Corrosion Damage and Wear Mechanisms in Long-Term Retrieved CoCr Femoral Components for Total Knee Arthroplasty.

Christina M Arnholt1, Daniel W MacDonald1, Arthur L Malkani2, Gregg R Klein3, Clare M Rimnac4, Steven M Kurtz5, Sevi B Kocagoz1, Jeremy L Gilbert6.   

Abstract

BACKGROUND: Metal debris and ion release has raised concerns in joint arthroplasty. The purpose of this study was to characterize the sources of metallic ions and particulate debris released from long-term (in vivo >15 years) total knee arthroplasty femoral components.
METHODS: A total of 52 CoCr femoral condyles were identified as having been implanted for more than 15 years. The femoral components were examined for incidence of 5 types of damage (metal-on-metal wear due to historical polyethylene insert failure, mechanically assisted crevice corrosion at taper interfaces, cement interface corrosion, third-body abrasive wear, and inflammatory cell-induced corrosion [ICIC]). Third-body abrasive wear was evaluated using the Hood method for polyethylene components and a similar method quantifying surface damage of the femoral condyle was used. The total area damaged by ICIC was quantified using digital photogrammetry.
RESULTS: Surface damage associated with corrosion and/or CoCr debris release was identified in 51 (98%) CoCr femoral components. Five types of damage were identified: 98% of femoral components exhibited third-body abrasive wear (mostly observed as scratching, n = 51/52), 29% of femoral components exhibited ICIC damage (n = 15/52), 41% exhibited cement interface damage (n = 11/27), 17% exhibited metal-on-metal wear after wear-through of the polyethylene insert (n = 9/52), and 50% of the modular femoral components exhibited mechanically assisted crevice corrosion taper damage (n = 2/4). The total ICIC-damaged area was an average of 0.11 ± 0.12 mm2 (range: 0.01-0.46 mm2).
CONCLUSION: Although implant damage in total knee arthroplasty is typically reported with regard to the polyethylene insert, the results of this study demonstrate that abrasive and corrosive damage occurs on the CoCr femoral condyle in vivo.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  CoCr femoral component TKA; ICIC damage; fretting corrosion; revision TKA; third-body wear

Mesh:

Substances:

Year:  2016        PMID: 27426028      PMCID: PMC5107165          DOI: 10.1016/j.arth.2016.05.006

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  16 in total

1.  A multicenter retrieval study of the taper interfaces of modular hip prostheses.

Authors:  Jay R Goldberg; Jeremy L Gilbert; Joshua J Jacobs; Thomas W Bauer; Wayne Paprosky; Sue Leurgans
Journal:  Clin Orthop Relat Res       Date:  2002-08       Impact factor: 4.176

2.  In vivo degradation of polyethylene liners after gamma sterilization in air.

Authors:  Steven M Kurtz; Clare M Rimnac; William J Hozack; Joseph Turner; Michele Marcolongo; Victor M Goldberg; Matthew J Kraay; Avram A Edidin
Journal:  J Bone Joint Surg Am       Date:  2005-04       Impact factor: 5.284

3.  Digital photogrammetry for quantitative wear analysis of retrieved TKA components.

Authors:  J C Grochowsky; L W Alaways; R Siskey; E Most; S M Kurtz
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2006-11       Impact factor: 3.368

4.  2006 Otto Aufranc Award Paper: significance of in vivo degradation for polyethylene in total hip arthroplasty.

Authors:  Steven M Kurtz; William J Hozack; James J Purtill; Michele Marcolongo; Matthew J Kraay; Victor M Goldberg; Peter F Sharkey; Javad Parvizi; Clare M Rimnac; Avram A Edidin
Journal:  Clin Orthop Relat Res       Date:  2006-12       Impact factor: 4.176

5.  Characterisation of the surface topography, tomography and chemistry of fretting corrosion product found on retrieved polished femoral stems.

Authors:  M Bryant; M Ward; R Farrar; R Freeman; K Brummitt; J Nolan; A Neville
Journal:  J Mech Behav Biomed Mater       Date:  2013-12-07

6.  Is increased modularity associated with increased fretting and corrosion damage in metal-on-metal total hip arthroplasty devices?: a retrieval study.

Authors:  Genymphas B Higgs; Josa A Hanzlik; Daniel W MacDonald; Jeremy L Gilbert; Clare M Rimnac; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2013-08-01       Impact factor: 4.757

Review 7.  Corrosion of metal orthopaedic implants.

Authors:  J J Jacobs; J L Gilbert; R M Urban
Journal:  J Bone Joint Surg Am       Date:  1998-02       Impact factor: 5.284

8.  Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses.

Authors:  R W Hood; T M Wright; A H Burstein
Journal:  J Biomed Mater Res       Date:  1983-09

9.  Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty.

