Wear in total knee arthroplasty--just a question of polyethylene?: Metal ion release in total knee arthroplasty.

Jan Philippe Kretzer1, Joern Reinders, Robert Sonntag, Sebastien Hagmann, Marcus Streit, Sebastian Jeager, Babak Moradi.   

Abstract

PURPOSE: Biological reactions against wear particles are a common cause for revision in total knee arthroplasty. To date, wear has mainly been attributed to polyethylene. However, the implants have large metallic surfaces that also could potentially lead to metal wear products (metal ions and debris). The aim of this study was to determine the local release of cobalt, chromium, molybdenum and titanium in total knee arthroplasty during a standard knee wear test.
METHODS: Four moderately conforming fixed-bearing implants were subjected to physiological loadings and motions for 5×10(6) walking cycles in a knee wear simulator. Polyethylene wear was determined gravimetrically and the release of metallic wear products was measured using high resolution-inductively coupled plasma-mass spectrometry.
RESULTS: A polyethylene wear rate of 7.28 ± 0.27 mg/10(6) cycles was determined and the cumulative mass of released metals measured 1.63 ± 0.28 mg for cobalt, 0.47 ± 0.06 mg for chromium, 0.42 ± 0.06 mg for molybdenum and 1.28 ± 0.14 mg for titanium.
CONCLUSION: For other metallic implants such as metal-on-metal total hip arthroplasty, the metal wear products can interact with the immune system, potentially leading to immunotoxic effects. In this study about 12 % by weight of the wear products were metallic, and these particles and ions may become clinically relevant for patients sensitive to these materials in particular. Non-metallic materials (e.g. ceramics or suitable coatings) may be considered for an alternative treatment for those patients.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24218115      PMCID: PMC3923952          DOI: 10.1007/s00264-013-2162-4

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  27 in total

1.  Potential thermal artifacts in hip joint wear simulators.

Authors:  Z Lu; H McKellop; P Liao; P Benya
Journal:  J Biomed Mater Res       Date:  1999

2.  Surface roughness of retrieved CoCrMo alloy femoral components from PCA artificial total knee joints.

Authors:  L Que; L D Topoleski; N L Parks
Journal:  J Biomed Mater Res       Date:  2000

3.  Oxidative properties and surface damage mechanisms of remelted highly crosslinked polyethylenes in total knee arthroplasty.

Authors:  Daniel W MacDonald; Genymphas Higgs; Javad Parvizi; Gregg Klein; Mark Hartzband; Harlan Levine; Matthew Kraay; Clare M Rimnac; Steven M Kurtz
Journal:  Int Orthop       Date:  2013-02-10       Impact factor: 3.075

4.  Knee biomechanics and total knee replacement.

Authors:  T P Andriacchi; T S Stanwyck; J O Galante
Journal:  J Arthroplasty       Date:  1986       Impact factor: 4.757

5.  Static and dynamic loading patterns in knee joints with deformities.

Authors:  I J Harrington
Journal:  J Bone Joint Surg Am       Date:  1983-02       Impact factor: 5.284

6.  The effects of proteins on the friction and lubrication of artificial joints.

Authors:  S C Scholes; A Unsworth
Journal:  Proc Inst Mech Eng H       Date:  2006-08       Impact factor: 1.617

7.  Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study.

Authors:  Hans-Georg Willert; Gottfried H Buchhorn; Afshin Fayyazi; Renata Flury; Markus Windler; Georg Köster; Christoph H Lohmann
Journal:  J Bone Joint Surg Am       Date:  2005-01       Impact factor: 5.284

Review 8.  Health-related quality of life in total hip and total knee arthroplasty. A qualitative and systematic review of the literature.

Authors:  Olivier Ethgen; Olivier Bruyère; Florent Richy; Charles Dardennes; Jean-Yves Reginster
Journal:  J Bone Joint Surg Am       Date:  2004-05       Impact factor: 5.284

Review 9.  Mechanism and clinical significance of wear debris-induced osteolysis.

