effects model) of the a- and b-wave amplitudes, as well as the corresponding implicit times of the photopic red ERGs for both electrode types
| ERG Recordings | Electrodes | Mean ± SD | 95 % confidence interval | P-values between eyes | P-values between elctrodes |
|---|---|---|---|---|---|
| A-wave amplitudes | |||||
| 0.3 cd.s/m2 | DTL | −11.46 ± 12.56 | −14.96,−8.15 | 0.787 | 0.002* |
| 0.5 cd.s/m2 | DTL | −14.44 ± 17.73 | −18.56,−9.37 | 0.185 | 0.005* |
| 0.9 cd.s/m2 | DTL | −20.89 ± 18.35 | −25.16,−15.56 | 0.250 | <0.001* |
| 3.0 cd.s/m2 | DTL | −36.30 ± 17.36 | −40.53,−31.14 | 0.082 | <0.001* |
| B-wave amplitudes | |||||
| 0.3 cd.s/m2 | DTL | 35.41 ± 15.40 | 31.41, 39.80 | 0.784 | <0.001* |
| 0.5 cd.s/m2 | DTL | 57.06 ± 23.26 | 50.64, 63.17 | 0.668 | <0.001* |
| 0.9 cd.s/m2 | DTL | 83.93 ± 31.70 | 75.31, 92.33 | 0.818 | <0.001* |
| 3.0 cd.s/m2 | DTL | 100.75 ± 29.05 | 94.02, 109.42 | 0.241 | <0.001* |
| A-wave implicit times | |||||
| 0.3 cd.s/m2 | DTL | 17.22 ± 1.79 | 16.17, 18.23 | 0.231 | 0.001* |
| 0.5 cd.s/m2 | DTL | 16.62 ± 1.51 | 15.86, 17.28 | 0.210 | 0.018* |
| 0.9 cd.s/m2 | DTL | 15.97 ± 1.21 | 14.82, 17.02 | 0.816 | 0.034* |
| 3.0 cd.s/m2 | DTL | 15.02 ± 0.84 | 14.07, 15.94 | 0.241 | 0.039* |
| B-wave implicit times | |||||
| 0.3 cd.s/m2 | DTL | 27.92 ± 1.85 | 27.14, 28.84 | 0.245 | 0.034* |
| 0.5 cd.s/m2 | DTL | 28.76 ± 1.21 | 28.21, 29.29 | 0.657 | 0.016* |
| 0.9 cd.s/m2 | DTL | 29.50 ± 1.54 | 28.76, 30.26 | 0.410 | 0.005* |
| 3.0 cd.s/m2 | DTL | 32.22 ± 1.27 | 31.62, 32.83 | 0.828 | 0.003* |
Statistically significant p-values (<0.05) are labeled with asteriks