Authors:  R C Wasielewski; J O Galante; R M Leighty; R N Natarajan; A G Rosenberg
Journal:  Clin Orthop Relat Res       Date:  1994-02       Impact factor: 4.176

10.  Wear in total knee arthroplasty--just a question of polyethylene?: Metal ion release in total knee arthroplasty.

Authors:  Jan Philippe Kretzer; Joern Reinders; Robert Sonntag; Sebastien Hagmann; Marcus Streit; Sebastian Jeager; Babak Moradi
Journal:  Int Orthop       Date:  2013-11-12       Impact factor: 3.075

View more
  14 in total

1.  What Is the Incidence of Cobalt-Chromium Damage Modes on the Bearing Surface of Contemporary Femoral Component Designs for Total Knee Arthroplasty?

Authors:  Christina M Arnholt; Daniel W MacDonald; Gregg R Klein; Harold E Cates; Clare M Rimnac; Steven M Kurtz; Sevi Kocagoz; Antonia F Chen
Journal:  J Arthroplasty       Date:  2018-05-17       Impact factor: 4.757

Review 2.  Total Alloplastic Temporomandibular Joint Replacement.

Authors:  Poonam Yadav; Ajoy Roychoudhury; Rudra Deo Kumar; Ongkila Bhutia; Tsering Bhutia; Bhaskar Aggarwal
Journal:  J Maxillofac Oral Surg       Date:  2021-08-18

3.  The influence of HLA genotype on the development of metal hypersensitivity following joint replacement.

Authors:  David J Langton; Rohan M Bhalekar; Thomas J Joyce; Stephen P Rushton; Benjamin J Wainwright; Matthew E Nargol; Nish Shyam; Benedicte A Lie; Moreica B Pabbruwe; Alan J Stewart; Susan Waller; Shonali Natu; Renne Ren; Rachelle Hornick; Rebecca Darlay; Edwin P Su; Antoni V F Nargol
Journal:  Commun Med (Lond)       Date:  2022-06-24

4.  Wear Performance of Calcium Carbonate-Containing Knee Spacers.

Authors:  Ulrike Mueller; Joern Reinders; Sydney Smith-Romanski; Jan Philippe Kretzer
Journal:  Materials (Basel)       Date:  2017-07-15       Impact factor: 3.623

5.  Diagnostic guidelines for the histological particle algorithm in the periprosthetic neo-synovial tissue.

Authors:  G Perino; S Sunitsch; M Huber; D Ramirez; J Gallo; J Vaculova; S Natu; J P Kretzer; S Müller; P Thomas; M Thomsen; M G Krukemeyer; H Resch; T Hügle; W Waldstein; F Böettner; T Gehrke; S Sesselmann; W Rüther; Z Xia; E Purdue; V Krenn
Journal:  BMC Clin Pathol       Date:  2018-08-25

6.  High wear resistance of femoral components coated with titanium nitride: a retrieval analysis.

Authors:  Christian Fabry; Carmen Zietz; Axel Baumann; Reinhard Ehall; Rainer Bader
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-05-20       Impact factor: 4.342

7.  Taper Corrosion and Adverse Local Tissue Reactions in Patients with a Modular Knee Prosthesis.

Authors:  Tom Christiner; Moreica B Pabbruwe; Alan M Kop; Jeremy Parry; Gavin Clark; Dermot Collopy
Journal:  JB JS Open Access       Date:  2018-11-13

8.  No improvement in reducing outliers in coronal axis alignment with patient-specific instrumentation.

Authors:  Uwe Maus; Carlos J Marques; David Scheunemann; Frank Lampe; Djordje Lazovic; Hagen Hommel; Dennis Vogel; Martin Haunschild; Tilman Pfitzner
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-10-25       Impact factor: 4.342

Review 9.  The contribution of the histopathological examination to the diagnosis of adverse local tissue reactions in arthroplasty.

Authors:  Giorgio Perino; Ivan De Martino; Lingxin Zhang; Zhidao Xia; Jiri Gallo; Shonali Natu; David Langton; Monika Huber; Anastasia Rakow; Janosch Schoon; Enrique Gomez-Barrena; Veit Krenn
Journal:  EFORT Open Rev       Date:  2021-06-28

10.  Retrieval Analysis of Modern Knee Tumor Megaendoprosthesis Shows Considerable Volumetric Metal Wear Generated at the Rotating Hinge.

Authors:  Therese Bormann; Sebastian Jäger; J Philippe Kretzer; Laura Nebel; Lucas Clarius; Georg Omlor; Rudi Bitsch; Burkhard Lehner
Journal:  Materials (Basel)       Date:  2020-03-26       Impact factor: 3.623