Authors:  H C Amstutz; P Campbell; N Kossovsky; I C Clarke
Journal:  Clin Orthop Relat Res       Date:  1992-03       Impact factor: 4.176

10.  In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling.

Authors:  J L Gilbert; C A Buckley; J J Jacobs
Journal:  J Biomed Mater Res       Date:  1993-12
View more
  22 in total

1.  Knee arthroplasty today.

Authors:  Francesco Falez
Journal:  Int Orthop       Date:  2014-02       Impact factor: 3.075

2.  Are Serum Metal Ion Levels a Concern at Mid-term Followup of Revision Knee Arthroplasty With a Metal-on-metal Hinge Design?

Authors:  Antonio Klasan; Esko Meine; Susanne Fuchs-Winkelmann; Turgay Efe; Friedrich Boettner; Thomas Jan Heyse
Journal:  Clin Orthop Relat Res       Date:  2019-09       Impact factor: 4.176

Review 3.  When 1+1>2: Nanostructured composites for hard tissue engineering applications.

Authors:  Vuk Uskoković
Journal:  Mater Sci Eng C Mater Biol Appl       Date:  2015-08-01       Impact factor: 7.328

4.  What Is the Incidence of Cobalt-Chromium Damage Modes on the Bearing Surface of Contemporary Femoral Component Designs for Total Knee Arthroplasty?

Authors:  Christina M Arnholt; Daniel W MacDonald; Gregg R Klein; Harold E Cates; Clare M Rimnac; Steven M Kurtz; Sevi Kocagoz; Antonia F Chen
Journal:  J Arthroplasty       Date:  2018-05-17       Impact factor: 4.757

5.  Does Kinematic Alignment Increase Polyethylene Wear Compared With Mechanically Aligned Components? A Wear Simulation Study.

Authors:  Stefan Schroeder; Mareike Schonhoff; Maximilian Uhler; Steffen Braun; Sebastian Jaeger; Tobias Renkawitz; Jan Philippe Kretzer
Journal:  Clin Orthop Relat Res       Date:  2022-05-17       Impact factor: 4.755

6.  Functionalised High-Performance Oxide Ceramics with Bone Morphogenic Protein 2 (BMP-2) Induced Ossification: An In Vivo Study.

Authors:  Filippo Migliorini; Jörg Eschweiler; Nicola Maffulli; Frank Hildebrand; Hanno Schenker
Journal:  Life (Basel)       Date:  2022-06-09

7.  The influence of HLA genotype on the development of metal hypersensitivity following joint replacement.

Authors:  David J Langton; Rohan M Bhalekar; Thomas J Joyce; Stephen P Rushton; Benjamin J Wainwright; Matthew E Nargol; Nish Shyam; Benedicte A Lie; Moreica B Pabbruwe; Alan J Stewart; Susan Waller; Shonali Natu; Renne Ren; Rachelle Hornick; Rebecca Darlay; Edwin P Su; Antoni V F Nargol
Journal:  Commun Med (Lond)       Date:  2022-06-24

8.  Ion release in ceramic bearings for total hip replacement: Results from an in vitro and an in vivo study.

Authors:  Jan Philippe Kretzer; Ulrike Mueller; Marcus R Streit; Hartmuth Kiefer; Robert Sonntag; Robert M Streicher; Joern Reinders
Journal:  Int Orthop       Date:  2017-07-20       Impact factor: 3.075

9.  Corrosion Damage and Wear Mechanisms in Long-Term Retrieved CoCr Femoral Components for Total Knee Arthroplasty.

Authors:  Christina M Arnholt; Daniel W MacDonald; Arthur L Malkani; Gregg R Klein; Clare M Rimnac; Steven M Kurtz; Sevi B Kocagoz; Jeremy L Gilbert
Journal:  J Arthroplasty       Date:  2016-05-12       Impact factor: 4.757

10.  Large Soft-tissue Mass Formation After Revision Total Knee Arthroplasty: An Unusual Case of Adverse Reaction to Metal Debris and Review of the Literature.

Authors:  Petri Bellova; Jessica Pablik; Maik Stiehler; Adrian Dragu; Jörg Lützner
Journal:  Arthroplast Today       Date:  2021-06-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.